Multicenter evaluation of a new closed system drug-transfer device in reducing surface contamination by antineoplastic hazardous drugs

被引:28
作者
Bartel, Sylvia B. [1 ]
Tyler, Timothy G. [2 ]
Power, Luci A. [3 ]
机构
[1] Dana Farber Canc Inst, Pharm, Boston, MA 02115 USA
[2] Desert Reg Med Ctr, Ctr Comprehens Canc, Palm Springs, CA USA
[3] Power Enterprises, San Francisco, CA 92831 USA
关键词
antineoplastic drugs; closed system drug-transfer device; CSTD; cyclophosphamide; hazardous drugs; surface contamination; wipe sampling; HEALTH-CARE WORKERS; ENVIRONMENTAL CONTAMINATION; HOSPITAL PHARMACIES; CYCLOPHOSPHAMIDE; EXPOSURE; AGENTS; IMPLEMENTATION; REDUCTION; PERSONNEL; EFFICACY;
D O I
10.2146/ajhp160948
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Purpose. Results of a study to evaluate the effectiveness of a recently introduced closed system drug-transfer device (CSTD) in reducing surface contamination during compounding and simulated administration of antineoplastic hazardous drugs (AHDs) are reported. Methods. Wipe samples were collected from 6 predetermined surfaces in compounding and infusion areas of 13 U.S. cancer centers to establish preexisting levels of surface contamination by 2 marker AHDs (cyclophosphamide and fluorouracil). Stainless steel templates were placed over the 6 previously sampled surfaces, and the marker drugs were compounded and infused per a specific protocol using all components of the CSTD. Wipe samples were collected from the templates after completion of tasks and analyzed for both marker AHDs. Results. Aggregated results of wipe sampling to detect preexisting contamination at the 13 study sites showed that overall, 66.7% of samples (104 of 156) had detectable levels of at least 1 marker AHD; subsequent testing after CSTD use per protocol found a sample contamination rate of 5.8% (9 of 156 samples). In the administration areas alone, the rate of preexisting contamination was 78% (61 of 78 samples); with use of the CSTD protocol, the contamination rate was 2.6%. Twenty-six participants rated the CSTD for ease of use, with 100% indicating that they were satisfied or extremely satisfied. Conclusion. A study involving a rigorous protocol and 13 cancer centers across the United States demonstrated that the CSTD reduced surface contamination by cyclophosphamide and fluorouracil during compounding and simulated administration. Participants reported that the CSTD was easy to use.
引用
收藏
页码:201 / 213
页数:13
相关论文
共 27 条
[1]  
Chu Winnie C, 2012, J Oncol Pharm Pract, V18, P46, DOI 10.1177/1078155211402106
[2]  
Clark Bernadette A, 2013, J Oncol Pharm Pract, V19, P99, DOI 10.1177/1078155212468367
[3]   Effectiveness of a closed-system device in containing surface contamination with cyclophosphamide and ifosfamide in an i.v. admixture area [J].
Connor, TH ;
Anderson, RW ;
Sessink, PJ ;
Spivey, SM .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2002, 59 (01) :68-72
[4]   Surface wipe sampling for antineoplastic (chemotherapy) and other hazardous drug residue in healthcare settings: Methodology and recommendations [J].
Connor, Thomas H. ;
Zock, Matthew D. ;
Snow, Amy H. .
JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL HYGIENE, 2016, 13 (09) :658-667
[5]   Evaluation of Antineoplastic Drug Exposure of Health Care Workers at Three University-Based US Cancer Centers [J].
Connor, Thomas H. ;
DeBord, Gayle ;
Pretty, Jack R. ;
Oliver, Marc S. ;
Roth, Tracy S. ;
Lees, Peter S. J. ;
Krieg, Edward F., Jr. ;
Rogers, Bonnie ;
Escalante, Carmen P. ;
Toennis, Christine A. ;
Clark, John C. ;
Johnson, Belinda C. ;
McDiarmid, Melissa A. .
JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL MEDICINE, 2010, 52 (10) :1019-1027
[6]   Dermal exposure to cyclophosphamide in hospitals during preparation, nursing and cleaning activities [J].
Fransman, W ;
Vermeulen, R ;
Kromhout, H .
INTERNATIONAL ARCHIVES OF OCCUPATIONAL AND ENVIRONMENTAL HEALTH, 2005, 78 (05) :403-412
[7]   Impact and appreciation of two methods aiming at reducing hazardous drug environmental contamination: The centralization of the priming of IV tubing in the pharmacy and use of a closed-system transfer device [J].
Guillemette, Annie ;
Langlois, Helene ;
Voisine, Maxime ;
Merger, Delphine ;
Therrien, Roxane ;
Mercier, Genevieve ;
Lebel, Denis ;
Bussieres, Jean-Francois .
JOURNAL OF ONCOLOGY PHARMACY PRACTICE, 2014, 20 (06) :426-432
[8]   Comparison of surface contamination with cyclophosphamide and fluorouracil using a closed-system drug transfer device versus standard preparation techniques [J].
Harrison, Bruce R. ;
Peters, Byron G. ;
Bing, Michael R. .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2006, 63 (18) :1736-1744
[9]   Antineoplastic drug contamination in the urine of Canadian healthcare workers [J].
Hon, Chun-Yip ;
Teschke, Kay ;
Shen, Hui ;
Demers, Paul A. ;
Venners, Scott .
INTERNATIONAL ARCHIVES OF OCCUPATIONAL AND ENVIRONMENTAL HEALTH, 2015, 88 (07) :933-941
[10]   Occupational Exposure to Antineoplastic Drugs: Identification of Job Categories Potentially Exposed throughout the Hospital Medication System [J].
Hon, Chun-Yip ;
Teschke, Kay ;
Chua, Prescillia ;
Venners, Scott ;
Nakashima, Lynne .
SAFETY AND HEALTH AT WORK, 2011, 2 (03) :273-281