Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers

被引:253
作者
Grulkowski, Ireneusz [1 ,2 ]
Liu, Jonathan J. [1 ,2 ]
Potsaid, Benjamin [1 ,2 ,3 ]
Jayaraman, Vijaysekhar [4 ]
Lu, Chen D. [1 ,2 ]
Jiang, James [3 ]
Cable, Alex E. [3 ]
Duker, Jay S. [5 ]
Fujimoto, James G. [1 ,2 ]
机构
[1] MIT, Dept Elect Engn & Comp Sci, Cambridge, MA 02139 USA
[2] MIT, Elect Res Lab, Cambridge, MA 02139 USA
[3] Thorlabs Inc, Adv Imaging Grp, Newton, NJ USA
[4] Praevium Res Inc, Santa Barbara, CA USA
[5] Tufts Univ, New England Eye Ctr, Boston, MA 02111 USA
基金
美国国家卫生研究院;
关键词
OPTICAL COHERENCE TOMOGRAPHY; FREQUENCY-DOMAIN REFLECTOMETRY; IN-VIVO; MEGAHERTZ OCT; AXIAL SCANS; 200; KHZ;
D O I
10.1364/BOE.3.002733
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We demonstrate swept source OCT utilizing vertical-cavity surface emitting laser (VCSEL) technology for in vivo high speed retinal, anterior segment and full eye imaging. The MEMS tunable VCSEL enables long coherence length, adjustable spectral sweep range and adjustable high sweeping rate (50-580 kHz axial scan rate). These features enable integration of multiple ophthalmic applications into one instrument. The operating modes of the device include: ultrahigh speed, high resolution retinal imaging (up to 580 kHz); high speed, long depth range anterior segment imaging (100 kHz) and ultralong range full eye imaging (50 kHz). High speed imaging enables wide-field retinal scanning, while increased light penetration at 1060 nm enables visualization of choroidal vasculature. Comprehensive volumetric data sets of the anterior segment from the cornea to posterior crystalline lens surface are also shown. The adjustable VCSEL sweep range and rate make it possible to achieve an extremely long imaging depth range of similar to 50 mm, and to demonstrate the first in vivo 3D OCT imaging spanning the entire eye for non-contact measurement of intraocular distances including axial eye length. Swept source OCT with VCSEL technology may be attractive for next generation integrated ophthalmic OCT instruments. (C) 2012 Optical Society of America
引用
收藏
页码:2733 / 2751
页数:19
相关论文
共 51 条
[21]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[22]   Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomography [J].
Huber, R ;
Wojtkowski, M ;
Fujimoto, JG .
OPTICS EXPRESS, 2006, 14 (08) :3225-3237
[23]   Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging: design and scaling principles [J].
Huber, R ;
Wojtkowski, M ;
Taira, K ;
Fujimoto, JG ;
Hsu, K .
OPTICS EXPRESS, 2005, 13 (09) :3513-3528
[25]   MICROMETER-SCALE RESOLUTION IMAGING OF THE ANTERIOR EYE IN-VIVO WITH OPTICAL COHERENCE TOMOGRAPHY [J].
IZATT, JA ;
HEE, MR ;
SWANSON, EA ;
LIN, CP ;
HUANG, D ;
SCHUMAN, JS ;
PULIAFITO, CA ;
FUJIMOTO, JG .
ARCHIVES OF OPHTHALMOLOGY, 1994, 112 (12) :1584-1589
[26]  
Jayaraman V, 2011, CONF LASER ELECTR
[27]   Extended in vivo anterior eye-segment imaging with full-range complex spectral domain optical coherence tomography [J].
Jungwirth, Johannes ;
Baumann, Bernhard ;
Pircher, Michael ;
Goetzinger, Erich ;
Hitzenberger, Christoph K. .
JOURNAL OF BIOMEDICAL OPTICS, 2009, 14 (05)
[28]   Spectral optical coherence tomography: a new imaging technique in contact lens practice [J].
Kaluzny, BJ ;
Kaluzny, JJ ;
Szkulmowska, A ;
Gorczynska, I ;
Szkulmowski, M ;
Bajraszewski, T ;
Targowski, P ;
Kowalczyk, A .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2006, 26 (02) :127-132
[29]   Megahertz OCT for ultrawide-field retinal imaging with a 1050nm Fourier domain mode-locked laser [J].
Klein, Thomas ;
Wieser, Wolfgang ;
Eigenwillig, Christoph M. ;
Biedermann, Benjamin R. ;
Huber, Robert .
OPTICS EXPRESS, 2011, 19 (04) :3044-3062
[30]   Performance of fourier domain vs. time domain optical coherence tomography [J].
Leitgeb, R ;
Hitzenberger, CK ;
Fercher, AF .
OPTICS EXPRESS, 2003, 11 (08) :889-894