ACCURACY OF END-TIDAL CO2 CAPNOMETERS IN POST-CARDIAC SURGERY PATIENTS DURING CONTROLLED MECHANICAL VENTILATION

被引:19
|
作者
Heines, Serge J. H. [1 ]
Strauch, Ulrich [1 ]
Roekaerts, Paul M. H. J. [1 ]
Winkens, Bjorn [2 ]
Bergmans, Dennis C. J. J. [1 ]
机构
[1] Maastricht Univ, Dept Intens Care, Med Ctr, NL-6202 AZ Maastricht, Netherlands
[2] Maastricht Univ, Dept Methodol & Stat, NL-6202 AZ Maastricht, Netherlands
关键词
end-tidal carbon dioxide; capnography; capnometer; monitoring; carbon dioxide; blood gas analysis; CARBON-DIOXIDE; ARTERIAL; TRAUMA; CAPNOGRAPHY; CARE;
D O I
10.1016/j.jemermed.2012.11.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The determination of end-tidal carbon dioxide (etCO(2)) is very helpful in cardiac resuscitation for confirmation and monitoring of endotracheal tube placement and as an indicator of return of circulation and effectiveness of chest compressions. There is now also widespread use of capnometry on-site at emergency and trauma fields. Objective: We studied the accuracy and correlation of three capnometers (EMMA, Medtronic, and Evita) with partial pressure of arterial CO2 (PaCO2) measurements. Methods: The three capnometers were placed in-line in the ventilator tubing of the patient. Forty sedated and mechanically ventilated post-cardiac surgery patients were studied. Twenty consecutive etCO(2) values were collected simultaneously from all three monitors while drawing an arterial blood sample. Paired sample t-test and Pearson correlation were used to compare the capnometers and their correlation with PaCO2. Results: The correlation of etCO(2) measurements between all three capnometers was good (Emma vs. Evita: 0.874, Emma vs. Medtronic: 0.949, Evita vs. Medtronic: 0.878). The correlation of PaCO2 with the Evita is the lowest (0.671) as compared to the EMMA (0.693) and the Medtronic (0.727). The lowest dispersion of the difference between etCO(2) and PaCO2 was seen in EMMA (3.30), the highest in Evita (3.98). Conclusions: A good correlation between etCO(2) and PaCO2 was shown with the three capnometers in the present study. However, etCO(2) measurements were not valid to estimate PaCO2 in these patients. Therefore, capnometry cannot be used to replace serial blood gas analyses completely, but may be a good cardiopulmonary trend monitor and alerting system in catastrophic events. (c) 2013 Elsevier Inc.
引用
收藏
页码:130 / 134
页数:5
相关论文
共 50 条
  • [1] Use of Maximum End-Tidal CO2 Values to Improve End-Tidal CO2 Monitoring Accuracy
    Galia, Fabrice
    Brimioulle, Serge
    Bonnier, Frederic
    Vandenbergen, Nicolas
    Dojat, Michel
    Vincent, Jean-Louis
    Brochard, Laurent J.
    RESPIRATORY CARE, 2011, 56 (03) : 278 - 283
  • [2] An Update on End-Tidal CO2 Monitoring
    Selby, Samuel T.
    Abramo, Thomas
    Hobart-Porter, Nicholas
    PEDIATRIC EMERGENCY CARE, 2018, 34 (12) : 888 - 892
  • [3] Comparison of ventilator-integrated end-tidal CO2 and transcutaneous CO2 monitoring in home-ventilated neuromuscular patients
    Orlikowski, David
    Prigent, Helene
    Ambrosi, Xavier
    Vaugier, Isabelle
    Pottier, Sandra
    Annane, Djillali
    Lofaso, Frederic
    Ogna, Adam
    RESPIRATORY MEDICINE, 2016, 117 : 7 - 13
  • [4] Clinical usefulness of end-tidal CO2 measured using a portable capnometer in patients with respiratory disease
    Suzuki, Manabu
    Fujimoto, Shota
    Sakamoto, Keita
    Tamura, Kentaro
    Ishii, Satoru
    Iikura, Motoyasu
    Izumi, Shinyu
    Takeda, Yuichiro
    Hojo, Masayuki
    Sugiyama, Haruhito
    CLINICAL RESPIRATORY JOURNAL, 2023, 17 (02) : 96 - 104
  • [5] Accuracy of End-tidal CO2 Measurement Through the Nose and Pharynx in Nonintubated Patients During Digital Subtraction Cerebral Angiography
    Zhang, Chunyan
    Wang, Maohua
    Wang, Rurong
    Wang, Wenjian
    JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 2013, 25 (02) : 191 - 196
  • [6] Misleading end-tidal CO2 tensions
    Wahba, RWM
    Tessler, MJ
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1996, 43 (08): : 862 - 866
  • [7] The gradient between arterial and end-tidal carbon dioxide predicts in-hospital mortality in post-cardiac arrest patient
    Kim, Yong Won
    Hwang, Sung Oh
    Kang, Hee Seung
    Cha, Kyoung-Chul
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2019, 37 (01) : 1 - 4
  • [8] End-Tidal CO2 Pressure and Cardiac Performance during Exercise in Heart Failure
    Myers, Jonathan
    Gujja, Pradeep
    Neelagaru, Suresh
    Hsu, Leon
    Vittorio, Timothy
    Jackson-Nelson, Tamika
    Burkhoff, Daniel
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2009, 41 (01) : 18 - 24
  • [9] End-tidal CO2 measurement in emergency management. In nonintubated patients as well?
    Casu, S.
    Schweigkofler, U.
    NOTFALL & RETTUNGSMEDIZIN, 2017, 20 (08): : 668 - 672
  • [10] Implementing Microstream End-Tidal CO2 in the PACU
    Latham, Kimberly
    Bird, Tamera
    Burke, Joyce
    JOURNAL OF PERIANESTHESIA NURSING, 2018, 33 (01) : 23 - 27