SIZE MEASUREMENT OF TISSUE DEBRIS PARTICLES GENERATED FROM PULSED ULTRASOUND CAVITATIONAL THERAPY - HISTOTRIPSY

被引:58
作者
Xu, Zhen [1 ]
Fan, Zhenzhen [1 ]
Hall, Timothy L. [1 ]
Winterroth, Frank [1 ]
Fowlkes, J. Brian [1 ,2 ]
Cain, Charles A. [1 ,3 ]
机构
[1] Univ Michigan, Dept Biomed Engn, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Radiol, Ann Arbor, MI 48109 USA
[3] Univ Michigan, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
基金
美国国家卫生研究院;
关键词
Histotripsy; Therapeutic ultrasound; Cavitation; Pulsed ultrasound; Tissue fractionation; Embolization; FOCUSED ULTRASOUND; HIGH-INTENSITY; IN-VIVO; THRESHOLD; EROSION; SYSTEM; BACKSCATTER; ABLATION; KIDNEY;
D O I
10.1016/j.ultrasmedbio.2008.09.002
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Extensive mechanical tissue fractionation can be achieved using successive high intensity ultrasound pulses ("histotripsy"). Histotripsy has many potential medical applications where noninvasive tissue removal is desired (e.g., tumor ablation). There is a concern that debris generated by histotripsy-induced tissue fractionation might be an embolization hazard. The aim of this study is to measure the size distribution of these tissue debris particles. Histotripsy pulses were produced by a 513-element 1 MHz array transducer, an 18-element 750 kHz array transducer and a 788 kHz single element transducer. Peak negative pressures of 11 to 25 MPa, pulse durations of 3 to 50 cycles, pulse repetition frequencies of 100 Hz to 2 kHz were used. Tissue debris particles created by histotripsy were collected and measured with a particle sizing system. In the resulting samples, debris <6 mu m in diameter constituted >99% of the total number of tissue particles. The largest particle generated by one of the parameter sets tested was 54 mu m in diameter, which is smaller than the clinic filter size (100 mu m) used to prevent embolization. The largest particles generated using other parameter sets were larger than 60 mu m but the value could not be specified using our current setup. Exposures with shorter pulses produced lower percentages of large tissue debris (>30 mu m) in comparison to longer pulses. These results suggest that the tissue debris particle size distribution is adjustable by altering acoustic parameters if necessary. (E-mail: zhenx@umich.edu) (c) 2009 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:245 / 255
页数:11
相关论文
共 28 条
[1]  
CHAPELON JY, 1992, CANCER RES, V52, P6353
[2]   THE CAVITATION THRESHOLD OF HUMAN TISSUE EXPOSED TO 0.2-MHZ PULSED ULTRASOUND - PRELIMINARY MEASUREMENTS BASED ON A STUDY OF CLINICAL LITHOTRIPSY [J].
COLEMAN, AJ ;
KODAMA, T ;
CHOI, MJ ;
ADAMS, T ;
SAUNDERS, JE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (03) :405-417
[3]   TREATMENT OF THE DUNNING PROSTATE RAT-TUMOR R3327-AT1 WITH PULSED HIGH-ENERGY ULTRASOUND SHOCK-WAVES (PHEUS) - GROWTH DELAY AND HISTOMORPHOLOGIC CHANGES [J].
DEBUS, J ;
PESCHKE, P ;
HAHN, EW ;
LORENZ, WJ ;
LORENZ, A ;
IFFLAENDER, H ;
ZABEL, HJ ;
VANKAICK, G ;
PFEILER, M .
JOURNAL OF UROLOGY, 1991, 146 (04) :1143-1146
[4]   ULTRASONIC THRESHOLD DOSAGES FOR MAMMALIAN CENTRAL NERVOUS SYSTEM [J].
DUNN, F ;
FRY, FJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1971, BM18 (04) :253-&
[5]   ACOUSTIC GENERATION OF BUBBLES IN EXCISED CANINE URINARY BLADDERS [J].
FOWLKES, JB ;
CARSON, PL ;
CHIANG, EH ;
RUBIN, JM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1991, 89 (06) :2740-2744
[6]   INVOLVEMENT OF ULTRASONICALLY INDUCED CAVITATION IN THE PRODUCTION OF HINDLIMB PARALYSIS OF THE MOUSE NEONATE [J].
FRIZZELL, LA ;
LEE, CS ;
ASCHENBACH, PD ;
BORRELLI, MJ ;
MORIMOTO, RS ;
DUNN, F .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1983, 74 (03) :1062-1065
[7]  
FRY F J, 1970, Journal of the Acoustical Society of America, V48, P1413, DOI 10.1121/1.1912301
[8]   A real-time measure of cavitation induced tissue disruption by ultrasound imaging backscatter reduction [J].
Hall, Timothy L. ;
Fowlkes, J. Brian ;
Cain, Charles A. .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (03) :569-575
[9]  
HALL TL, 2007, P JOINT ANN M ISMRM
[10]  
HALL TL, 2005, P INT S THER ULTR BO, P445