Celect Inferior Vena Cava Wall Strut Perforation Begets Additional Strut Perforation

被引:25
|
作者
Dowell, Joshua D. [1 ,2 ,3 ]
Castle, Jordan C. [1 ,2 ]
Schickel, Maureen [3 ]
Andersson, Urbina K. [1 ,2 ]
Zielinski, Rachel [3 ]
McLoney, Eric [1 ,2 ]
Guy, Gregory [1 ,2 ]
Yang, Xiangyu [1 ,2 ]
Ghadiali, Samir [3 ]
机构
[1] Ohio State Univ, Div Intervent Radiol, Columbus, OH 43210 USA
[2] Ohio State Univ, Dept Radiol, Columbus, OH 43210 USA
[3] Ohio State Univ, Dept Biomed Engn, Wexner Med Ctr, Columbus, OH 43210 USA
关键词
GUNTHER-TULIP; FILTER; PENETRATION; EXPERIENCE; RETRIEVAL; COMPLICATIONS;
D O I
10.1016/j.jvir.2015.06.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To identify risk factors for strut perforation following Celect inferior vena cava (IVC) filter (IVCF) placement and to use finite element modeling to predict the mechanical impact of long-dwelling filters. Materials and Methods: Ninety-one patients With three computed tomography (CT) studies were evaluated following Celect IVCF placement (2007-2013). Three-dimensional finite element models of the Celect IVCF were developed to simulate mechanical deformation of the IVCF encountered in vivo. Simulated forces applied by the primary struts on the IVC wall were measured as a function of luminal area and tilt angle. Results: Although 33 patients (36%) showed primary strut perforation on initial follow-up CT, 60 patients (66%) showed progressive perforation over time (P < .0001), with 72 patients (79%) showing primary strut perforation on the final CT (average, 554 d). Female patients (P = .004) and those with malignancy history (P = .01) had significantly higher perforation rates at a given time. Caval area also decreased after primary filter strut perforation, and we therefore proposed that this was the mechanism for progressive perforation. Consistent with this mechanism, three-dimensional finite element modeling demonstrated increasing strut force with decreasing IVC diameter. Conclusions: Celect IVCF primary strut perforation is progressive over time and is more common in female patients and those with a history of malignancy. In addition, this progressive perforation may be predicted by three-dimensional finite element modeling. These patient populations may require closer follow-up after IVCF placement to prevent or reduce the risk for filter complication or worsening perforation.
引用
收藏
页码:1510 / 1518
页数:9
相关论文
共 50 条
  • [1] Perforation of inferior vena cava and duodenum by strut of inferior vena cava filter A case report
    Park, Hyun Oh
    Choi, Jun Young
    Jang, In Seok
    Kim, Jong Duk
    Kim, Jong Woo
    Byun, Joung Hun
    Kim, Sung Hwan
    Yang, Jun Ho
    Moon, Seong Ho
    Kim, Ki Nyun
    Kang, Dong Hun
    Jung, Jae Jun
    Choi, See Min
    Kim, Ji Yoon
    Lee, Chung Eun
    MEDICINE, 2019, 98 (47)
  • [2] Reporting the impact of inferior vena cava perforation by filters
    Wood, Emily A.
    Malgor, Rafael D.
    Gasparis, Antonios P.
    Labropoulos, Nicos
    PHLEBOLOGY, 2014, 29 (07) : 471 - 475
  • [3] Symptomatic duodenal perforation by inferior vena cava filter
    Baptista Sincos, Anna P. W.
    Sincos, Igor R.
    Labropoulos, Nicos
    Donega, Bruno C.
    Klepacz, Andrea
    Aun, Ricardo
    PHLEBOLOGY, 2018, 33 (08) : 523 - 533
  • [4] Inferior vena cava perforation during percutaneous filter removal
    Hicks, Adam C.
    Sangroula, Daisy
    Dwivedi, Amit J.
    Wayne, Erik J.
    Sigdel, Abindra
    VASCULAR, 2024, 32 (01) : 190 - 194
  • [5] Strut Penetration: Local Complications, Breakthrough Pulmonary Embolism, and Retrieval Failure in Patients with Celect Vena Cava Filters
    Bos, Aaron
    Thuong Van Ha
    van Beek, Darren
    Ginsburg, Michael
    Zangan, Steven
    Navuluri, Rakesh
    Lorenz, Jonathan
    Funaki, Brian
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2015, 26 (01) : 101 - 106
  • [6] Symptomatic caval penetration by a Celect inferior vena cava filter
    Conor O. Bogue
    Philip R. John
    Bairbre L. Connolly
    David J. Rea
    Joao G. Amaral
    Pediatric Radiology, 2009, 39 : 1110 - 1113
  • [7] Caval Perforation from a Malpositioned Inferior Vena Cava Filter
    Holleck, Jurgen L.
    Chen, Eric Y.
    Bonomo, Jason
    JOURNAL OF GENERAL INTERNAL MEDICINE, 2019, 34 (07) : 1358 - 1359
  • [8] Caval Perforation from a Malpositioned Inferior Vena Cava Filter
    Jürgen L. Holleck
    Eric Y. Chen
    Jason Bonomo
    Journal of General Internal Medicine, 2019, 34 : 1358 - 1359
  • [9] Symptomatic caval penetration by a Celect inferior vena cava filter
    Bogue, Conor O.
    John, Philip R.
    Connolly, Bairbre L.
    Rea, David J.
    Amaral, Joao G.
    PEDIATRIC RADIOLOGY, 2009, 39 (10) : 1110 - 1113
  • [10] Penetration of Celect Inferior Vena Cava Filters: Retrospective Review of CT Scans in 265 Patients
    Zhou, Dayong
    Moon, Eunice
    Bullen, Jennifer
    Sands, Mark
    Levitin, Abraham
    Wang, Weiping
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (03) : 643 - 647