Comparison between the Apnea-Hypopnea Indices Determined by the REMstar Auto M Series and Those Determined by Standard In-Laboratory Polysomnography in Patients with Obstructive Sleep Apnea

被引:29
作者
Ikeda, Yukiko [1 ]
Kasai, Takatoshi [1 ]
Kawana, Fusae [1 ]
Kasagi, Satoshi [1 ]
Takaya, Hisashi [1 ]
Ishiwata, Sugao [1 ]
Narui, Koji [1 ]
机构
[1] Juntendo Univ, Sch Med, Dept Cardiol, Tokyo, Japan
关键词
apnea; hypopnea; polysomnography; obstructive sleep apnea; POSITIVE AIRWAY PRESSURE; PRACTICE PARAMETERS; APNOEA/HYPOPNOEA SYNDROME; BREATHING DISORDERS; CPAP; DIAGNOSIS; DEVICE; ADULTS;
D O I
10.2169/internalmedicine.51.8249
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective Continuous positive airway pressure (CPAP) is an effective treatment for obstructive sleep apnea (OSA). After performing an initial titration study, most physicians do not have the CPAP equipment retitrated unless the patient complains about the CPAP use. Several automated CPAP devices are used clinically that can detect upper airway obstructive events and provide information about residual events while patients are on CPAP. The aim of this study was to compare the apnea-hypopnea index (AHI) determined by automated CPAP devices to that obtained from polysomnography. Methods Patients with OSA underwent polysomnography for CPAP titration using the REMstar Auto M-series. The initial two hours of CPAP titration were spent at a subtherapeutic pressure of 4 cmH(2)O so that more breathing events could be observed. The correlations between the simultaneous determination of the AHI with polysomnography (AHI-PSG) and the automated device (AHI-RAM) during the subtherapeutic, therapeutic and overall phases were evaluated. In addition, the apnea index (AI) and the hypopnea index (HI) were each evaluated separately. Results Sixty patients were enrolled. The mean AHI on diagnostic PSG was 35.2+/-2.6 events/hour. Strong correlations were observed between the AHI-PSG and the AHI-RAM (subtherapeutic: r=0.958, p<0.001; therapeutic: r=0.824, p<0.001; overall: r=0.927, p<0.001). A slightly stronger correlation was observed between the AI values, whereas a weaker correlation was observed between the HI values in all three phases. Conclusion Strong correlations between the AHI-PSG and the AHI-RAM were observed. The correlations were weakened when the analysis was limited to the HI and the therapeutic phase.
引用
收藏
页码:2877 / 2885
页数:9
相关论文
共 21 条
[1]  
[Anonymous], 1997, SLEEP, V20, P406
[2]  
[Anonymous], 1999, SLEEP, V22, P667
[3]  
Baltzan MA, 2006, SLEEP, V29, P557
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   COMPARISON OF POLYSOMNOGRAPHY WITH RESCARE AUTOSET IN THE DIAGNOSIS OF THE SLEEP-APNEA HYPOPNEA SYNDROME [J].
BRADLEY, PA ;
MORTIMORE, IL ;
DOUGLAS, NJ .
THORAX, 1995, 50 (11) :1201-1203
[6]   EFFECT OF CONTINUOUS POSITIVE AIRWAY PRESSURE TREATMENT ON DAYTIME FUNCTION IN SLEEP APNOEA/HYPOPNOEA SYNDROME [J].
ENGLEMAN, HM ;
MARTIN, SE ;
DEARY, IJ ;
DOUGLAS, NJ .
LANCET, 1994, 343 (8897) :572-575
[7]   ACCURACY OF AN INTELLIGENT CPAP MACHINE WITH IN-BUILT DIAGNOSTIC ABILITIES IN DETECTING APNEAS - A COMPARISON WITH POLYSOMNOGRAPHY [J].
GUGGER, M ;
MATHIS, J ;
BASSETTI, C .
THORAX, 1995, 50 (11) :1199-1201
[8]  
Gugger M, 1997, EUR RESPIR J, V10, P587
[9]   Nasal CPAP improves the quality of life and lessens the depressive symptoms in patients with obstructive sleep apnea syndrome [J].
Kawahara, S ;
Akashiba, T ;
Akahoshi, T ;
Horie, T .
INTERNAL MEDICINE, 2005, 44 (05) :422-427
[10]   Practice parameters for the indications for polysomnography and related procedures: An update for 2005 [J].
Kushida, CA ;
Littner, MR ;
Morgenthaler, T ;
Alessi, CA ;
Bailey, D ;
Coleman, J ;
Friedman, L ;
Hirshkowitz, M ;
Kapen, S ;
Kramer, M ;
Lee-Chiong, T ;
Loube, DL ;
Owens, J ;
Pancer, JP ;
Wise, M .
SLEEP, 2005, 28 (04) :499-521