The Quick EXAFS Setup at Beamline P64 at PETRA III for up to 200 Spectra per Second

被引:26
作者
Bornmann, Benjamin [1 ]
Klaes, Jonas [1 ]
Mueller, Oliver [2 ]
Luetzenkirchen-Hecht, Dirk [1 ]
Frahm, Ronald [1 ]
机构
[1] Univ Wuppertal, Fac Phys 4, D-42097 Wuppertal, Germany
[2] SSRL, 2575 Sand Hill Rd, Menlo Pk, CA 94025 USA
来源
13TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION (SRI2018) | 2019年 / 2054卷
关键词
IN-SITU;
D O I
10.1063/1.5084609
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The layout and the characteristics of the Quick scanning EXAFS (QEXAFS, ref. [1]) setup at beamline P64 at the storage ring PETRA III, DESY, Germany are described. The quickly oscillating cryo-cooled monochromator allows measurements from 4.5 keV to 44 keV with frequencies up to 100 Hz. The effect of tapering of the undulator on spectra as well as the different detection schemes with transmission, reflectivity and fluorescence are discussed. The data acquisition is explained in detail. Reference spectra at different frequencies and examples of time resolved user experiments are shown. The influence of noise and filtering on the spectra is analysed.
引用
收藏
页数:5
相关论文
共 5 条
[2]  
Frahm R., 2013, J PHYS C SER, V425
[3]   Tuning the Pt/CeO2 Interface by in Situ Variation of the Pt Particle Size [J].
Gaenzler, Andreas M. ;
Casapu, Maria ;
Maurer, Florian ;
Stoermer, Heike ;
Gerthsen, Dagmar ;
Ferre, Geraldine ;
Vernaux, Philippe ;
Bornmann, Benjamin ;
Frahm, Ronald ;
Murzin, Vadim ;
Nachtegaal, Maarten ;
Votsmeier, Martin ;
Grunwaldt, Jan-Dierk .
ACS CATALYSIS, 2018, 8 (06) :4800-4811
[4]   Quick scanning monochromator for millisecond in situ and in operando X-ray absorption spectroscopy [J].
Mueller, O. ;
Luetzenkirchen-Hecht, D. ;
Frahm, R. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2015, 86 (09)
[5]   Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution [J].
Mueller, Oliver ;
Nachtegaal, Maarten ;
Just, Justus ;
Luetzenkirchen-Hecht, Dirk ;
Frahm, Ronald .
JOURNAL OF SYNCHROTRON RADIATION, 2016, 23 :260-266