ACCURACY OF END-TIDAL CO2 CAPNOMETERS IN POST-CARDIAC SURGERY PATIENTS DURING CONTROLLED MECHANICAL VENTILATION

被引:19
|
作者
Heines, Serge J. H. [1 ]
Strauch, Ulrich [1 ]
Roekaerts, Paul M. H. J. [1 ]
Winkens, Bjorn [2 ]
Bergmans, Dennis C. J. J. [1 ]
机构
[1] Maastricht Univ, Dept Intens Care, Med Ctr, NL-6202 AZ Maastricht, Netherlands
[2] Maastricht Univ, Dept Methodol & Stat, NL-6202 AZ Maastricht, Netherlands
关键词
end-tidal carbon dioxide; capnography; capnometer; monitoring; carbon dioxide; blood gas analysis; CARBON-DIOXIDE; ARTERIAL; TRAUMA; CAPNOGRAPHY; CARE;
D O I
10.1016/j.jemermed.2012.11.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The determination of end-tidal carbon dioxide (etCO(2)) is very helpful in cardiac resuscitation for confirmation and monitoring of endotracheal tube placement and as an indicator of return of circulation and effectiveness of chest compressions. There is now also widespread use of capnometry on-site at emergency and trauma fields. Objective: We studied the accuracy and correlation of three capnometers (EMMA, Medtronic, and Evita) with partial pressure of arterial CO2 (PaCO2) measurements. Methods: The three capnometers were placed in-line in the ventilator tubing of the patient. Forty sedated and mechanically ventilated post-cardiac surgery patients were studied. Twenty consecutive etCO(2) values were collected simultaneously from all three monitors while drawing an arterial blood sample. Paired sample t-test and Pearson correlation were used to compare the capnometers and their correlation with PaCO2. Results: The correlation of etCO(2) measurements between all three capnometers was good (Emma vs. Evita: 0.874, Emma vs. Medtronic: 0.949, Evita vs. Medtronic: 0.878). The correlation of PaCO2 with the Evita is the lowest (0.671) as compared to the EMMA (0.693) and the Medtronic (0.727). The lowest dispersion of the difference between etCO(2) and PaCO2 was seen in EMMA (3.30), the highest in Evita (3.98). Conclusions: A good correlation between etCO(2) and PaCO2 was shown with the three capnometers in the present study. However, etCO(2) measurements were not valid to estimate PaCO2 in these patients. Therefore, capnometry cannot be used to replace serial blood gas analyses completely, but may be a good cardiopulmonary trend monitor and alerting system in catastrophic events. (c) 2013 Elsevier Inc.
引用
收藏
页码:130 / 134
页数:5
相关论文
共 50 条
  • [31] A Novel Method of Distal End-Tidal CO2 Capnography in Intubated Infants: Comparison With Arterial CO2 and With Proximal Mainstream End-Tidal CO2
    Kugelman, Amir
    Zeiger-Aginsky, Dana
    Bader, David
    Shoris, Irit
    Riskin, Arieh
    PEDIATRICS, 2008, 122 (06) : E1219 - E1224
  • [32] Elevations in End-Tidal CO2 With CO2 Use During Pediatric Endoscopy With Airway Protection: Is This Physiologically Significant?
    Dike, Chinenye R.
    Pacheco, Andrew Huang
    Lyden, Elizabeth
    Freestone, David
    Choudhry, Ojasvini
    Bishop, Warren P.
    Shukry, Mohanad
    JOURNAL OF PEDIATRIC GASTROENTEROLOGY AND NUTRITION, 2023, 76 (05) : 660 - 666
  • [33] Evaluation of fully automated ventilation: a randomized controlled study in post-cardiac surgery patients
    Lellouche, Francois
    Bouchard, Pierre-Alexandre
    Simard, Serge
    L'Her, Erwan
    Wysocki, Marc
    INTENSIVE CARE MEDICINE, 2013, 39 (03) : 463 - 471
  • [34] Assessment of end-tidal CO2 in the pleural chest tube during lung volume reduction surgery
    Raffin, L
    Moutafis, M
    Dalibon, N
    ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2003, 22 (05): : 484 - 486
  • [35] Prehospital end-tidal CO2 as an early marker for transfusion requirement in trauma patients
    Wilson, Bryan R.
    Bruno, John
    Duckwitz, Marcus
    Akers, Natalie
    Jeanmonod, Donald
    Jeanmonod, Rebecca
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2021, 45 : 254 - 257
  • [36] End-tidal CO2 monitoring in emergency medicine
    Brambrink, AM
    ANAESTHESIST, 1997, 46 (07): : 604 - 612
  • [37] Continuous measurement of arterial and end-tidal carbon dioxide during cardiac surgery: Pa-ETCO2 gradient
    Myles, PS
    Story, DA
    Higgs, MA
    Buckland, MR
    ANAESTHESIA AND INTENSIVE CARE, 1997, 25 (05) : 459 - 463
  • [38] ARTERIAL TO END-TIDAL CO2 TENSION DIFFERENCE AFTER BILATERAL LUNG TRANSPLANTATION
    JELLINEK, H
    HIESMAYR, M
    SIMON, P
    KLEPETKO, W
    HAIDER, W
    CRITICAL CARE MEDICINE, 1993, 21 (07) : 1035 - 1040
  • [39] Resting end-tidal CO2 association with age, gender, and personality
    Dhokalia, A
    Parsons, DJ
    Anderson, DE
    PSYCHOSOMATIC MEDICINE, 1998, 60 (01): : 33 - 37
  • [40] Effect of Heliox on End-Tidal CO2 Measurement in Healthy Adults
    Waugh, Jonathan B.
    Gardner, Donna D.
    Vines, David L.
    RESPIRATORY CARE, 2013, 58 (11) : 1945 - 1948