Stent salvage using the Enterprise stent for procedure-related complication during coil embolization of ruptured intracranial aneurysms

被引:9
作者
Chung, Joonho [1 ]
Kim, Yong Bae [1 ]
Hong, Chang-Ki [1 ]
Joo, Jin Yang [1 ]
Shin, Yong Sam [2 ]
Lim, Yong Cheol [3 ]
机构
[1] Yonsei Univ, Coll Med, Dept Neurosurg, Gangnam Severance Hosp, Seoul, South Korea
[2] Catholic Univ Korea, Seoul St Marys Hosp, Dept Neurosurg, Seoul, South Korea
[3] Ajou Univ Sch Med & Hosp, Dept Neurosurg, Suwon 442721, South Korea
关键词
Coil embolization; Enterprise stent; Intracranial aneurysm; Stent salvage; GUGLIELMI DETACHABLE COIL; WIDE-NECKED ANEURYSMS; CEREBRAL ANEURYSMS; NEUROFORM STENT; TECHNICAL NOTE; FOLLOW-UP; EXPERIENCE; HERNIATION; PLACEMENT; RETRIEVAL;
D O I
10.1007/s00701-012-1528-5
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Despite accumulated experience and improved understanding of the tools, endovascular treatment of intracranial aneurysms continues to have risks linked to the technique itself, and induces procedure-related complications. The purpose of this study was to report our series of stent salvage using the Enterprise stent for procedure-related complication during coil embolization in patients with ruptured intracranial aneurysms. Parent artery thrombosis, parent artery dissection, and coil protrusion were considered to be the procedure-related complications. There were 18 consecutive cases (3 unruptured and 15 ruptured aneurysms) with procedure-related complications rescued by the Enterprise stent from December 2008 to December 2011. Follow-up angiography was performed in 14 of the 15 patients with ruptured aneurysms between 6 and 30 months (mean 14.6 months) after the procedure. The procedure-related complications were parent artery dissection (n = 1), parent artery thrombosis (n = 4), and coil protrusion (n = 10). There was no complication related to delivering or deploying of the Enterprise stent. Initial radiographic results showed 8 cases of complete occlusion and 7 cases of neck remnant. There was no change in the angiographic results during the follow-up periods. Facing with procedure-related complications during coil embolization of ruptured intracranial aneurysms, the closed-cell designed Enterprise stent might be a useful option for the salvage technique by restoring blood flow and minimizing thromboembolic events.
引用
收藏
页码:223 / 229
页数:7
相关论文
共 18 条
[11]   SPONTANEOUS DELAYED PROXIMAL MIGRATION OF ENTERPRISE STENT AFTER STAGED TREATMENT OF WIDE-NECKED BASILAR ANEURYSM: TECHNICAL CASE REPORT [J].
Lavine, Sean D. ;
Meyers, Philip M. ;
Connolly, E. Sander ;
Solomon, Robert S. .
NEUROSURGERY, 2009, 64 (05) :1012-1012
[12]   Treatment of intracranial aneurysms with the Enterprise stent: a multicenter registry [J].
Mocco, J. ;
Snyder, Kenneth V. ;
Albuquerque, Felipe C. ;
Bendok, Bernard R. ;
Boulos, Alan S. ;
Carpenter, Jeffrey S. ;
Fiorella, David J. ;
Hoti, Brian L. ;
Howington, Jay U. ;
Jankowitz, Brian T. ;
Liebman, Kenneth M. ;
Rai, Ansaar T. ;
Rodriguez-Mercado, Rafael ;
Siddiqui, Adnan H. ;
Veznedaroglu, Erol ;
Hopkins, L. Nelson ;
Levy, Elad I. .
JOURNAL OF NEUROSURGERY, 2009, 110 (01) :35-39
[13]   Coil Migration through a Neuroform 3 Stent during Endovascular Coiling A Case Report [J].
O'Hare, A. ;
Thornton, J. ;
Brennan, P. .
INTERVENTIONAL NEURORADIOLOGY, 2009, 15 (02) :219-222
[14]   Proximal stent fixation of fractured coils:: technical note [J].
Schütz, A ;
Solymosi, L ;
Vince, GH ;
Bendszus, M .
NEURORADIOLOGY, 2005, 47 (11) :874-878
[15]   One-stage embolization in patients with acutely ruptured poor-grade aneurysm [J].
Shin, YS ;
Kim, SY ;
Kim, SH ;
Ahn, YH ;
Yoon, SH ;
Cho, KH ;
Cho, KG .
SURGICAL NEUROLOGY, 2005, 63 (02) :149-155
[16]   Horizontal stent-assisted coil embolisation of wide-necked intracranial aneurysms with the Enterprise stent-a case series with early angiographic follow-up [J].
Siddiqui, Mohammed Aslam ;
Bhattacharya, Joti J. ;
Lindsay, Kenneth W. ;
Jenkins, Sarah .
NEURORADIOLOGY, 2009, 51 (06) :411-418
[17]   RETRIEVAL OF A GUGLIELMI DETACHABLE COIL AFTER UNRAVELING AND FRACTURE - CASE-REPORT AND EXPERIMENTAL RESULTS [J].
STANDARD, SC ;
CHAVIS, TD ;
WAKHLOO, AK ;
AHUJA, A ;
GUTERMAN, LR ;
HOPKINS, LN .
NEUROSURGERY, 1994, 35 (05) :994-998
[18]   Stent salvage for parent vessel coil herniation during intracranial aneurysm embolization [J].
Wong, George K. C. ;
Yu, Simon C. H. ;
Wong, Ka-Tak ;
Poon, Wai-Sang .
SURGICAL PRACTICE, 2009, 13 (04) :114-118