摘要
A simple analytical method sufficient for the determination of Na, K, Ca, Mg, Al, B, Ba, Cd, Co, Cr, Cu, Fe, Li, Mn, Mo, Ni, Pb, Sr, V and Zn at mg kg(-1) levels in pure NaOH, KOH, NaCl and KCl using the optical emission spectrometry was developed. The results of direct determination with multi-elemental aqueous standards were compared with them obtained by the internal standardisation, by the standard addition methods and with the maximal allowable contents of above-mentioned elements in pure chemicals. The method was shown to be very sensitive and exhibits following limits of detection: Na 0.90, K 3.0, Ca 0.064, Mg 0.026, Al 0.43, B 0.13, Ba 0.015, Cd 0.023, Co 0.056, Cr 0.041, Cu 0.063, Fe 0.060, Li 0.017, Mn 0.035, Mo 0.19, Ni 0.055, Pb 0.39, Sr 0.030, V 0.065 and Zn 0.043 (all in mg kg(-1)). The method presents a satisfactory precision (relative standard deviation 4-11 %), high analytical recoveries, linear responses of at least four orders of magnitude, accuracy and low contamination susceptibility. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:208 / 213
页数:6
相关论文
共 24 条
[1]   Determination of metals in highly saline matrices by solid-phase extraction and slurry-sampling inductively coupled plasma-atomic emission spectrometry [J].
Abollino, O ;
Aceto, M ;
Bruzzoniti, MC ;
Mentasti, E ;
Sarzanini, C .
ANALYTICA CHIMICA ACTA, 1998, 375 (03) :293-298
[2]   Correction for non-spectroscopic matrix effects in inductively coupled plasma-atomic emission spectroscopy by internal standardization using spectral lines of the same analyte [J].
Al-Ammar, AS ;
Barnes, RM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (11) :1583-1593
[3]   Determination of trace metals in seawater by inductively coupled plasma mass spectrometry after off-line dithiocarbamate solvent extraction [J].
Batterham, GJ ;
Munksgaard, NC ;
Parry, DL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (11) :1277-1280
[4]   Determination of lead in seawater by inductively coupled plasma optical emission spectrometry after separation and pre-concentration with cocrystallized naphthalene alizarin [J].
Bispo, MS ;
Korn, MDA ;
Morte, ESD ;
Teixeira, LSG .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (12) :2175-2180
[5]   Direct determination of vanadium in high saline produced waters from offshore petroleum exploration by electrothermal atomic absorption spectrometry [J].
Cassella, RJ ;
Oliveira, EP ;
Magalhaes, OIB .
TALANTA, 2006, 69 (01) :48-54
[6]   Empirical procedure for the reduction of mixed-matrix effects in inductively coupled plasma atomic emission spectrometry using an internal standard and proportional correction [J].
deBoer, JLM ;
Velterop, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1996, 356 (06) :362-370
[7]   Determination of heavy metals by inductively coupled plasma mass spectrometry after on-line separation and preconcentration [J].
Dressler, VL ;
Pozebon, D ;
Curtius, AJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (11) :1527-1539
[8]   Nickel determination in saline matrices by ICP-AES after sorption on Amberlite XAD-2 loaded with PAN [J].
Ferreira, SLC ;
de Brito, CF ;
Dantas, AF ;
de Araújo, NML ;
Costa, ACS .
TALANTA, 1999, 48 (05) :1173-1177
[9]   Study of matrix interferences in trace element analysis of environmental samples by inductively coupled plasma atomic emission spectrometry with ultrasonic nebulization [J].
Hoenig, M ;
Docekalova, H ;
Baeten, H .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (03) :195-199
[10]   Group separation of trace rare-earth elements by countercurrent chromatography for their determination in high-purity calcium chloride [J].
Ignatova, SN ;
Maryutina, TA ;
Spivakov, BY ;
Karandashev, VK .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 370 (08) :1109-1113