Ex Vivo Biomechanical Comparison of Titanium Locking Plate, Stainless Steel Nonlocking Plate, and Tie-in External Fixator Applied by a Dorsal Approach on Ostectomized Humeri of Pigeons (Columba livia)

被引:4
作者
Darrow, Brett G. [1 ]
Weigel, Joseph P. [1 ]
Greenacre, Cheryl B. [1 ]
Xie, Xie [2 ,3 ]
Liaw, Peter K. [2 ,3 ]
Biskup, Jeffery J. [1 ]
机构
[1] Univ Tennessee, Coll Vet Med, Small Anim Clin Sci, 2407 River Dr C-247, Knoxville, TN 37996 USA
[2] Univ Tennessee, Coll Engn, Dept Mat Sci, 414 Ferris Hall,1508 Middle Dr, Knoxville, TN 37996 USA
[3] Univ Tennessee, Coll Engn, Dept Engn, 414 Ferris Hall,1508 Middle Dr, Knoxville, TN 37996 USA
关键词
fracture; osteosynthesis; humerus; locking plate; nonlocking plate; external skeletal fixator; orthopedic; avian; pigeon; Columba livia; EXPERIMENTALLY-INDUCED ULNAR; MINIPLATE SYSTEMS; RADIAL FRACTURES; STABILIZATION; REPAIR;
D O I
10.1647/2017-305
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
To compare the bending strength of a locking plate (LP), nonlocking plate (NLP), and an external skeletal fixator intramedullary pin (ESF-IM) tie-in fixation applied by a dorsal approach in an avian humerus fracture model, 5 left humeri obtained from pigeon (Columba livia) cadavers were randomly assigned to each repair technique (n = 15). The ESF-IM group was repaired with a 0.062-inch intramedullary pin tied-in with two 0.035-inch positive profile transfixation pins using acrylic filled plastic tubing. The LP group was repaired with a dorsally applied titanium 1.6-mm screw 7-hole locking plate (1 bicortical and 2 monocortical screws in each segment). The NLP group was repaired with a dorsally applied 6-hole stainless steel 1.5-mm dynamic compression plate (all bicortical screws). All constructs were applied before complete ostectomy to allow perfect reconstruction. Constructs were cyclically tested nondestructively for 1000 cycles in four-point bending before being tested to failure. Outcome measures included stiffness, strength, and strain energy. All specimens cycled without failure. The ESF-IM specimens were significantly stiffer and stronger than the plated repair groups. Plated constructs had significantly higher strain energies than ESF-IM. LP and NLP were of equal stiffness, strength, and strain energies. This study demonstrated that bending biomechanical properties of the ESF-IM configuration were superior to those of the dorsal plate fixation. Exact properties of fixation required to facilitate avian fracture healing are largely unknown. Further study, including assessments of optimal plate position and configuration, and torsional and in vivo studies in avian species are warranted.
引用
收藏
页码:29 / 37
页数:9
相关论文
共 20 条
[1]   Influence of materials for fixation implants on local infection - An experimental study of steel versus titanium DCP in rabbits [J].
Arens, S ;
Schlegel, U ;
Printzen, G ;
Ziegler, WJ ;
Perren, SM ;
Hansis, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (04) :647-651
[2]  
Beaufrère H, 2009, J AVIAN MED SURG, V23, P173, DOI 10.1647/2007-023.1
[3]  
Bennert BM, 2016, J AVIAN MED SURG, V30, P111, DOI 10.1647/2015-082
[4]  
BENNETT RA, 1992, J ZOO WILDLIFE MED, V23, P5
[5]   IN-VIVO STRAIN IN THE HUMERUS OF PIGEONS (COLUMBA-LIVIA) DURING FLIGHT [J].
BIEWENER, AA ;
DIAL, KP .
JOURNAL OF MORPHOLOGY, 1995, 225 (01) :61-75
[6]   Single cycle to failure in bending of three standard and five locking plates and plate constructs [J].
Blake, C. A. ;
Boudrieau, R. J. ;
Torrance, B. S. ;
Tacvorian, E. K. ;
Cabassu, J. B. ;
Gaudette, G. R. ;
Kowaleski, M. P. .
VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2011, 24 (06) :408-417
[7]   Biomechanical Concepts Applicable to Minimally Invasive Fracture Repair in Small Animals [J].
Chao, Peini ;
Lewis, Daniel D. ;
Kowaleski, Michael P. ;
Pozzi, Antonio .
VETERINARY CLINICS OF NORTH AMERICA-SMALL ANIMAL PRACTICE, 2012, 42 (05) :853-+
[8]   Evaluation of a maxillofacial miniplate compact 1.0 for stabilization of the ulna in experimentally induced ulnar and radial fractures in pigeons (Columba livia) [J].
Christen, C ;
Fischer, I ;
von Rechenberg, B ;
Flückiger, M ;
Hatt, JM .
JOURNAL OF AVIAN MEDICINE AND SURGERY, 2005, 19 (03) :185-190
[9]   Ex vivo biomechanical evaluation of pigeon (Columba livia) cadaver intact humeri and ostectomized humeri stabilized with caudally applied titanium locking plate or stainless steel nonlocking plate constructs [J].
Darrow, Brett G. ;
Biskup, Jeffrey J. ;
Weigel, Joseph P. ;
Jones, Michael P. ;
Xie, Xie ;
Liaw, Peter K. ;
Tharpe, Josh L. ;
Sharma, Aashish ;
Penumadu, Dayakar .
AMERICAN JOURNAL OF VETERINARY RESEARCH, 2017, 78 (05) :570-578
[10]   Holding power of different pin designs and pin insertion methods in avian cortical bone [J].
Degernes, LA ;
Roe, SC ;
Abrams, CF .
VETERINARY SURGERY, 1998, 27 (04) :301-306