Lack of interaction between magnetic resonance imaging and the Copper-T380A IUD

被引:10
作者
Pasquale, SA
Russer, TJ
Foldesy, R
Mezrich, RS
机构
[1] UNIV MED & DENT NEW JERSEY,ROBERT WOOD JOHNSON MED SCH,DEPT RADIOL,NEW BRUNSWICK,NJ 08901
[2] WYETH AYERST LABS,PHILADELPHIA,PA
[3] HOSP UNIV PENN,DEPT RADIOL,PHILADELPHIA,PA 19104
关键词
copper-T380A IUD; ParaGard; MRI;
D O I
10.1016/S0010-7824(97)00019-X
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
To determine if the CuT380A (ParaGard(R)) IUD is affected by magnetic resonance imaging (MRI), in vitro studies utilizing a CuT380A IUD (ParaGard(R)), a copper-bearing IUD, and a Signa 1.5T HR system were used to evaluate whether the dynamic magnetic forces generated by the MRI resulted in movement, torque, or heat when the IUD was exposed to the magnetic field generated by the MRI. There was no deflection, turning motion (torque), or temperature change when the IUD was exposed to a magnetic field. There appears to be no reason to exclude women with IUDs of the type examined from an MRI system or its environs. (C) 1997 Elsevier Science Inc.
引用
收藏
页码:169 / 173
页数:5
相关论文
共 8 条
  • [1] OBSERVATION BY MR IMAGING OF INVIVO TEMPERATURE-CHANGES INDUCED BY RADIO-FREQUENCY HYPERTHERMIA
    HALL, AS
    PRIOR, MV
    HAND, JW
    YOUNG, IR
    DICKINSON, RJ
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1990, 14 (03) : 430 - 436
  • [2] KANAL E, 1990, SOC MAGN RES, V1, P274
  • [3] PLACEMENT OF A FERROMAGNETIC INTRACEREBRAL ANEURYSM CLIP IN A MAGNETIC-FIELD WITH A FATAL OUTCOME
    KLUCZNIK, RP
    CARRIER, DA
    PYKA, R
    HAID, RW
    [J]. RADIOLOGY, 1993, 187 (03) : 855 - 856
  • [4] MARK AS, 1987, RADIOLOGY, P311
  • [5] MITCHELL M D, 1986, Magnetic Resonance Imaging, V4, P263, DOI 10.1016/0730-725X(86)91068-4
  • [6] NUCLEAR MAGNETIC-RESONANCE IMAGING AND ITS SAFETY IMPLICATIONS - FOLLOW-UP OF 181 PATIENTS
    REID, A
    SMITH, FW
    HUTCHISON, JMS
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1982, 55 (658) : 784 - 786
  • [7] MR PROCEDURES AND BIOMEDICAL IMPLANTS, MATERIALS, AND DEVICES - 1993 UPDATE
    SHELLOCK, FG
    MORISOLI, S
    KANAL, E
    [J]. RADIOLOGY, 1993, 189 (02) : 587 - 599
  • [8] PARAGARD T380A PRESC