摘要
Spectral drift of any origin consisting of detectors and nuclear electronics affects the long-term stability in a gamma-ray spectrometer, which is essential to maintaining the high quality of gamma energy spectra. The feedback regulation according to the quantity of spectral drift can restore the system's stability. A new algorithm for spectral drift calculation is given in this paper. A digital logic unit is developed in a FPGA chip based on the algorithm, and a MCU coordinating with it can fast calculate the spectral drift within only less than 110 instruction periods.
引用
收藏
页码:705 / +
页数:2
相关论文
共 50 条
  • [41] A PRECISION CURVED CRYSTAL GAMMA-RAY SPECTROMETER
    RYDE, N
    ANDERSSON, B
    PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION B, 1955, 68 (12): : 1117 - &
  • [42] GE(LI) SPECTROMETER FOR GAMMA-RAY ASTRONOMY
    HICKS, DB
    JACOBSON, AS
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1974, NS21 (01) : 169 - 177
  • [43] GAMMA-RAY MEASUREMENTS WITH A MAGNETIC LENS SPECTROMETER
    HORNYAK, WF
    LAURITSEN, T
    RASMUSSEN, VK
    PHYSICAL REVIEW, 1949, 76 (06): : 731 - 739
  • [44] STUDIES ON THE LINEARITY OF A SCINTILLATION GAMMA-RAY SPECTROMETER
    ERIKSEN, VO
    JENSSEN, G
    PHYSICAL REVIEW, 1952, 85 (01): : 150 - 150
  • [45] AN ULTRALOW BACKGROUND GERMANIUM GAMMA-RAY SPECTROMETER
    BRODZINSKI, RL
    BROWN, DP
    EVANS, JC
    HENSLEY, WK
    REEVES, JH
    WOGMAN, NA
    AVIGNONE, FT
    MILEY, HS
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1985, 239 (02): : 207 - 213
  • [46] A COAXIAL GE(LI) GAMMA-RAY SPECTROMETER
    SAKAI, E
    SUZUKI, T
    GOTOH, H
    SEKIGUCH.N
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY-TOKYO, 1966, 3 (12): : 534 - &
  • [47] CERENKOV SPECTROMETER FOR MEASURING GAMMA-RAY ENERGIES
    HACHATURJAN, MN
    PANTUJEV, VS
    NUCLEAR INSTRUMENTS & METHODS, 1963, 25 (01): : 173 - 176
  • [48] COMPUTER SIMULATION OF AN AIRBORNE GAMMA-RAY SPECTROMETER
    CLARK, RB
    ADAMS, JAS
    DUVAL, JS
    JOURNAL OF GEOPHYSICAL RESEARCH, 1972, 77 (17): : 3021 - +
  • [49] Germanium Gamma-Ray Spectrometer on SELENE (KAGUYA)
    Yamashita, Naoyuki
    Hasebe, Nobuyuki
    Shibamura, Eido
    Miyachi, Takashi
    Takashima, Takeshi
    Kobayashi, Masanori
    Okudaira, Osamu
    Kobayashi, Shingo
    Hareyama, Makoto
    Karouji, Yuzuru
    Kodaira, Satoshi
    Sakurai, Kunitomo
    Iwabuchi, Kazuya
    Hayatsu, Kanako
    Nemoto, Shinpei
    Ebihara, Mitsuru
    Hihara, Takeshi
    D'Uston, Claude
    Maurice, Sylvestre
    Gasnault, Olivier
    Forni, Olivier
    Diez, Benedicte
    Reedy, Robert C.
    Kim, Kyeong J.
    JOURNAL OF THE PHYSICAL SOCIETY OF JAPAN, 2009, 78 : 153 - 156
  • [50] BOREHOLE GAMMA-RAY SPECTROMETER FOR URANIUM EXPLORATION
    GEORGE, DC
    EVANS, HB
    GEOPHYSICS, 1979, 44 (03) : 413 - 413