Quantification of myocardial uptake rate constants in dynamic small-animal SPECT using a cardiac phantom

被引:4
|
作者
Johnson, Lindsay C. [1 ]
Guerraty, Marie A. [2 ]
Moore, Stephen C. [1 ]
Metzler, Scott D. [1 ]
机构
[1] Univ Penn, Dept Radiol, Perelman Sch Med, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Med, Div Cardiovasc Med, Perelman Sch Med, Philadelphia, PA 19104 USA
基金
美国国家卫生研究院;
关键词
SPECT; cardiac imaging; phantom; kinetic modeling; POSITRON-EMISSION-TOMOGRAPHY; BLOOD-FLOW; PROGNOSTIC VALUE; RESERVE;
D O I
10.1088/1361-6560/ab0472
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Myocardial blood flow and myocardial blood flow reserve (MBFR) measurements are often used clinically to quantify coronary microvascular function. Developing imaging-based methods to measure MBFR for research in mice would be advantageous for evaluating new treatment methods for coronary microvascular disease (CMVD), yet this is more challenging in mice than in humans. This work investigates microSPECT's quantitative capabilities of cardiac imaging by utilizing a multi-part cardiac phantom and applying a known kinetic model to synthesize kinetic data from static data, allowing for assessment of kinetic modeling accuracy. The phantom was designed with four main components: two left-ventricular (LV) myocardial sections and two LV blood-pool sections, sized for end-systole (ES) and end-diastole (ED). Each section of the phantom was imaged separately while acquiring list-mode data. These static, separate-compartment data were manipulated into synthetic dynamic data using a kinetic model representing the myocardium and blood-pool activity concentrations over time and then combined into a set of dynamic image frames and reconstructed. Regions of interest were drawn on the resulting images, and kinetic parameters were estimated. This process was performed for three tracer uptake values (K-1), three myocardial wall thicknesses, ten filter parameters, and 20 iterations for 25 noise ensembles. The degree of filtering and iteration number were optimized to minimize the root mean-squared error (RMSE) of K-1 values, with the largest number of iterations and minimal filtering yielding the lowest error. Using the optimized parameters, K-1 was determined with reasonable error (similar to 3% RMSE) over all wall thicknesses and K-1 input values. This work demonstrates that accurate and precise measurements of K-1 are possible for the U-SPECT+ system used in this study, for several different uptake rates and LV dimensions. Additionally, it allows for future investigation utilizing other imaging systems, including PET studies with any radiotracer, as well as with additional phantom parts containing lesions.
引用
收藏
页数:15
相关论文
共 24 条
  • [11] Comparison of ejection fraction calculation between CT and SPECT at high heart rate: A dynamic cardiac phantom study
    Chien-Cheng Chen
    Thau-Yun Shen
    Christine B. Peterson
    Guang-Uei Hung
    Tinsu Pan
    Journal of Nuclear Cardiology, 2021, 28 : 311 - 316
  • [12] Quantification of myocardial perfusion reserve using dynamic SPECT images of patients with chronic kidney disease
    Tsuda, Noriko
    Shiraishi, Shinya
    Sakamoto, Fumi
    Yuki, Hideaki
    Ogasawara, Koji
    Yoshida, Morikatsu
    Tomiguchi, Seiji
    Tsujita, Kenichi
    Yamashita, Yasuyuki
    JOURNAL OF CARDIOLOGY, 2018, 71 (1-2) : 174 - 180
  • [13] Effect of image quality on myocardial extracellular volume quantification using cardiac computed tomography: a phantom study
    Funama, Yoshinori
    Oda, Seitaro
    Kidoh, Masafumi
    Sakabe, Daisuke
    Nakaura, Takeshi
    ACTA RADIOLOGICA, 2022, 63 (02) : 159 - 165
  • [14] Assessment of infarct size and severity by quantitative myocardial SPECT: Results from a multicenter study using a cardiac phantom
    O'Connor, MK
    Leong, LK
    Gibbons, RJ
    JOURNAL OF NUCLEAR MEDICINE, 2000, 41 (08) : 1383 - 1390
  • [15] Validation of gated blood-pool SPECT cardiac measurements tested using a biventricular dynamic physical phantom
    De Bondt, P
    Nichols, K
    Vandenberghe, S
    Segers, P
    De Winter, O
    Van de Wiele, C
    Verdonck, P
    Shazad, A
    Shoyeb, AH
    De Sutter, J
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (06) : 967 - 972
  • [16] Approach to Assessing Myocardial Perfusion in Rats Using Static [13N]-Ammonia Images and a Small-Animal PET
    Jose Vaquero, Juan
    Gao, Dong-Wei
    Garcia-Villaba, Carmen
    Bacharach, Stephen
    VanBrocklin, Henry
    Fang, Qizhi
    Desco, Manuel
    Lee, Randall
    Dae, Michael
    MOLECULAR IMAGING AND BIOLOGY, 2012, 14 (05) : 541 - 545
  • [17] Assessment of global cardiac I-123 MIBG uptake and washout using volumetric quantification of SPECT acquisitions
    Berlinda J. van der Veen
    Imad Al Younis
    Albert de Roos
    Marcel P. M. Stokkel
    Journal of Nuclear Cardiology, 2012, 19 : 752 - 762
  • [18] Design for a high-resolution small-animal SPECT system using pixellated Si(Li) detectors for In vivo 125I imaging
    Choong, WS
    Moses, WW
    Tindall, CS
    Luke, PN
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2315 - 2319
  • [19] Detection and quantification of acute reperfused myocardial infarction in rabbits using DISA-SPECT/CT and 3.0 T cardiac MRI
    Feng, Yuanbo
    Cona, Marlein Miranda
    Vunckx, Kathleen
    Li, Yue
    Chen, Feng
    Nuyts, Johan
    Gheysens, Olivier
    Zhou, Lin
    Xie, Yi
    Oyen, Raymond
    Ni, Yicheng
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 168 (04) : 4191 - 4198
  • [20] Application of Post Reconstruction Dual Respiratory and Cardiac Motion Compensation for 4D High-Resolution Small Animal Myocardial SPECT
    Lee, Taek-Soo
    Rittenbach, Andrew
    Feng, Tao
    Tsui, Benjamin M. W.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,