Transtracheal ventilation with a novel ejector-based device (Ventrain) in open, partly obstructed, or totally closed upper airways in pigs

被引:32
作者
Paxian, M. [1 ]
Preussler, N. P. [1 ]
Reinz, T. [1 ]
Schlueter, A. [1 ]
Gottschall, R. [1 ]
机构
[1] Jena Univ Hosp, Dept Anaesthesiol & Intens Care Med, D-07747 Jena, Germany
关键词
airway obstruction; equipment; ventilator; ventilation; transtracheal; JET VENTILATION; STRATEGIES; MODEL; COMPLICATIONS; MANAGEMENT; CATHETER; INTUBATE;
D O I
10.1093/bja/aev200
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Transtracheal access and subsequent jet ventilation are among the last options in a 'cannot intubate-cannot oxygenate' scenario. These interventions may lead to hypercapnia, barotrauma, and haemodynamic failure in the event of an obstructed upper airway. The aim of the present study was to evaluate the efficacy and the haemodynamic effects of the Ventrain, a manually operated ventilation device that provides expiratory ventilation assistance. Transtracheal ventilation was carried out with the Ventrain in different airway scenarios in live pigs, and its performance was compared with a conventional jet ventilator. Methods: Pigs with open, partly obstructed, or completely closed upper airways were transtracheally ventilated either with the Ventrain or by conventional jet ventilation. Airway pressures, haemodynamic parameters, and blood gases obtained in the different settings were compared. Results: Mean (SD) alveolar minute ventilation as reflected by arterial partial pressure of CO2 was superior with the Ventrain in partly obstructed airways after 6 min in comparison with traditional manual jet ventilation [4.7 (0.19) compared with 7.1 (0.37) kPa], and this was also the case in all simulated airway conditions. At the same time, peak airway pressures were significantly lower and haemodynamic parameters were altered to a lesser extent with the Ventrain. Conclusions: The results of this study suggest that the Ventrain device can ensure sufficient oxygenation and ventilation through a small-bore transtracheal catheter when the airway is open, partly obstructed, or completely closed. Minute ventilation and avoidance of high airway pressures were superior in comparison with traditional hand-triggered jet ventilation, particularly in the event of complete upper airway obstruction.
引用
收藏
页码:308 / 316
页数:9
相关论文
共 17 条
[1]   Percutaneous transtracheal ventilation in an obstructed airway model in post-apnoeic sheep [J].
Berry, M. ;
Tzeng, Y. ;
Marsland, C. .
BRITISH JOURNAL OF ANAESTHESIA, 2014, 113 (06) :1039-1045
[2]   Major complications of airway management in the UK: results of the Fourth National Audit Project of the Royal College of Anaesthetists and the Difficult Airway Society. Part 1: Anaesthesia [J].
Cook, T. M. ;
Woodall, N. ;
Frerk, C. .
BRITISH JOURNAL OF ANAESTHESIA, 2011, 106 (05) :617-631
[3]   2 CASES OF BAROTRAUMA ASSOCIATED WITH TRANSTRACHEAL JET VENTILATION [J].
CRAFT, TM ;
CHAMBERS, PH ;
WARD, ME ;
GOAT, VA .
BRITISH JOURNAL OF ANAESTHESIA, 1990, 64 (04) :524-527
[4]   Difficult airway management practice patterns among anesthesiologists practicing in the United States: Have we made any progress? [J].
Ezri, T ;
Szmuk, P ;
Warters, RD ;
Katz, J ;
Hagberg, CA .
JOURNAL OF CLINICAL ANESTHESIA, 2003, 15 (06) :418-422
[5]   Comparison of different methods of ventilation via cannula cricothyroidotomy in a trachea-lung model† [J].
Flint, N. J. ;
Russell, W. C. ;
Thompson, J. P. .
BRITISH JOURNAL OF ANAESTHESIA, 2009, 103 (06) :891-895
[6]   Jet ventilation in upper airway obstruction: Description and model lung testing of a new jetting device [J].
Garry, B ;
Woo, P ;
Perrault, DF ;
Shapshay, SM ;
Wurm, WH .
ANESTHESIA AND ANALGESIA, 1998, 87 (04) :915-920
[7]   Equipment and strategies for emergency tracheal access in the adult patient [J].
Hamaekers, A. E. ;
Henderson, J. J. .
ANAESTHESIA, 2011, 66 :65-80
[8]   Ventrain: an ejector ventilator for emergency use [J].
Hamaekers, A. E. W. ;
Borg, P. A. J. ;
Enk, D. .
BRITISH JOURNAL OF ANAESTHESIA, 2012, 108 (06) :1017-1021
[9]   Ventilation through a small-bore catheter: optimizing expiratory ventilation assistance [J].
Hamaekers, A. E. W. ;
Borg, P. A. J. ;
Goetz, T. ;
Enk, D. .
BRITISH JOURNAL OF ANAESTHESIA, 2011, 106 (03) :403-409
[10]   Achieving an adequate minute volume through a 2 mm transtracheal catheter in simulated upper airway obstruction using a modified industrial ejector [J].
Hamaekers, A. E. W. ;
Goetz, T. ;
Borg, P. A. J. ;
Enk, D. .
BRITISH JOURNAL OF ANAESTHESIA, 2010, 104 (03) :382-386