Inner-volume imaging in vivo using three-dimensional parallel spatially selective excitation

被引:38
作者
Schneider, Johannes T. [1 ,2 ]
Kalayciyan, Raffi [1 ]
Haas, Martin [2 ]
Herrmann, Sarah R. [1 ]
Ruhm, Wolfgang [1 ]
Hennig, Juergen [2 ]
Ullmann, Peter [1 ]
机构
[1] Bruker BioSpin MRI GmbH, DE-76275 Ettlingen, Germany
[2] Univ Med Ctr Freiburg, Dept Radiol, Freiburg, Germany
关键词
inner-volume imaging; parallel excitation; multidimensional spatially selective excitation pulses; three-dimensional k-space trajectory; K-SPACE; RF PULSES; MR-ANGIOGRAPHY; TRANSMISSION; TRAJECTORIES; DESIGN; LOCALIZATION; SPECTROSCOPY; CHANNELS; ECHO;
D O I
10.1002/mrm.24381
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work describes the first experimental realization of three-dimensional spatially selective excitation using parallel transmission in vivo. For the design of three-dimensional parallel excitation pulses with short durations and high excitation accuracy, the choice of a suitable transmit k-space trajectory is crucial. For this reason, the characteristics of a stack-of-spirals trajectory and of a concentric-shells trajectory were examined in an initial simulation study. It showed that, especially when undersampling the trajectories in combination with parallel transmission, experimental parameters such as transmit-coil geometry and off-resonance conditions have an essential impact on the suitability of the selected trajectory and undersampling scheme. Both trajectories were applied in MR inner-volume imaging experiments which demonstrate that acceptably short and robust three-dimensional selective pulses can be achieved if the trajectory is temporally optimized and its actual path is measured and considered during pulse calculation. Pulse durations as short as 3.2 ms were realized and such pulses were appropriate to accurately excite arbitrarily shaped volumes in a corn cob and in a rat in vivo. Reduced field-of-view imaging of these selectively excited targets allowed high spatial resolution and significantly reduced measurement times and furthermore demonstrates the feasibility of three-dimensional parallel excitation in realistic MRI applications in vivo. Magn Reson Med, 2013. (c) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:1367 / 1378
页数:12
相关论文
共 46 条
  • [1] k-space undersampling in PROPELLER imaging
    Arfanakis, K
    Tamhane, AA
    Pipe, JG
    Anastasio, MA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (03) : 675 - 683
  • [2] Multispectral imaging with three-dimensional rosette trajectories
    Bucholz, Elizabeth K.
    Song, Jiayu
    Johnson, Allan
    Hancu, Ileana
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (03) : 581 - 589
  • [3] Local SAR Reduction in Parallel Excitation Based on Channel-Dependent Tikhonov Parameters
    Cloos, Martijn Anton
    Luong, Michel
    Ferrand, Guillaume
    Amadon, Alexis
    Le Bihan, Denis
    Boulant, Nicolas
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 32 (05) : 1209 - 1216
  • [4] Accelerated Multidimensional Radiofrequency Pulse Design for Parallel Transmission Using Concurrent Computation on Multiple Graphics Processing Units
    Deng, Weiran
    Yang, Cungeng
    Stenger, V. Andrew
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (02) : 363 - 369
  • [5] Simple correction method for k-space trajectory deviations in MRI
    Duyn, JH
    Yang, YH
    Frank, JA
    van der Veen, JW
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1998, 132 (01) : 150 - 153
  • [6] INNER VOLUME MR IMAGING - TECHNICAL CONCEPTS AND THEIR APPLICATION
    FEINBERG, DA
    HOENNINGER, JC
    CROOKS, LE
    KAUFMAN, L
    WATTS, JC
    ARAKAWA, M
    [J]. RADIOLOGY, 1985, 156 (03) : 743 - 747
  • [7] Single-shot MR imaging using trapezoidal-gradient-based Lissajous trajectories
    Feng, HH
    Gu, H
    Silbersweig, D
    Stern, E
    Yang, YH
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (08) : 925 - 932
  • [8] Graesslin I., 2007, Proc. 15th Ann. Meet. Intl. Soc. Mag. Reson. Med, P1688
  • [9] Spatial domain method for the design of RF pulses in multicoil parallel excitation
    Grissom, William
    Yip, Chun-yu
    Zhang, Zhenghui
    Stenger, V. Andrew
    Fessler, Jeffrey A.
    Noll, Douglas C.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (03) : 620 - 629
  • [10] Haas M, 2011, P 19 ANN M ISMRM MON, P4433