Celect Inferior Vena Cava Wall Strut Perforation Begets Additional Strut Perforation

被引:25
|
作者
Dowell, Joshua D. [1 ,2 ,3 ]
Castle, Jordan C. [1 ,2 ]
Schickel, Maureen [3 ]
Andersson, Urbina K. [1 ,2 ]
Zielinski, Rachel [3 ]
McLoney, Eric [1 ,2 ]
Guy, Gregory [1 ,2 ]
Yang, Xiangyu [1 ,2 ]
Ghadiali, Samir [3 ]
机构
[1] Ohio State Univ, Div Intervent Radiol, Columbus, OH 43210 USA
[2] Ohio State Univ, Dept Radiol, Columbus, OH 43210 USA
[3] Ohio State Univ, Dept Biomed Engn, Wexner Med Ctr, Columbus, OH 43210 USA
关键词
GUNTHER-TULIP; FILTER; PENETRATION; EXPERIENCE; RETRIEVAL; COMPLICATIONS;
D O I
10.1016/j.jvir.2015.06.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To identify risk factors for strut perforation following Celect inferior vena cava (IVC) filter (IVCF) placement and to use finite element modeling to predict the mechanical impact of long-dwelling filters. Materials and Methods: Ninety-one patients With three computed tomography (CT) studies were evaluated following Celect IVCF placement (2007-2013). Three-dimensional finite element models of the Celect IVCF were developed to simulate mechanical deformation of the IVCF encountered in vivo. Simulated forces applied by the primary struts on the IVC wall were measured as a function of luminal area and tilt angle. Results: Although 33 patients (36%) showed primary strut perforation on initial follow-up CT, 60 patients (66%) showed progressive perforation over time (P < .0001), with 72 patients (79%) showing primary strut perforation on the final CT (average, 554 d). Female patients (P = .004) and those with malignancy history (P = .01) had significantly higher perforation rates at a given time. Caval area also decreased after primary filter strut perforation, and we therefore proposed that this was the mechanism for progressive perforation. Consistent with this mechanism, three-dimensional finite element modeling demonstrated increasing strut force with decreasing IVC diameter. Conclusions: Celect IVCF primary strut perforation is progressive over time and is more common in female patients and those with a history of malignancy. In addition, this progressive perforation may be predicted by three-dimensional finite element modeling. These patient populations may require closer follow-up after IVCF placement to prevent or reduce the risk for filter complication or worsening perforation.
引用
收藏
页码:1510 / 1518
页数:9
相关论文
共 50 条
  • [11] Differences in radial expansion force among inferior vena cava filter models support documented perforation rates
    Robins, J. Eli
    Ragai, Ihab
    Yamaguchi, Dean J.
    JOURNAL OF VASCULAR SURGERY-VENOUS AND LYMPHATIC DISORDERS, 2018, 6 (03) : 368 - 371
  • [12] A systematic review of symptomatic duodenal perforation by inferior vena cava filters
    Malgor, Rafael D.
    Labropoulos, Nicos
    JOURNAL OF VASCULAR SURGERY, 2012, 55 (03) : 856 - U552
  • [13] Perforation of the Heart During Elective Removal of an Inferior Vena Cava Filter
    Desai, Sapna P.
    Yu, Christine Z.
    Evangelista, Frank
    Gill, James R.
    AMERICAN JOURNAL OF FORENSIC MEDICINE AND PATHOLOGY, 2024, 45 (04) : 347 - 351
  • [14] Safety and Effectiveness of the Celect Inferior Vena Cava Filter: Preliminary Results
    Sangwaiya, Minal Jagtiani
    Marentis, Theodore C.
    Walker, T. Gregory
    Stecker, Michael
    Wicky, Stephan T.
    Kalva, Sanjeeva P.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 20 (09) : 1188 - 1192
  • [15] Filter-Associated Inferior Vena Cava Thrombosis with Duodenal Perforation: Case Report and Literature Review
    Chassin-Trubert, Lucien
    Prouse, Giorgio
    Ozdemir, Baris Ata
    Lounes, Youcef
    Alonso, William
    Clapies, Myriam
    Alric, Pierre
    Canaud, Ludovic
    ANNALS OF VASCULAR SURGERY, 2019, 58 : 383.e1 - 383.e6
  • [16] Upper Gastrointestinal Bleeding Secondary to Duodenal Wall Perforation by Inferior Vena Cava Filter: A Rare Clinical Presentation
    Abosheaishaa, Hazem
    Saha, Utsow
    Abdelhalim, Omar
    Al-Howthi, Nuha
    Elhawary, Ahmed
    Abusuliman, Mohammed
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (11)
  • [17] Retrospective Review of 120 Celect Inferior Vena Cava Filter Retrievals: Experience at a Single Institution
    Zhou, Dayong
    Spain, James
    Moon, Eunice
    Mclennan, Gordon
    Sands, Mark J.
    Wang, Weiping
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 23 (12) : 1557 - 1563
  • [18] Long-term computed tomography follow-up results strut penetration of inferior vena cava filters
    Yang, Shin Seok
    Yun, Woo-Sung
    JOURNAL OF VASCULAR SURGERY-VENOUS AND LYMPHATIC DISORDERS, 2019, 7 (05) : 646 - 652
  • [19] Vena Cava Filter Retrieval with Aorto-Iliac Arterial Strut Penetration
    Holly, Brian P.
    Gaba, Ron C.
    Lessne, Mark L.
    Lewandowski, Robert J.
    Ryu, Robert K.
    Desai, Kush R.
    Sing, Ronald F.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2018, 41 (08) : 1184 - 1188
  • [20] Bidirectional Pull-Back Technique for Retrieval of Strut-Embedded Cylinder-Shaped Filters in Inferior Vena Cava
    Du, Xiaolong
    Li, Chenglong
    Qian, Aimin
    Zhang, Yeqing
    Li, Wendong
    Yu, Huiying
    Li, Xiaoqiang
    Kong, Lingshang