ACCURACY OF END-TIDAL CO2 CAPNOMETERS IN POST-CARDIAC SURGERY PATIENTS DURING CONTROLLED MECHANICAL VENTILATION

被引:19
|
作者
Heines, Serge J. H. [1 ]
Strauch, Ulrich [1 ]
Roekaerts, Paul M. H. J. [1 ]
Winkens, Bjorn [2 ]
Bergmans, Dennis C. J. J. [1 ]
机构
[1] Maastricht Univ, Dept Intens Care, Med Ctr, NL-6202 AZ Maastricht, Netherlands
[2] Maastricht Univ, Dept Methodol & Stat, NL-6202 AZ Maastricht, Netherlands
关键词
end-tidal carbon dioxide; capnography; capnometer; monitoring; carbon dioxide; blood gas analysis; CARBON-DIOXIDE; ARTERIAL; TRAUMA; CAPNOGRAPHY; CARE;
D O I
10.1016/j.jemermed.2012.11.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The determination of end-tidal carbon dioxide (etCO(2)) is very helpful in cardiac resuscitation for confirmation and monitoring of endotracheal tube placement and as an indicator of return of circulation and effectiveness of chest compressions. There is now also widespread use of capnometry on-site at emergency and trauma fields. Objective: We studied the accuracy and correlation of three capnometers (EMMA, Medtronic, and Evita) with partial pressure of arterial CO2 (PaCO2) measurements. Methods: The three capnometers were placed in-line in the ventilator tubing of the patient. Forty sedated and mechanically ventilated post-cardiac surgery patients were studied. Twenty consecutive etCO(2) values were collected simultaneously from all three monitors while drawing an arterial blood sample. Paired sample t-test and Pearson correlation were used to compare the capnometers and their correlation with PaCO2. Results: The correlation of etCO(2) measurements between all three capnometers was good (Emma vs. Evita: 0.874, Emma vs. Medtronic: 0.949, Evita vs. Medtronic: 0.878). The correlation of PaCO2 with the Evita is the lowest (0.671) as compared to the EMMA (0.693) and the Medtronic (0.727). The lowest dispersion of the difference between etCO(2) and PaCO2 was seen in EMMA (3.30), the highest in Evita (3.98). Conclusions: A good correlation between etCO(2) and PaCO2 was shown with the three capnometers in the present study. However, etCO(2) measurements were not valid to estimate PaCO2 in these patients. Therefore, capnometry cannot be used to replace serial blood gas analyses completely, but may be a good cardiopulmonary trend monitor and alerting system in catastrophic events. (c) 2013 Elsevier Inc.
引用
收藏
页码:130 / 134
页数:5
相关论文
共 50 条
  • [41] The effect of procedural end-tidal CO2 on infarct expansion during anterior circulation thrombectomy
    Parr, Matthew S.
    Salehani, Arsalaan
    Ogilvie, Mark
    Tabibian, B. Ethan
    Rahm, Sage
    Hale, Andrew T.
    Tsemo, Georges Bouobda
    Aluri, Akshay
    Kim, Jinsuh
    Mathru, Mali
    Jones, Jesse G. A.
    INTERVENTIONAL NEURORADIOLOGY, 2024, 30 (05) : 689 - 693
  • [42] Initial end-tidal CO2 partial pressure predicts outcomes of in-hospital cardiac arrest
    Wang, An-Yi
    Huang, Chien-Hua
    Chang, Wei-Tien
    Tsai, Min-Shan
    Wang, Chih-Hung
    Chen, Wen-Jone
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2016, 34 (12) : 2367 - 2371
  • [43] End-tidal CO2 in the diagnosis of fluid responsiveness - a systematic review
    Oberg, Emilie B.
    Jorgensen, Bent G.
    Berthelsen, Preben G.
    DANISH MEDICAL JOURNAL, 2019, 66 (09):
  • [44] The relationship between minute ventilation and end tidal CO2 in intubated and spontaneously breathing patients undergoing procedural sedation
    Mehta, Jaideep H.
    Williams, George W., II
    Harvey, Brian C.
    Grewal, Navneet K.
    George, Edward E.
    PLOS ONE, 2017, 12 (06):
  • [45] End-tidal CO2 and transcutaneous CO2: Are we ready to replace arterial CO2 in awake children?
    Dicembrino, Manuela
    Alejandra Barbieri, Irene
    Pereyra, Carla
    Leske, Vivian
    PEDIATRIC PULMONOLOGY, 2021, 56 (02) : 486 - 494
  • [46] Initial end-tidal CO2 is markedly elevated during cardiopulmonary resuscitation after asphyxial cardiac arrest
    Berg, RA
    Henry, C
    Otto, CW
    Sanders, AB
    Kern, KB
    Hilwig, RW
    Ewy, GA
    PEDIATRIC EMERGENCY CARE, 1996, 12 (04) : 245 - 248
  • [47] Sleep Transcutaneous vs. End-Tidal CO2 Monitoring for Patients with Neuromuscular Disease
    Won, Yu Hui
    Choi, Won Ah
    Lee, Jang Woo
    Bach, John Robert
    Park, Jinyoung
    Kang, Seong-Woong
    AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 2016, 95 (02) : 91 - 95
  • [48] Accuracy of capnography with a 30 foot nasal cannula for monitoring respiratory rate and end-tidal CO2 in children
    Mason, KP
    Burrows, PE
    Dorsey, MM
    Zurakowski, D
    Krauss, B
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2000, 16 (04) : 259 - 262
  • [49] Accuracy of Capnography with a 30 Foot Nasal Cannula for Monitoring Respiratory Rate and End-Tidal CO2 in Children
    Keira P. Mason
    Patricia E. Burrows
    Maureen M. Dorsey
    David Zurakowski
    Baruch Krauss
    Journal of Clinical Monitoring and Computing, 2000, 16 : 259 - 262
  • [50] Can end-tidal CO2 measurement replace arterial partial CO2 in emergency department respiratory distress management?
    Selki, Kudret
    Demir, Mehmet Cihat
    Senguldur, Erdinc
    Erdem, Emre
    Guldal, Hatice
    Tasdemir, Murat
    Kiciroglu, Alp Kaan Furkan
    Bogan, Mustafa
    MEDICINA INTENSIVA, 2024, 48 (09) : 511 - 519