Shape selectivity: A key factor in comprehensive two-dimensional gas chromatographic analysis of toxic PCBs

被引:38
作者
Haglund, P [1 ]
Harju, M
Ong, R
Marriott, P
机构
[1] Umea Univ, SE-90187 Umea, Sweden
[2] Royal Melbourne Inst Technol, Melbourne, Vic 3001, Australia
关键词
comprehensive two-dimensional gas chromatography; fast analysis; peak compression; shape selectivity; toxic PCBs;
D O I
10.1002/mcs.10011
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A number of toxic planar PCBs, that is, PCB 77, 105, 118, 126, 156, and 169, were successfully separated from other PCB congeners present in technical PCB formulations using comprehensive GC X GC. The planar PCBs were selectively retained using a liquid crystal column, which was used as the first-dimension column. A short and narrow bore (0.25 In. X 0.1 mm) nonpolar 5%-phenyl-methylpoly-siloxane column was used as the second-dimension column. A longitudinally modulating cryogenic system (LMCS) was used to modulate the first-dimension signal. Since the planar PCBs elute from the first column at relatively high oven temperatures they were rapidly eluted from the second column. Altogether, this resulted in a significant peak sharpening, and a 20-fold increase in the signal-to-noise ratio, as compared to standard one-dimensional GC. This column set also seems to separate seven frequently measured congeners, that is, PCBs 28, 52, 101, 118, 138, 153, and 180, from other major PCB congeners. (C) 2001 John Wiley & Sons, Inc.
引用
收藏
页码:306 / 311
页数:6
相关论文
共 6 条
[1]   Comprehensive, quantitative, congener-specific analyses of eight aroclors and complete PCB congener assignments on DB-1 capillary GC columns [J].
Frame, GM ;
Wagner, RE ;
Carnahan, JC ;
Brown, JF ;
May, RJ ;
Smullen, LA ;
Bedard, DL .
CHEMOSPHERE, 1996, 33 (04) :603-623
[2]  
HARJU MT, 1998, ORGANOHALOGEN COMPOU, V35, P111
[3]   CRITICAL-REVIEW OF THE ANALYSIS OF NON-ORTHO-CHLOROBIPHENYLS AND MONO-ORTHO-CHLOROBIPHENYLS [J].
HESS, P ;
DEBOER, J ;
COFINO, WP ;
LEONARDS, PEG ;
WELLS, DE .
JOURNAL OF CHROMATOGRAPHY A, 1995, 703 (1-2) :417-465
[4]   Longitudinally modulated cryogenic system. A generally applicable approach to solute trapping and mobilization in gas chromatography [J].
Marriott, PJ ;
Kinghorn, RM .
ANALYTICAL CHEMISTRY, 1997, 69 (13) :2582-2588
[5]  
Patterson D., COMMUNICATION
[6]   POLYCHLORINATED-BIPHENYLS (PCBS) - ENVIRONMENTAL-IMPACT, BIOCHEMICAL AND TOXIC RESPONSES, AND IMPLICATIONS FOR RISK ASSESSMENT [J].
SAFE, SH .