Fatigue and cyclic deformation behaviour of surface-modified titanium alloys in simulated physiological media

被引:105
作者
Leinenbach, C
Eifler, D
机构
[1] Univ Kaiserslautern, Inst Mat Sci & Engn, D-67663 Kaiserslautern, Germany
[2] EMPA, Swiss Fed Labs Mat Testing & Res, Dubendorf, Switzerland
关键词
titanium alloys; surface modification; titanium oxide; surface roughness; corrosion fatigue;
D O I
10.1016/j.biomaterials.2005.08.012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this investigation, the cyclic deformation behaviour of the binary titanium alloys Ti-6Al-4V and Ti-6Al-7Nb was characterized in axial stress-controlled constant amplitude and load increase tests as well as in rotating bending tests. The influence of different clinically relevant surface treatments (polishing, corundum grit blasting, thermal and anodic oxidizing) on the fatigue behaviour was investigated. All tests were realized in oxygen-saturated Ringer's solution. The cyclic deformation behaviour was characterized by mechanical hysteresis measurements. In addition, the change of the free corrosion potential and the corrosion current during testing in simulated physiological media indicated surface damages such as slip bands, intrusions and extrusions or finally microcracks. Microstructural changes on the specimen surfaces were examined by scanning electron microscopy (SEM). (c) 2005 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1200 / 1208
页数:9
相关论文
共 30 条
[1]   The effect of sandblasting treatment on endurance properties of titanium alloy hip prostheses [J].
Baleani, M ;
Viceconti, M ;
Toni, A .
ARTIFICIAL ORGANS, 2000, 24 (04) :296-299
[2]   CRYSTAL-STRUCTURE OF THE SURFACE OXIDE LAYER ON TITANIUM AND ITS CHANGES ARISING FROM IMMERSION [J].
EFFAH, EAB ;
BIANCO, PD ;
DUCHEYNE, P .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (01) :73-80
[3]   Anodic plasma-chemical treatment of CP titanium surfaces for biomedical applications [J].
Frauchiger, VM ;
Schlottig, F ;
Gasser, B ;
Textor, M .
BIOMATERIALS, 2004, 25 (04) :593-606
[4]   Effect of thermal oxidation on corrosion and corrosion-wear behaviour of a Ti-6A1-4V alloy [J].
Güleryüz, H ;
Çimenoglu, H .
BIOMATERIALS, 2004, 25 (16) :3325-3333
[5]  
HANAGARTH H, 1988, P ICSP 3 OB DGM, P477
[6]  
Hickling J, 1998, MATER CORROS, V49, P651, DOI 10.1002/(SICI)1521-4176(199809)49:9<651::AID-MACO651>3.0.CO
[7]  
2-M
[8]  
HIRSCH T, 1985, P 5 INT C TIT OB DGM, V3, P2163
[9]   CORROSION MONITORING-SYSTEM BASED ON MEASUREMENT AND ANALYSIS OF ELECTROCHEMICAL NOISE [J].
LEGAT, A ;
DOLECEK, V .
CORROSION, 1995, 51 (04) :295-300
[10]   Characterization of the cyclic deformation behaviour and fatigue crack initiation on titanium in physiological media by electrochemical techniques [J].
Leinenbach, C ;
Eifler, D .
ZEITSCHRIFT FUR METALLKUNDE, 2004, 95 (06) :535-541