Photoresist cross-sectional shape change caused by scanning electron microscope-induced shrinkage

被引:13
|
作者
Ohashi, Takeyoshi [1 ]
Sekiguchi, Tomoko [1 ]
Yamaguchi, Atsuko [1 ]
Tanaka, Junichi [1 ]
Kawada, Hiroki [2 ]
机构
[1] Hitachi Ltd, Cent Res Lab, Kokubunji, Tokyo 1858601, Japan
[2] Hitachi High Technol Corp, Hitachinaka, Ibaraki 3128504, Japan
来源
JOURNAL OF MICRO-NANOLITHOGRAPHY MEMS AND MOEMS | 2015年 / 14卷 / 03期
关键词
photoresist; shrinkage; scanning transmission electron microscope; cross-sectional measurement; metrology; critical-dimension scanning electron microscope; ARF-RESIST; SEM MEASUREMENT; IMPACT; MECHANISM; METROLOGY; ACCURACY;
D O I
10.1117/1.JMM.14.3.034001
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Change in the cross-sectional profile of a photoresist (PR) pattern due to shrinkage was evaluated to investigate the mechanism of electron beam-induced shrinkage. A scanning transmission electron microscope (STEM) was used to observe the cross-sectional profiles of PR lines after atomic-layer deposition of metal oxide and carbon deposition on the sample surface. A HfO2 thin layer enhanced the profile contrast in the STEM measurements without blurring the edge, which enabled the precise cross-sectional measurement of the PR patterns. We found interesting features associated with shrinkage from the detailed profile change obtained using this method, such as a rounding of the pattern top, a necking of the sidewall profile, a rounding of the foot in the pattern on the organic underlying layer, and voltage-independent sidewall shrinkage under a large electron beam dose. These behaviors along with the results from a Monte Carlo simulation are discussed. Consequently, these observations experimentally clarified that the elastic deformation effect and the impact of the secondary electrons emitted from the spaces around the pattern into the sidewall are important to interpret the change in the shape of the pattern induced by shrinkage. (C) The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License.
引用
收藏
页数:8
相关论文
共 4 条
  • [1] Precise Measurement of Photoresist Cross-sectional Shape Change Caused by SEM-induced Shrinkage
    Ohashi, Takeyoshi
    Sekiguchi, Tomoko
    Yamaguchi, Atsuko
    Tanaka, Junichi
    Kawada, Hiroki
    METROLOGY, INSPECTION, AND PROCESS CONTROL FOR MICROLITHOGRAPHY XXVII, 2013, 8681
  • [2] Precise Cross-Sectional Measurement of Photoresist Shrinkage Caused by Electron Beam Irradiation
    Ohashi, Takeyoshi
    Sekiguchi, Tomoko
    Yamaguchi, Atsuko
    Tanaka, Junichi
    Kawada, Hiroki
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2013, 52 (06)
  • [3] Effect of Cross-Sectional Shape of Small Level Change on Walkability
    Goto, Yoshiaki
    Miyake, Mai
    Sato, Katsushi
    Kose, Satoshi
    ADVANCES IN DESIGN FOR INCLUSION, 2016, 500 : 315 - 322
  • [4] The influence of a change in the meniscus cross-sectional shape on the medio-lateral translation of the knee joint and meniscal extrusion
    Luczkiewicz, Piotr
    Daszkiewicz, Karol
    Witkowski, Wojciech
    Chroscielewski, Jacek
    Ferenc, Tomasz
    Baczkowski, Boguslaw
    PLOS ONE, 2018, 13 (02):