Effect of X-ray energies on induced photo-neutron doses

被引:4
作者
Khaled, N. E. [1 ]
Ghanim, E. H. [2 ]
Shinashin, Kh [3 ]
El-Sersy, A. R. [1 ]
机构
[1] Natl Inst Stand, Dept Ionizing Radiat Metrol, Giza, Egypt
[2] Beni Suef Univ, Fac Ind Educ, Dept Basic Sci, Bani Suwayf, Egypt
[3] Suez Canal Univ, Dept Phys, Faulty Sci, Port Saed, Egypt
来源
RADIATION EFFECTS AND DEFECTS IN SOLIDS | 2014年 / 169卷 / 03期
关键词
BN1 converter film; high-energy photons; dosimetry; photoneutrons; NTDs; LINEAR ACCELERATORS; RADIOTHERAPY; DOSIMETRY;
D O I
10.1080/10420150.2013.849250
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Photoneutrons induced by two high energies range from the Elekta medical linear accelerator (10 and 18 MV) were measured by nuclear track detectors (NTDs). CR-39 NTD in contact with converter screen slide films, natural boron of thickness 40 mu m coated on the polyester film (BN1). Detectors were exposed at 100 cm SSD with field size 20x20 on the patient table, with chest phantom and with build-up Perspex used for high-energy exposure. CR-39 registers the thermal neutron by the (n-alpha) reaction with the thin layer of boron and the fast neutron was measured through the (n-p) elastic scattering with the H-2 molecules in the CR-39 constituents.It was found that the total neutron dose (thermal and fast) from the 18 MV X-ray is higher than that of 10 MV. The measured thermal neutron dose is relatively smaller than the fast neutron dose in the case of direct exposure at the two X-ray energies. On the other hand, in the case of measurements on phantom and upon the use of build-up Perspex sheets, the ratio of fast to that of thermal is less than that of direct exposure.
引用
收藏
页码:239 / 248
页数:10
相关论文
共 8 条
[1]   Further investigations on CR-39 fast neutron personal dosemeter [J].
Djeffal, S ;
Lounis, Z ;
Allab, M ;
Izerrouken, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 398 (2-3) :343-350
[2]  
El-Sersy A. R., 2006, J BIOPHYS, V12, P131
[3]   Determination of CR-39 detection efficiency for fast neutron registration and the absolute neutron dosimetry [J].
El-Sersy, AR ;
Khaled, NE ;
Eman, SA .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 215 (3-4) :443-448
[4]   Calculation of photoneutrons produced in the targets of electron linear accelerators for radiography and radiotherapy applications [J].
Huang, WL ;
Li, QF ;
Lin, YZ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2005, 229 (3-4) :339-347
[5]   Dosimetry and fast neutron energies characterization of photoneutrons produced in some medical linear accelerators [J].
Khaled, N. E. ;
Attalla, E. M. ;
Ammar, H. ;
Khalil, W. .
RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2011, 166 (12) :908-917
[6]   Assessment and measurement of the photoneutron field produced in the varian medical linear accelerator [J].
Kim, Hong Suk ;
Lee, Jai Ki .
JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2007, 44 (01) :95-101
[7]   Investigation of photoneutron dose equivalent from high-energy photons in radiotherapy [J].
Lin, Jao-Perng ;
Liu, Wei-Chung ;
Lin, Chun-Chih .
APPLIED RADIATION AND ISOTOPES, 2007, 65 (05) :599-604
[8]   Analysis of photoneutron spectra produced in medical accelerators [J].
Ongaro, C ;
Zanini, A ;
Nastasi, U ;
Ródenas, J ;
Ottaviano, G ;
Manfredotti, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (12) :L55-L61