Assessment of scatter radiation dose and absorbed doses in eye lens and thyroid gland during digital breast tomosynthesis

被引:14
作者
Chusin, Thunyarat [1 ,2 ]
Matsubara, Kosuke [3 ]
Takemura, Akihiro [3 ]
Okubo, Rena [1 ]
Ogawa, Yoshinori [1 ]
机构
[1] Kanazawa Univ, Grad Sch Med Sci, Div Hlth Sci, Dept Quantum Med Technol, Kanazawa, Ishikawa, Japan
[2] Naresuan Univ, Fac Allied Hlth Sci, Dept Radiol Technol, Phitsanulok, Thailand
[3] Kanazawa Univ, Fac Hlth Sci, Inst Med, Dept Quantum Med Technol, Kanazawa, Ishikawa, Japan
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2019年 / 20卷 / 01期
关键词
digital breast tomosynthesis; eye lens dose; mammography; optically stimulated luminescence dosimeter; scatter radiation; MONTE-CARLO; MAMMOGRAPHY; DOSIMETRY;
D O I
10.1002/acm2.12486
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Digital breast tomosynthesis (DBT) is an alternative tool for breast cancer screening; however, the magnitude of peripheral organs dose is not well known. This study aimed to measure scattered dose and estimate organ dose during mammography under conventional (CM) and Tomo (TM) modes in a specific DBT system. Optically stimulated luminescence dosimeters (OSLDs), whose responses were corrected using a parallel-plate ionization chamber, were pasted on the surface of custom-made polymethyl methacrylate (PMMA) and RANDO phantoms to measure entrance surface air kerma (ESAK). ESAK measurements were also acquired with a 4.5-cm thick breast phantom for a standard mammogram. Organ dose conversion factors (CFD) were determined as ratio of air kerma at a specific depth to that at the surface for the PMMA phantom and multiplied by the ratio of mass energy absorption coefficients of tissue to air. Normalized eye lens and thyroid gland doses were calculated using the RANDO phantom by multiplying CFD and ESAK values. Maximum variability in OSLD response to scatter radiation from the DBT system was 33% in the W/Rh spectrum and variations in scattered dose distribution were observed between CM and TM. The CFD values for eye lens and thyroid gland ranged between 0.58 to 0.66 and 0.29 to 0.33, respectively. Mean organ doses for two-view unilateral imaging were 0.24 (CM) and 0.18 (TM) mu Gy/mAs for the eye lens and 0.24 (CM) and 0.25 (TM) mu Gy/mAs for the thyroid gland. Higher organ doses were observed during TM compared to CM as the automatic exposure control (AEC) system resulted in greater total mAs values in TM.
引用
收藏
页码:340 / 347
页数:8
相关论文
共 18 条
  • [1] [Anonymous], ANN ICRP
  • [2] [Anonymous], 2000089 JAERI
  • [3] Breast Tomosynthesis: State-of-the-Art and Review of the Literature
    Baker, Jay A.
    Lo, Joseph Y.
    [J]. ACADEMIC RADIOLOGY, 2011, 18 (10) : 1298 - 1310
  • [4] Dosimetric characterization and organ dose assessment in digital breast tomosynthesis: Measurements and Monte Carlo simulations using voxel phantoms
    Baptista, Mariana
    Di Maria, Salvatore
    Barros, Silvia
    Figueira, Catarina
    Sarmento, Marta
    Orvalho, Lurdes
    Vaz, Pedro
    [J]. MEDICAL PHYSICS, 2015, 42 (07) : 3788 - 3800
  • [5] Average glandular dose in digital mammography and digital breast tomosynthesis: comparison of phantom and patient data
    Bouwman, R. W.
    van Engen, R. E.
    Young, K. C.
    den Heeten, G. J.
    Broeders, M. J. M.
    Schopphoven, S.
    Jeukens, C. R. L. P. N.
    Veldkamp, W. J. H.
    Dance, D. R.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (20) : 7893 - 7907
  • [6] Impact on the recall rate of digital breast tomosynthesis as an adjunct to digital mammography in the screening setting. A double reading experience and review of the literature
    Carbonaro, Luca A.
    Di Leo, Giovanni
    Clauser, Paola
    Trimboli, Rubina M.
    Verardi, Nicola
    Fedeli, Maria P.
    Girometti, Rossano
    Tafa, Alfredo
    Bruscoli, Paola
    Saguatti, Gianni
    Bazzocchi, Massimo
    Sardanelli, Francesco
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (04) : 808 - 814
  • [7] CT dosimetry: Getting the best from the adult cristy phantom
    Castellano, IA
    Dance, DR
    Evans, PM
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 321 - 325
  • [8] Scatter Radiation Dose From Digital Screening Mammography Measured in a Representative Patient Population
    Mautz, Alan
    Budovec, Joseph J.
    Mulligan, Margaret
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 206 (02) : 365 - 365
  • [9] Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study
    Ciatto, Stefano
    Houssami, Nehmat
    Bernardi, Daniela
    Caumo, Francesca
    Pellegrini, Marco
    Brunelli, Silvia
    Tuttobene, Paola
    Bricolo, Paola
    Fanto, Carmine
    Valentini, Marvi
    Montemezzi, Stefania
    Macaskill, Petra
    [J]. LANCET ONCOLOGY, 2013, 14 (07) : 583 - 589
  • [10] Overview of the evidence on digital breast tomosynthesis in breast cancer detection
    Houssami, Nehmat
    Skaane, Per
    [J]. BREAST, 2013, 22 (02) : 101 - 108