Modular junction may be more problematic than bearing wear in metal-on-metal total hip arthroplasty

被引:14
|
作者
Vendittoli, Pascal-Andre [1 ]
Masse, Vincent [1 ]
Kiss, Marc-Olivier [1 ]
Lusignan, Daniel [1 ]
Lavigne, Martin [1 ]
机构
[1] Univ Montreal, Dept Surg, Hosp Maisonneuve Rosemont, Montreal, PQ, Canada
关键词
Adverse reaction to metal debris; metal ions; metal-on-metal joint prostheses; modular stem; taper junction; total hip arthroplasty; CHROMIUM SERUM-LEVELS; ION LEVELS; NECKS ANALYSIS; CERAMIC THR; LARGE-HEAD; CORROSION; COBALT; REPLACEMENT; INTERFACE; RELEASE;
D O I
10.1177/1120700018808696
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: In total hip arthroplasty (THA), local adverse reaction to metal debris (ARMD) may be caused by abnormal metal ion release from a metal-on-metal (MoM) bearing, or by wear and corrosion of the implant's modular junction. The aim of this study was to compare ion levels and rate of ARMD between patients sharing the same MoM bearing but 1 group having monoblock stems versus another having modular stems. Materials and methods: Whole blood cobalt (Co) and chromium (Cr) ion concentrations, ARMD rate, revision rate, and function measured by UCLA and WOMAC scores were compared between groups. Results: ARMD rate was significantly higher in the modular group (46%) compared with the monoblock group (16%, p = 0.031). Revision for ARMD was performed at 52.8 +/- 8.1 months in the modular group versus 98.2 +/- 15.5 months after primary THA in the monoblock group. ARMD originated from wear and corrosion of the junction between stem and femoral head adapter sleeve in all monoblock cases, and the junction between stem and modular neck in all the modular ones. Cr and Co ions levels were significantly higher in the modular stem group (p < 0.001 for both). Conclusions: Although both groups had MoM bearings, corrosion at stem/neck or neck/head junctions combining dissimilar metal (Ti and Cr-Co) was seen as the source of excess metal ions release leading to ARMD. Poor performance of the modular junction may be more deleterious than wear of the bearing. To avoid such complications, THA femoral stem modular junctions should be eliminated (return to a full monoblock implant) or have improved junction design.
引用
收藏
页码:262 / 269
页数:8
相关论文
共 50 条
  • [21] Metal on metal bearing in total hip arthroplasty
    Alghamdi, A.
    Lavigne, M.
    Vendittoli, P. -A.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2012, 63 (02) : 85 - 99
  • [22] Adverse effects after metal-on-metal bearings in total hip arthroplasty
    Friesenbichler, J.
    Maurer-Ertl, W.
    Pirker-Fruehauf, U.
    Wolf, M.
    Leithner, A.
    Weger, C.
    Pechmann, M.
    Sadoghi, P.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2012, 63 (05) : 379 - 385
  • [23] Five- to Ten-Year Outcomes for Modular Metal-On-Metal Total Hip Arthroplasty
    Bernasek, Thomas L.
    Polikandriotis, John A.
    Levering, Melissa F.
    Dalury, David F.
    Fisher, David A.
    Adler, Michael J.
    JOURNAL OF ARTHROPLASTY, 2013, 28 (07): : 1231 - 1234
  • [24] Long-Term Outcome of Metal-on-Metal Total Hip Arthroplasty with Modular Neck Stem
    Wakabayashi, Hiroki
    Hasegawa, Masahiro
    Naito, Yohei
    Tone, Shine
    Sudo, Akihiro
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (06)
  • [25] Serum metal ion levels after second-generation metal-on-metal total hip arthroplasty
    Imanishi, Takao
    Hasegawa, Masahiro
    Sudo, Akihiro
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2010, 130 (12) : 1447 - 1450
  • [26] Revision in Metal-on-Metal Total Hip Arthroplasty: Current Knowledge
    Stiehler, Maik
    Guenther, Klaus-Peter
    EUROPEAN INSTRUCTIONAL LECTURES, VOL 15, 2015, 15 : 143 - 151
  • [27] Metal-on-Metal Total Hip Arthroplasty: Patient Evaluation and Treatment
    Bolognesi, Michael P.
    Ledford, Cameron K.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2015, 23 (12) : 724 - 731
  • [28] The Relationship Between Cobalt/Chromium Ratios and the High Prevalence of Head-Stem Junction Corrosion in Metal-on-Metal Total Hip Arthroplasty
    Hothi, Harry S.
    Berber, Reshid
    Whittaker, Robert K.
    Blunn, Gordon W.
    Skinner, John A.
    Hart, Alister J.
    JOURNAL OF ARTHROPLASTY, 2016, 31 (05): : 1123 - 1127
  • [29] Large-Diameter Modular Metal-on-Metal Total Hip Arthroplasty Incidence of Revision for Adverse Reaction to Metallic Debris
    Barrett, William P.
    Kindsfater, Kirk A.
    Lesko, James P.
    JOURNAL OF ARTHROPLASTY, 2012, 27 (06): : 976 - 983
  • [30] Medium-term results following large diameter metal-on-metal total hip arthroplasty: increasing failure after 6 years
    Dhotare, Satish V.
    Shivarathre, Deepak G.
    Croitoru, Cristina
    Armstrong, Catherine
    Kapoor, Birender
    Peter, Viju K.
    HIP INTERNATIONAL, 2016, 26 (03) : 226 - 232