A portable wireless near-infrared spatially resolved spectroscopy system for use on brain and muscle

被引:28
作者
Everdell, N. L. [1 ]
Airantzis, D. [1 ]
Kolvya, C. [1 ]
Suzuki, T. [2 ]
Elwell, C. E. [1 ]
机构
[1] UCL, Dept Med Phys & Bioengn, London WC1E 6BT, England
[2] Hamamatsu Photon KK, Naka Ku, Hamamatsu, Shizuoka 4308587, Japan
基金
英国工程与自然科学研究理事会;
关键词
Near-infrared spectroscopy; Oximetry; TISSUE OXYGENATION; MOTOR IMAGERY; BLOOD-FLOW; RESPONSES; EXERCISE;
D O I
10.1016/j.medengphy.2013.04.011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have designed, built and successfully tested a prototype portable and wireless near-infrared spectroscopy system. It takes forward the well-established series of NIRO spectroscopy instruments made by Hamamatsu Photonics (Hamamatsu City, Japan). It uses an identical optical probe, and has a data acquisition rate of 10 Hz. It illuminates the tissue with laser diode sources at 3 wavelengths of 775, 810 and 850 nm, and detects the reflected light with 2 silicon photodiode detectors at 2 different separations, enabling spatially resolved spectroscopy to be performed. We have tested it with both in vitro and in vivo experiments to establish its basic functionality for use in studies of both brain and muscle. (C) 2013 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1692 / 1697
页数:6
相关论文
共 26 条
[1]   Deep and surface hemodynamic signal from functional time resolved transcranial near infrared spectroscopy compared to skin flowmotion [J].
Aletti, Federico ;
Re, Rebecca ;
Pace, Vincenzo ;
Contini, Davide ;
Molteni, Erika ;
Cerutti, Sergio ;
Bianchi, Anna Maria ;
Torricelli, Alessandro ;
Spinelli, Lorenzo ;
Cubeddu, Rinaldo ;
Baselli, Giuseppe .
COMPUTERS IN BIOLOGY AND MEDICINE, 2012, 42 (03) :282-289
[2]   Development of wearable optical topography system for mapping the prefrontal cortex activation [J].
Atsumori, Hirokazu ;
Kiguchi, Masashi ;
Obata, Akiko ;
Sato, Hiroki ;
Katura, Takusige ;
Funane, Tsukasa ;
Maki, Atsushi .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2009, 80 (04)
[3]   Reproducibility and sensitivity of muscle reoxygenation and oxygen uptake recovery kinetics following running exercise in the field [J].
Buchheit, Martin ;
Ufland, Pierre ;
Haydar, Bachar ;
Laursen, Paul B. ;
Ahmaidi, Said .
CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2011, 31 (05) :337-346
[4]  
Cope M, 1991, THESIS U LONDON, P134
[5]   ESTIMATION OF OPTICAL PATHLENGTH THROUGH TISSUE FROM DIRECT TIME OF FLIGHT MEASUREMENT [J].
DELPY, DT ;
COPE, M ;
VANDERZEE, P ;
ARRIDGE, S ;
WRAY, S ;
WYATT, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (12) :1433-1442
[6]  
Elwell CE, 2011, PHILOS T R SOC A, V369, P4358, DOI 10.1098/rsta.2011.0302
[7]   Oxidative metabolism in muscle [J].
Ferrari, M ;
Binzoni, T ;
Quaresima, V .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY B-BIOLOGICAL SCIENCES, 1997, 352 (1354) :677-683
[8]   A brief review on the history of human functional near-infrared spectroscopy (fNIRS) development and fields of application [J].
Ferrari, Marco ;
Quaresima, Valentina .
NEUROIMAGE, 2012, 63 (02) :921-935
[9]   A DESIGN FOR A STABLE AND REPRODUCIBLE PHANTOM FOR USE IN NEAR-INFRARED IMAGING AND SPECTROSCOPY [J].
FIRBANK, M ;
DELPY, DT .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) :847-853
[10]   Brain responses associated with the Valsalva maneuver revealed by functional magnetic resonance imaging [J].
Henderson, LA ;
Macey, PM ;
Macey, KE ;
Frysinger, RC ;
Woo, MA ;
Harper, RK ;
Alger, JR ;
Yan-Go, FL ;
Harper, RM .
JOURNAL OF NEUROPHYSIOLOGY, 2002, 88 (06) :3477-3486