Investigation of dental samples using a 35 MHz focussed ultrasound piezocomposite transducer

被引:27
作者
Hughes, D. A. [1 ,5 ]
Girkin, J. M. [1 ]
Poland, S. [1 ]
Longbottom, C. [2 ]
Button, T. W. [3 ]
Elgoyhen, J. [3 ]
Hughes, H. [3 ]
Meggs, C. [3 ]
Cochran, S. [4 ]
机构
[1] Univ Strathclyde, SUPA, Inst Photon, Glasgow G4 0NW, Lanark, Scotland
[2] Univ Dundee, Ctr Clin Innovat, Dundee, Scotland
[3] Appl Funct Mat Ltd, Birmingham B15 2SQ, W Midlands, England
[4] Univ Dundee, IMSAT, Dundee DD2 1FD, Scotland
[5] Univ Strathclyde, Dept Bioengn, Glasgow G4 0NW, Lanark, Scotland
基金
英国工程与自然科学研究理事会;
关键词
Human teeth; Biomedical imaging; High frequency; HUMAN TEETH; ENAMEL THICKNESS; EROSION; PREVENTION; VELOCITY;
D O I
10.1016/j.ultras.2008.08.007
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Dental erosion and decay are increasingly prevalent but as yet there is no quantitative monitoring tool. Such a tool would allow earlier diagnosis and treatment and ultimately the prevention of more serious disease and pain. Despite ultrasound having been demonstrated as a method of probing the internal structures of teeth more than 40 years ago, development of a clinical tool has been slow. The aim of the study reported here was to investigate the use of a novel high frequency ultrasound transducer and validate it using a known dental technique. A tooth extracted for clinical reasons was sectioned to provide a sample that contained an enamel and dentine layer such that the enamel-dentine junction (EDJ) was of a varying depth. The sample was then submerged in water and a B-scan recorded using a custom-designed piezocomposite ultrasound transducer with a centre frequency of 35 MHz and a -6 dB bandwidth of 24 MHz. The transducer has an axial resolution of 180 mu m and a spatial resolution of 110 mu m, a significant advance on previous work using lower frequencies. The depth of the EDJ was measured from the resulting data set and compared to measurements from the sequential grinding and imaging (SGI) method. The B-scan showed that the EDJ was of varying depth. Subsequently, the EDJ measurements were found to have a correlation of 0.89 (p < 0.01) against the SGI measurements. The results indicate that high frequency ultrasound is capable of measuring enamel thickness to an accuracy of within 10% of the total enamel thickness, whereas currently there is no clinical tool available to measure enamel thickness. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:212 / 218
页数:7
相关论文
共 26 条
  • [1] 1-3 connectivity piezoelectric ceramic-polymer composite transducers made with viscous polymer processing for high frequency ultrasound[J]. Abrar, A;Zhang, D;Su, B;Button, TW;Kirk, KJ;Cochran, S. ULTRASONICS, 2004(1-9)
  • [2] Dental erosion: possible approaches to prevention and control[J]. Amaechi, BT;Higham, SM. JOURNAL OF DENTISTRY, 2005(03)
  • [3] [Anonymous], 2002, The Essential Physics of Medical Imaging
  • [4] The measurement and prevention of erosion and abrasion[J]. Azzopardi, A;Bartlett, DW;Watson, TF;Sherriff, M. JOURNAL OF DENTISTRY, 2001(06)
  • [5] Surface preparation of 1-3 piezocomposite material for microfabrication of high frequency transducer arrays[J]. Bernassau, A. L.;McKay, S.;Hutson, D.;Demore, C. E. M.;Garcia-Gancedo, L.;Batton, T. W.;McAneny, J. J.;Cochran, C. 2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007
  • [6] Button TW, 2005, ULTRASON, P1625
  • [7] Development of a 35-MHz piezo-composite ultrasound array for medical imaging[J]. Cannata, JM;Williams, JA;Zhou, QF;Ritter, TA;Shung, KK. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006(01)
  • [8] Imaging of human tooth enamel using ultrasound[J]. Culjat, M;Singh, RS;Yoon, DC;Brown, ER. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003(04)
  • [9] DONALD I, 1958, LANCET, V1, P1188
  • [10] A history of medical and biological imaging with polyvinylidene fluoride (PVDF) transducers[J]. Foster, FS;Harasiewicz, EA;Sherar, MD. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000(06)