Effect of Multileaf Collimator Leaf Position Error Determined by Picket Fence Test on Gamma Index Value in Patient-Specific Quality Assurance of Volumetric-Modulated Arc Therapy Plans

被引:3
作者
Ceylan, Cemile [1 ,2 ]
Inal, Serpil Yondem [3 ]
Senol, Elif [3 ]
Yilmaz, Berrin [1 ]
Sahin, Sevim [4 ]
机构
[1] Istanbul Oncol Hosp, Radiat Oncol Dept, Istanbul, Turkey
[2] Yeditepe Univ, Hlth Sci Inst, Istanbul, Turkey
[3] Mem Bahcelievler Hosp, Radiat Oncol Dept, Istanbul, Turkey
[4] Fenerbahce Univ, Med Imaging Dept, Istanbul, Turkey
关键词
mlc error; patient-specific quality assurance; gamma-index evaluation; DYNAMIC IMRT; SLIDING WINDOW; MLC; ACCURACY; SENSITIVITY; PERFORMANCE; DELIVERY; CONSEQUENCES;
D O I
10.7759/cureus.12684
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim The correlation between the MLC QA (IBA Dosimetry, Germany) results of the picket fence test created with intentional errors and the patient's quality assurance (QA) evaluation was investigated to assess the impact of multileaf collimator (MLC) positioning error on patient QA. Materials and methods The picket fence, including error-free and intentional MLC errors, defined in Bank In, Bank Out, and Bank Both were analyzed using MLC QA. The QA of 15 plans consisting of stereotactic radiosurgery (SRS), stereotactic body radiotherapy (SBRT), and conventionally fractionated volumetric-modulated arc therapy (VMAT) acquired with electronic portal imaging devices (EPID) was evaluated in the presence of error-free and MLC errors. The QA of plans were analyzed with 2%/2 mm and 3%/3 mm criteria. Results The passing rates of the picket fence test were 97%, 92%, 91%, and 87% for error-free and intentional errors. The criterion of 3%/3 mm wasn't able to detect an MLC error for either SRS/SBRT or conventionally fractionated VMAT. The criterion of 2%/2mm was more sensitive to detect MLC error for the conventionally fractionated VMAT than SRS/SBRT. While only two of SBRT plans had <90%, four of conventionally fractionated VMAT plans had a <90% passing rate. Conclusion We found that the systematic MLC positioning errors defined with picket fence have a smaller but measurable impact on SRS/SBRT than the VMAT plan for a conventionally fractionated and relatively complex plan such as head and neck and endometrium cases.
引用
收藏
页数:13
相关论文
共 30 条
[1]   Evaluating the dosimetric consequences of MLC leaf positioning errors in dynamic IMRT treatments [J].
Agarwal, Arpita ;
Rastogi, Nikhil ;
Das, K. J. Maria ;
Yoganathan, S. A. ;
Udayakumar, D. ;
Naresh, R. ;
Kumar, Shaleen .
JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (03) :225-231
[2]   Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries [J].
Agnew, A. ;
Agnew, C. E. ;
Grattan, M. W. D. ;
Hounsell, A. R. ;
McGarry, C. K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09) :N49-N63
[3]  
Alber M, 2008, ESTRO BOOKLET, P89
[4]   Evaluation of the ability of a 2D ionisation chamber array and an EPID to detect systematic delivery errors in IMRT plans [J].
Bawazeer, Omemh ;
Gray, Alison ;
Arumugam, Sankar ;
Vial, Philip ;
Thwaites, David ;
Descallar, Joseph ;
Holloway, Lois .
XVII INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY (ICCR 2013), 2014, 489
[5]   Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator [J].
Bedford, James L. ;
Thomas, Michael D. R. ;
Smyth, Gregory .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :172-185
[6]   Requirements for leaf position accuracy for dynamic multileaf collimation [J].
Budgell, GJ ;
Mott, JHL ;
Williams, PC ;
Brown, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1211-1227
[7]   Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning system [J].
Cadman, P ;
Bassalow, R ;
Sidhu, NPS ;
Ibbott, G ;
Nelson, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16) :3001-3010
[8]   Automatic detection of MLC relative position errors for VMAT using the EPID-based picket fence test [J].
Christophides, Damianos ;
Davies, Alex ;
Fleckney, Mark .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (23) :8340-8359
[9]   Testing of dynamic multileaf collimation [J].
Chui, CS ;
Spirou, S ;
LoSasso, T .
MEDICAL PHYSICS, 1996, 23 (05) :635-641
[10]   Determination of the optimal tolerance for MLC positioning in sliding window and VMAT techniques [J].
Hernandez, V. ;
Abella, R. ;
Calvo, J. F. ;
Jurado-Bruggemann, D. ;
Sancho, I. ;
Carrasco, P. .
MEDICAL PHYSICS, 2015, 42 (04) :1911-1916