Correction of acetabular cup orientation measurements for X-ray beam offset

被引:31
作者
Derbyshire, Brian [1 ]
机构
[1] Wrightington Hosp, Ctr Hip Surg, Appley Bridge WN6 9EP, Wigan, England
关键词
Hip; Prosthesis; Acetabular cup; Orientation measurement; Version measurement; Inclination measurement; Correction; Algorithms;
D O I
10.1016/j.medengphy.2008.02.001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is important to achieve correct orientation of an acetabular cup in order to avoid impingement, dislocation and increased wear. Cup orientation is usually assessed from a pelvis radiograph in which the central X-ray beam has been focused on the pubic symphysis. However, the angular offset of this beam from a beam passing through the centre of the cup causes rotational artefact in the cup image. Three simple algorithms for correcting this 3D rotational artefact are proposed, and these were validated using 3D CAD techniques by simulating the X-ray focus, target point and film position, as well as cup orientation. The accuracy of the correction was assessed in different practical scenarios. Cup version could be corrected to within a few hundredths of a degree, and inclination to within 0.2 degrees, if the X-ray target position was accurately located. However, only a small error in the correction was incurred if this position was erroneously located by as much as 10 mm. The error in correcting the rotation position about the axis of symmetry depended on the cup size and anteversion setting, and it varied sinusoidally according to the position of measurement. Generally, it was within any possible measurement error. For the largest cup size considered (60 mm outer diameter), the error ranged between -0.4 degrees and 0.8 degrees at the highest anteversion setting (30 degrees). (C) 2008 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1119 / 1126
页数:8
相关论文
共 16 条
[1]   ANTEVERSION OF THE ACETABULAR CUP - MEASUREMENT OF ANGLE AFTER TOTAL HIP-REPLACEMENT [J].
ACKLAND, MK ;
BOURNE, WB ;
UHTHOFF, HK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (03) :409-413
[2]  
CLARKE IC, 1976, CLIN ORTHOP RELAT R, P83
[3]  
DERBYSHIRE B, 2003, HIP INT, V13, P52
[4]   Three-dimensional polyethylene wear of a press-fit titanium prosthesis - Factors influencing generation of polyethylene debris [J].
Devane, PA ;
Horne, JG ;
Martin, K ;
Coldham, G ;
Krause, B .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :256-266
[5]   EVALUATION OF ACETABULAR ANTEVERSION FOLLOWING TOTAL HIP ARTHROPLASTY - NECESSITY OF PROPER CENTERING [J].
GOERGEN, TG ;
RESNICK, D .
BRITISH JOURNAL OF RADIOLOGY, 1975, 48 (568) :259-260
[6]   RADIOGRAPHIC CALCULATION OF ANTEVERSION IN ACETABULAR PROSTHESES [J].
HASSAN, DM ;
JOHNSTON, GHF ;
DUST, WNC ;
WATSON, LG ;
CASSIDY, D .
JOURNAL OF ARTHROPLASTY, 1995, 10 (03) :369-372
[7]   Computer assisted measurement of cup placement in total hip replacement [J].
Jaramaz, B ;
DiGioia, AM ;
Blackwell, M ;
Nikou, C .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (354) :70-81
[8]   DISLOCATIONS AFTER TOTAL HIP-REPLACEMENT ARTHROPLASTIES [J].
LEWINNEK, GE ;
LEWIS, JL ;
TARR, R ;
COMPERE, CL ;
ZIMMERMAN, JR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1978, 60 (02) :217-220
[9]   THE DEFINITION AND MEASUREMENT OF ACETABULAR ORIENTATION [J].
MURRAY, DW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02) :228-232
[10]   Planar anteversion of the acetabular cup as determined from plain anteroposterior radiographs [J].
Pradhan, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (03) :431-435