CdTe and CdZnTe gamma ray detectors for medical and industrial imaging systems

被引:116
作者
Eisen, Y [1 ]
Shor, A [1 ]
Mardor, I [1 ]
机构
[1] Soreq NRC, IL-81800 Yavne, Israel
关键词
CdTe; CdZnTe detectors; imaging; nuclear camera;
D O I
10.1016/S0168-9002(99)00003-0
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
CdTe and CdZnTe X-ray and gamma ray detectors in the form of single elements or as segmented monolithic detectors have been shown to be useful in medical and industrial imaging systems. These detectors possess inherently better energy resolution than scintillators coupled to either photodiodes or photomultipliers, and together with application specific integrated circuits they lead to compact imaging systems of enhanced spatial resolution and better contrast resolution. Photopeak efficiencies of these detectors is greatly affected by a relatively low hole mobility-lifetime product. Utilizing these detectors as highly efficient good spectrometers, demands use of techniques to improve their charge collection properties, i.e., correct for variations in charge losses at different depths of interaction in the detector. The corrections for the large hole trapping are made either by applying electronic techniques or by fabricating detector or electrical contacts configurations which differ from the commonly used planar detectors. The following review paper is divided into three parts: The first part discusses detector contact configurations for enhancing photopeak efficiencies and the single carrier collection approach which leads to improved energy resolutions and photopeak efficiencies at high gamma ray energies. The second part demonstrates excellent spectroscopic results using thick CdZnTe segmented monolithic pad and strip detectors showing energy resolutions less than 2% FWHM at 356 keV gamma rays. The third part discusses advantages and disadvantages of CdTe and CdZnTe detectors in imaging systems and describes new developments for medical diagnostics imaging systems. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:158 / 170
页数:13
相关论文
共 28 条
[1]  
Anger H O., 1967, Instrumentation in Nuclear Medicine
[2]   SEMICONDUCTOR ARRAYS WITH MULTIPLEXER READOUT FOR GAMMA-RAY IMAGING - RESULTS FOR A 48X48 GE ARRAY [J].
BARBER, HB ;
AUGUSTINE, FL ;
BARRETT, HH ;
DERENIAK, EL ;
MATHERSON, KL ;
MEYERS, TJ ;
PERRY, DL ;
VENZON, JE ;
WOOLFENDEN, JM ;
YOUNG, ET .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 353 (1-3) :361-365
[3]   CHARGE-TRANSPORT IN ARRAYS OF SEMICONDUCTOR GAMMA-RAY DETECTORS [J].
BARRETT, HH ;
ESKIN, JD ;
BARBER, HB .
PHYSICAL REVIEW LETTERS, 1995, 75 (01) :156-159
[4]   Timing properties of CdZnTe detectors for positron emission tomography [J].
Bertolucci, E ;
Conti, M ;
Curto, CA ;
Russo, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 400 (01) :107-112
[5]   CHARGE-CARRIER MOBILITIES IN CD0.8ZN0.2TE SINGLE-CRYSTALS USED AS NUCLEAR RADIATION DETECTORS [J].
BURSHTEIN, Z ;
JAYATIRTHA, HN ;
BURGER, A ;
BUTLER, JF ;
APOTOVSKY, B ;
DOTY, FP .
APPLIED PHYSICS LETTERS, 1993, 63 (01) :102-104
[6]   PIXELLATED CDZNTE DETECTOR ARRAYS [J].
DOTY, FP ;
BARBER, HB ;
AUGUSTINE, FL ;
BUTLER, JF ;
APOTOVSKY, BA ;
YOUNG, ET ;
HAMILTON, W .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 353 (1-3) :356-360
[7]  
EISEN Y, 1993, MATER RES SOC SYMP P, V302, P487, DOI 10.1557/PROC-302-487
[8]   CORRECTION OF INCOMPLETE CHARGE COLLECTION IN CDTE DETECTORS [J].
EISEN, Y ;
HOROVITZ, Y .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 353 (1-3) :60-66
[9]   A gamma camera based on CdTe detectors [J].
Eisen, Y ;
Shor, A ;
Gilath, C ;
Tsabarim, M ;
Chouraqui, P ;
Hellman, C ;
Lubin, E .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) :474-478
[10]   Variations in pulse-height spectrum and pulse timing in CdZnTe pixel array detectors [J].
Eskin, JD ;
Barber, HB ;
Barrett, HH .
HARD X-RAY/GAMMA-RAY AND NEUTRON OPTICS, SENSORS, AND APPLICATIONS, 1996, 2859 :46-49