Simultaneous quantitative determination of cyclosporine A and its three main metabolites (AM1, AM4N and AM9) in human blood by liquid chromatography/mass spectrometry using a rapid sample processing method

被引:17
作者
Koseki, N [1 ]
Nakashima, A [1 ]
Nagae, Y [1 ]
Masuda, N [1 ]
机构
[1] Novartis Pharma KK, Tsukuba Res Inst, Drug Metab & Pharmacokinet, Tsukuba, Ibaraki 3002611, Japan
关键词
D O I
10.1002/rcm.2358
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We have developed a sensitive and specific liquid chromatography/mass spectrometry (LC/MS) method for the simultaneous determination of cyclosporine A (CsA) and its three main metabolites (AM1, AM4N and AM9) in human blood. Following protein precipitation, supernatant was directly injected into the LC/MS system. Chromatographic separation was accomplished on a Symmetry C8 (4.6 X 75 mm, 3.5 mu m) column with a linear gradient elution prior to detection by atmospheric pressure chemical ionization (APCI) MS using selected ion monitoring (SIM) in positive mode. This method can be applied to single mass equipment. The analytical range for each analyte was set at 1-2500 ng/mL using 100 mu L of blood sample. The analytical method was fully validated according to FDA guidance. Intra-day mean accuracy and precision were 95.2-113.5% and 0.9-8.9%, respectively. Inter-day mean accuracy and precision were 95.8-107.0% and 1.5-10.7%, respectively. In blood all analytes were stable during three freeze/thaw cycles, for 24 h at room temperature and for 12 months at or below -15 degrees C. Stability was also confirmed in processed samples for 24 h at 10 degrees C and for 6 months at VC in methanol. In addition, we confirmed the method could avoid matrix effects from transplant subjects' samples. This LC/MS technique provided an excellent method for simultaneous quantitative determination of CsA and its three metabolites for evaluation of their pharmacokinetic profiles. Copyright (c) 2006 John Wiley & Sons, Ltd.
引用
收藏
页码:733 / 740
页数:8
相关论文
共 39 条
[1]   Simple determination of cyclosporine in human whole blood by high-performance liquid chromatography [J].
Amini, H ;
Ahmadiani, A .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 795 (02) :209-214
[2]  
[Anonymous], GUID IND BIOAN METH
[3]   Alterations in cyclosporin A pharmacokinetics and metabolism during treatment with St John's wort in renal transplant patients [J].
Bauer, S ;
Störmer, E ;
Johne, A ;
Krüger, H ;
Budde, K ;
Neumayer, HH ;
Roots, I ;
Mai, I .
BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 2003, 55 (02) :203-211
[4]   High-performance liquid chromatographic method for therapeutic drug monitoring of cyclosporine A and its two metabolites in renal transplant patients [J].
Brozmanová, H ;
Grundmann, M ;
Safarcik, K ;
Jegorov, A .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2000, 749 (01) :93-100
[5]  
CARRUTHERS SG, 1983, CLIN CHEM, V29, P180
[6]   High-performance liquid chromatographic analysis of cyclosporin A in rat blood and liver using a commercially available internal standard [J].
Chimalakonda, AP ;
Shah, RB ;
Mehvar, R .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 772 (01) :107-114
[7]   Automated, fast and sensitive quantification of drugs in blood by liquid chromatography-mass spectrometry with on-line extraction: immunosuppressants [J].
Christians, U ;
Jacobsen, W ;
Serkova, N ;
Benet, LZ ;
Vidal, C ;
Sewing, KF ;
Manns, MP ;
Kirchner, GI .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2000, 748 (01) :41-53
[8]   CYCLOSPORINE METABOLISM IN TRANSPLANT PATIENTS [J].
CHRISTIANS, U ;
SEWING, KF .
PHARMACOLOGY & THERAPEUTICS, 1993, 57 (2-3) :291-345
[9]   Liquid chromatography/mass spectrometry for therapeutic drug monitoring of immunosuppressants [J].
Deters, M ;
Kaever, V ;
Kirchner, GI .
ANALYTICA CHIMICA ACTA, 2003, 492 (1-2) :133-145
[10]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC COLUMN-SWITCHING METHOD FOR 2 CYCLOSPORINE METABOLITES IN BLOOD [J].
GMUR, DJ ;
MEIER, P ;
YEE, GC .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 425 (02) :343-352