Assessment of dosimetric impact of system specific geometric distortion in an MRI only based radiotherapy workflow for prostate

被引:40
|
作者
Gustafsson, C. [1 ,2 ]
Nordstrom, F. [3 ]
Persson, E. [1 ]
Brynolfsson, J. [4 ]
Olsson, L. E. [2 ]
机构
[1] Skane Univ Hosp, Dept Hematol Oncol & Radiat Phys, Lund, Sweden
[2] Lund Univ, Dept Med Phys, Malmo, Sweden
[3] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, Gothenburg, Sweden
[4] Spectron Med AB, Helsingborg, Sweden
关键词
MRI only; distortion; MRI radiotherapy; QA; synthetic CT; MRI treatment planning; RADIATION-THERAPY; ONLY RADIOTHERAPY; PHANTOM; CANCER; CT; ALGORITHM; SIMULATION; IMAGES; BRAIN; FIELD;
D O I
10.1088/1361-6560/aa5fa2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dosimetric errors in a magnetic resonance imaging (MRI) only radiotherapy workflow may be caused by system specific geometric distortion from MRI. The aim of this study was to evaluate the impact on planned dose distribution and delineated structures for prostate patients, originating from this distortion. A method was developed, in which computer tomography (CT) images were distorted using the MRI distortion field. The displacement map for an optimized MRI treatment planning sequence was measured using a dedicated phantom in a 3 T MRI system. To simulate the distortion aspects of a synthetic CT (electron density derived from MR images), the displacement map was applied to CT images, referred to as distorted CT images. A volumetric modulated arc prostate treatment plan was applied to the original CT and the distorted CT, creating a reference and a distorted CT dose distribution. By applying the inverse of the displacement map to the distorted CT dose distribution, a dose distribution in the same geometry as the original CT images was created. For 10 prostate cancer patients, the dose difference between the reference dose distribution and inverse distorted CT dose distribution was analyzed in isodose level bins. The mean magnitude of the geometric distortion was 1.97 mm for the radial distance of 200-250 mm from isocenter. The mean percentage dose differences for all isodose level bins, were <= 0.02% and the radiotherapy structure mean volume deviations were < 0.2%. The method developed can quantify the dosimetric effects of MRI system specific distortion in a prostate MRI only radiotherapy workflow, separated from dosimetric effects originating from synthetic CT generation. No clinically relevant dose difference or structure deformation was found when 3D distortion correction and high acquisition bandwidth was used. The method could be used for any MRI sequence together with any anatomy of interest.
引用
收藏
页码:2976 / 2989
页数:14
相关论文
共 35 条
  • [21] Assessment of specific versus combined purpose knowledge based models in prostate radiotherapy
    Bossart, Elizabeth
    Duffy, Melissa
    Simpson, Garrett
    Abramowitz, Matthew
    Pollack, Alan
    Dogan, Nesrin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (06): : 209 - 216
  • [22] Clinical assessment of geometric distortion for a 0.35T MR-guided radiotherapy system
    Neylon, John
    Cook, Kiri A.
    Yang, Yingli
    Du, Dongsu
    Sheng, Ke
    Chin, Robert K.
    Kishan, Amar U.
    Lamb, James M.
    Low, Daniel A.
    Cao, Minsong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (08): : 303 - 309
  • [23] Dosimetric evaluation of synthetic CT relative to bulk density assignment-based magnetic resonance-only approaches for prostate radiotherapy
    Kim, Joshua
    Garbarino, Kim
    Schultz, Lonni
    Levin, Kenneth
    Movsas, Benjamin
    Siddiqui, M. Salim
    Chetty, Indrin J.
    Glide-Hurst, Carri
    RADIATION ONCOLOGY, 2015, 10
  • [24] Assessment of MRI-Based Attenuation Correction for MRI-Only Radiotherapy Treatment Planning of the Brain
    Ranta, Iiro
    Teuho, Jarmo
    Linden, Jani
    Klen, Riku
    Teras, Mika
    Kapanen, Mika
    Keyrilainen, Jani
    DIAGNOSTICS, 2020, 10 (05)
  • [25] The Impact of System-Related Magnetic Resonance Imaging Geometric Distortion in Stereotactic Radiotherapy: A Case Report
    Kumagai, Motoki
    Kawamura, Mariko
    Kato, Yutaka
    Okudaira, Kuniyasu
    Naganawa, Shinji
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (07)
  • [26] Interfractional Geometric Variations and Dosimetric Benefits of Stereotactic MRI Guided Online Adaptive Radiotherapy (SMART) of Prostate Bed after Radical Prostatectomy: Post-Hoc Analysis of a Phase II Trial
    Cao, Minsong
    Gao, Yu
    Yoon, Stephanie M.
    Yang, Yingli
    Sheng, Ke
    Ballas, Leslie K.
    Basehart, Vincent
    Sachdeva, Ankush
    Felix, Carol
    Low, Daniel A.
    Steinberg, Michael L.
    Kishan, Amar U.
    CANCERS, 2021, 13 (11)
  • [27] Dosimetric evaluation of synthetic CT relative to bulk density assignment-based magnetic resonance-only approaches for prostate radiotherapy
    Joshua Kim
    Kim Garbarino
    Lonni Schultz
    Kenneth Levin
    Benjamin Movsas
    M. Salim Siddiqui
    Indrin J. Chetty
    Carri Glide-Hurst
    Radiation Oncology, 10
  • [28] Optimal planning target margin for prostate radiotherapy based on interfractional and intrafractional variability assessment during 1.5T MRI-guided radiotherapy
    Kim, Jina
    Sung, Jiwon
    Lee, Seo Jin
    Cho, Kang Su
    Chung, Byung Ha
    Yang, Dongjoon
    Kim, Jihun
    Kim, Jun Won
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [29] A generative adversarial network-based (GAN-based) architecture for automatic fiducial marker detection in prostate MRI-only radiotherapy simulation images
    Singhrao, Kamal
    Fu, Jie
    Parikh, Neil R.
    Mikaeilian, Argin G.
    Ruan, Dan
    Kishan, Amar U.
    Lewis, John H.
    MEDICAL PHYSICS, 2020, 47 (12) : 6405 - 6413
  • [30] Cone beam computed tomography based image guidance and quality assessment of prostate cancer for magnetic resonance imaging-only radiotherapy in the pelvis