Quality assurance of MLC leaf position accuracy and relative dose effect at the MLC abutment region using an electronic portal imaging device

被引:15
作者
Sumida, Iori [1 ,2 ]
Yamaguchi, Hajime [2 ]
Kizaki, Hisao [2 ]
Koizumi, Masahiko [3 ]
Ogata, Toshiyuki [3 ]
Takahashi, Yutaka [3 ]
Yoshioka, Yasuo [3 ]
机构
[1] Osaka Univ, Dept Oral & Maxillofacial Radiol, Grad Sch Dent, Suita, Osaka 5650871, Japan
[2] NTT W Osaka Hosp, Dept Radiat Oncol, Tennoji Ku, Osaka 5438922, Japan
[3] Osaka Univ, Dept Radiat Oncol, Grad Sch Med, Suita, Osaka 5650871, Japan
关键词
MLC; IMRT; EPID; garden fence test; calibration; DYNAMIC MULTILEAF COLLIMATION; RADIATION-THERAPY; VERIFICATION; IMPACT; ERRORS;
D O I
10.1093/jrr/rrs038
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
We investigated an electronic portal image device (EPID)-based method to see whether it provides effective and accurate relative dose measurement at abutment leaves in terms of positional errors of the multi-leaf collimator (MLC) leaf position. A Siemens ONCOR machine was used. For the garden fence test, a rectangular field (0.2 x 20 cm) was sequentially irradiated 11 times at 2-cm intervals. Deviations from planned leaf positions were calculated. For the nongap test, relative doses at the MLC abutment region were evaluated by sequential irradiation of a rectangular field (2 x 20 cm) 10 times with a MLC separation of 2 cm without a leaf gap. The integral signal in a region of interest was set to position A (between leaves) and B (neighbor of A). A pixel value at position B was used as background and the pixel ratio (A/B x 100) was calculated. Both tests were performed at four gantry angles (0, 90, 180 and 270) four times over 1 month. For the nongap test the difference in pixel ratio between the first and last period was calculated. Regarding results, average deviations from planned positions with the garden fence test were within 0.5 mm at all gantry angles, and at gantry angles of 90 and 270 tended to decrease gradually over the month. For the nongap test, pixel ratio tended to increase gradually in all leaves, leading to a decrease in relative doses at abutment regions. This phenomenon was affected by both gravity arising from the gantry angle, and the hardware-associated contraction of field size with this type of machine.
引用
收藏
页码:798 / 806
页数:9
相关论文
共 22 条
[1]   MLC quality assurance techniques for IMRT applications [J].
Bayouth, JE ;
Wendt, D ;
Morrill, SM .
MEDICAL PHYSICS, 2003, 30 (05) :743-750
[2]  
Bhardwaj Anup K, 2007, J Cancer Res Ther, V3, P23
[3]  
Boyer A., 2001, Basic Applications of multileaf collimators
[4]   Improving IMRT quality control efficiency using an amorphous silicon electronic portal imager [J].
Budgell, GJ ;
Zhang, Q ;
Trouncer, RJ ;
Mackay, RI .
MEDICAL PHYSICS, 2005, 32 (11) :3267-3278
[5]   Use of EPID for leaf position accuracy QA of dynamic multi-leaf collimator (DMLC) treatment [J].
Chang, J ;
Obcemea, CH ;
Sillanpaa, J ;
Mechalakos, J ;
Burman, C .
MEDICAL PHYSICS, 2004, 31 (07) :2091-2096
[6]   Testing of dynamic multileaf collimation [J].
Chui, CS ;
Spirou, S ;
LoSasso, T .
MEDICAL PHYSICS, 1996, 23 (05) :635-641
[7]   Beam characteristics of a retrofitted double-focused multileaf collimator [J].
Das, IJ ;
Desobry, GE ;
McNeeley, SW ;
Cheng, EC ;
Schultheiss, TE .
MEDICAL PHYSICS, 1998, 25 (09) :1676-1684
[8]   Task Group 142 report: Quality assurance of medical accelerators [J].
Klein, Eric E. ;
Hanley, Joseph ;
Bayouth, John ;
Yin, Fang-Fang ;
Simon, William ;
Dresser, Sean ;
Serago, Christopher ;
Aguirre, Francisco ;
Ma, Lijun ;
Arjomandy, Bijan ;
Liu, Chihray ;
Sandin, Carlos ;
Holmes, Todd .
MEDICAL PHYSICS, 2009, 36 (09) :4197-4212
[9]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927
[10]   Characterization of a commercial multileaf collimator used for intensity modulated radiation therapy [J].
Low, DA ;
Sohn, JW ;
Klein, EE ;
Markman, J ;
Mutic, S ;
Dempsey, JF .
MEDICAL PHYSICS, 2001, 28 (05) :752-756