Repeatability of In Vivo Parafoveal Cone Density and Spacing Measurements

被引:126
|
作者
Garrioch, Robert [2 ,3 ]
Langlo, Christopher [4 ]
Dubis, Adam M.
Cooper, Robert F. [5 ]
Dubra, Alfredo [5 ,6 ,7 ]
Carroll, Joseph [1 ,5 ,6 ,7 ]
机构
[1] Med Coll Wisconsin, Inst Eye, Dept Cell Biol Neurobiol & Anat, Milwaukee, WI 53226 USA
[2] Med Coll Wisconsin, Summer Program Undergrad Res, Milwaukee, WI 53226 USA
[3] Cardiff Univ, Cardiff, S Glam, Wales
[4] Med Coll Wisconsin, Med Scientist Training Program, Milwaukee, WI 53226 USA
[5] Marquette Univ, Dept Biomed Engn, Milwaukee, WI 53233 USA
[6] Med Coll Wisconsin, Dept Ophthalmol, Milwaukee, WI 53226 USA
[7] Med Coll Wisconsin, Dept Biophys, Milwaukee, WI 53226 USA
基金
美国国家卫生研究院;
关键词
retina; cones; adaptive optics; repeatability; photoreceptors; PHOTORECEPTOR PACKING DENSITY; ADAPTIVE-OPTICS; RETINAL MOSAICS; RESOLUTION; REFLECTANCE; DISRUPTION; VISION;
D O I
10.1097/OPX.0b013e3182540562
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To assess the repeatability and measurement error associated with cone density and nearest neighbor distance (NND) estimates in images of the parafoveal cone mosaic obtained with an adaptive optics scanning light ophthalmoscope (AOSLO). Methods. Twenty-one participants with no known ocular pathology were recruited. Four retinal locations, approximately 0.65 degrees eccentricity from the center of fixation, were imaged 10 times in randomized order with an AOSLO. Cone coordinates in each image were identified using an automated algorithm (with or without manual correction) from which cone density and NND were calculated. Owing to naturally occurring fixational instability, the 10 images recorded from a given location did not overlap entirely. We thus analyzed each image set both before and after alignment. Results. Automated estimates of cone density on the unaligned image sets showed a coefficient of repeatability of 11,769 cones/mm(2) (17.1%). The primary reason for this variability appears to be fixational instability, as aligning the 10 images to include the exact same retinal area results in an improved repeatability of 4358 cones/mm(2) (6.4%) using completely automated cone identification software. Repeatability improved further by manually identifying cones missed by the automated algorithm, with a coefficient of repeatability of 1967 cones/mm(2) (2.7%). NND showed improved repeatability and was generally insensitive to the undersampling by the automated algorithm. Conclusions. As our data were collected in a young, healthy population, this likely represents a best-case estimate for corresponding measurements in patients with retinal disease. Similar studies need to be carried out on other imaging systems (including those using different imaging modalities, wavefront correction technology, and/or image analysis software), as repeatability would be expected to be highly sensitive to initial image quality and the performance of cone identification algorithms. Separate studies addressing intersession repeatability and interobserver reliability are also needed. (Optom Vis Sci 2012;89:632-643)
引用
收藏
页码:632 / 643
页数:12
相关论文
共 50 条
  • [21] A 2-Year Longitudinal Study of Normal Cone Photoreceptor Density
    Jackson, Kevin
    Vergilio, Grace K.
    Cooper, Robert F.
    Ying, Gui-Shuang
    Morgan, Jessica I. W.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (05) : 1420 - 1430
  • [22] Use of multiple metrics to describe the parafoveal cone mosaic
    Lombardo, Marco
    Serrao, Sebastiano
    Ducoli, Pietro
    Lombardo, Giuseppe
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [23] Visual resolution and cone spacing in the nasal and inferior retina
    Woog, Kelly
    Legras, Richard
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2018, 38 (01) : 66 - 75
  • [24] The relationship between visual resolution and cone spacing in the human fovea
    Rossi, Ethan A.
    Roorda, Austin
    NATURE NEUROSCIENCE, 2010, 13 (02) : 156 - 157
  • [25] Assessment of parafoveal cone density in patients taking hydroxychloroquine in the absence of clinically documented retinal toxicity
    Debellemaniere, Guillaume
    Flores, Mathieu
    Tumahai, Perle
    Meillat, Mathieu
    Garnier, Melanie Bidaut
    Delbosc, Bernard
    Saleh, Maher
    ACTA OPHTHALMOLOGICA, 2015, 93 (07) : E534 - E540
  • [26] Parafoveal cone metrics and their relationship with multifocal electroretinogram
    Ziccardi, Lucia
    Lombardo, Giuseppe
    Parisi, Vincenzo
    Serrao, Sebastiano
    Lombardo, Marco
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [27] Reliability and Agreement Between Metrics of Cone Spacing in Adaptive Optics Images of the Human Retinal Photoreceptor Mosaic
    Giannini, Daniela
    Lombardo, Giuseppe
    Mariotti, Letizia
    Devaney, Nicholas
    Serrao, Sebastiano
    Lombardo, Marco
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (07) : 3127 - 3137
  • [28] Relationships Between Spatial Contrast Sensitivity and Parafoveal Cone Density in Normal Subjects and Patients With Retinal Degeneration
    Hirota, Masakazu
    Morimoto, Takeshi
    Kanda, Hiroyuki
    Lohmann, Tibor Karl
    Miyagawa, Suguru
    Endo, Takao
    Miyoshi, Tomomitsu
    Fujikado, Takashi
    OPHTHALMIC SURGERY LASERS & IMAGING RETINA, 2017, 48 (02) : 106 - 113
  • [29] Assessment of Different Sampling Methods for Measuring and Representing Macular Cone Density Using Flood-Illuminated Adaptive Optics
    Feng, Shu
    Gale, Michael J.
    Fay, Jonathan D.
    Faridi, Ambar
    Titus, Hope E.
    Garg, Anupam K.
    Michaels, Keith V.
    Erker, Laura R.
    Peters, Dawn
    Smith, Travis B.
    Pennesi, Mark E.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (10) : 5751 - 5763
  • [30] The effect of sampling window size on topographical maps of foveal cone density
    Warr, Emma
    Grieshop, Jenna
    Cooper, Robert F.
    Carroll, Joseph
    FRONTIERS IN OPHTHALMOLOGY, 2024, 4