Metal artifact reduction in MRI-based cervical cancer intracavitary brachytherapy

被引:10
作者
Rao, Yuan James [1 ]
Zoberi, Jacqueline E. [1 ]
Kadbi, Mo [2 ]
Grigsby, Perry W. [1 ]
Cammin, Jochen [1 ]
Mackey, Stacie L. [3 ]
Garcia-Ramirez, Jose [1 ]
Goddu, S. Murty [1 ]
Schwarz, Julie K. [1 ]
Gach, H. Michael [1 ,4 ]
机构
[1] Washington Univ, Sch Med, Dept Radiat Oncol, 4921 Parkview Pl,Campus Box 8224, St Louis, MO 63110 USA
[2] Philips Healthcare, Cleveland, OH 44143 USA
[3] Barnes Jewish Hosp, St Louis, MO 63110 USA
[4] Washington Univ, Sch Med, Dept Radiol, St Louis, MO 63110 USA
关键词
metal artifact reduction; MRI; brachytherapy; tandem; cervical cancer; MODULATED RADIATION-THERAPY; HIGH-DOSE-RATE; APPLICATOR RECONSTRUCTION; ARTHROPLASTY IMPLANTS; CLINICAL-OUTCOMES; TUMOR VOLUME; RECOMMENDATIONS; SEMAC; PARAMETERS;
D O I
10.1088/1361-6560/62/8/3011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Magnetic resonance imaging (MRI) plays an increasingly important role in brachytherapy planning for cervical cancer. Yet, metal tandem, ovoid intracavitary applicators, and fiducial markers used in brachytherapy cause magnetic susceptibility artifacts in standard MRI. These artifacts may impact the accuracy of brachytherapy treatment and the evaluation of tumor response by misrepresenting the size and location of the metal implant, and distorting the surrounding anatomy and tissue. Metal artifact reduction sequences (MARS) with high bandwidth RF selective excitations and turbo spin-echo readouts were developed for MRI of orthopedic implants. In this study, metal artifact reduction was applied to brachytherapy of cervical cancer using the orthopedic metal artifact reduction (O-MAR) sequence. O-MAR combined MARS features with view angle tilting and slice encoding for metal artifact correction (SEMAC) to minimize in-plane and through-plane susceptibility artifacts. O-MAR improved visualization of the tandem tip on T-2 and proton density weighted (PDW) imaging in phantoms and accurately represented the diameter of the tandem. In a pilot group of cervical cancer patients (N = 7), O-MAR significantly minimized the blooming artifact at the tip of the tandem in PDW MRI. There was no significant difference observed in artifact reduction between the weak (5 kHz, 7 z-phase encodes) and medium (10 kHz, 13 z-phase encodes) SEMAC settings. However, the weak setting allowed a significantly shorter acquisition time than the medium setting. O-MAR also reduced susceptibility artifacts associated with metal fiducial markers so that they appeared on MRI at their true dimensions.
引用
收藏
页码:3011 / 3024
页数:14
相关论文
共 33 条
[1]   TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11
[2]   Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (IV): Basic principles and parameters for MR imaging within the frame of image based adaptive cervix cancer brachytherapy [J].
Dimopoulos, Johannes C. A. ;
Petrow, Peter ;
Tanderup, Kari ;
Petric, Primoz ;
Berger, Daniel ;
Kirisits, Christian ;
Pedersen, Erik M. ;
van Limbergen, Erik ;
Haie-Meder, Christine ;
Poetter, Richard .
RADIOTHERAPY AND ONCOLOGY, 2012, 103 (01) :113-122
[3]   Cervical Gross Tumor Volume Dose Predicts Local Control Using Magnetic Resonance Imaging/Diffusion-Weighted Imaging-Guided High-Dose-Rate and Positron Emission Tomography/Computed Tomography-Guided Intensity Modulated Radiation Therapy [J].
Dyk, Pawel ;
Jiang, Naomi ;
Sun, Baozhou ;
DeWees, Todd A. ;
Fowler, Kathryn J. ;
Narra, Vamsi ;
Garcia-Ramirez, Jose L. ;
Schwarz, Julie K. ;
Grigsby, Perry W. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 (04) :794-801
[4]   Comparison of apparent diffusion coefficient maps to T2-weighted images for target delineation in cervix cancer brachytherapy [J].
Esthappan, Jacqueline ;
Ma, Daniel J. ;
Narra, Vamsidhar R. ;
Raptis, Constantine A. ;
Grigsby, Perry W. .
JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2011, 3 (04) :193-198
[5]   A comparison of two methods for measuring the signal to noise ratio on MR images [J].
Firbank, MJ ;
Coulthard, A ;
Harrison, RM ;
Williams, ED .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (12) :N261-N264
[6]   Compressed Sensing SEMAC: 8-fold Accelerated High Resolution Metal Artifact Reduction MRI of Cobalt-Chromium Knee Arthroplasty Implants [J].
Fritz, Jan ;
Ahlawat, Shivani ;
Demehri, Shadpour ;
Thawait, Gaurav K. ;
Raithel, Esther ;
Gilson, Wesley D. ;
Nittka, Mathias .
INVESTIGATIVE RADIOLOGY, 2016, 51 (10) :666-676
[7]   Advanced metal artifact reduction MRI of metal-on-metal hip resurfacing arthroplasty implants: compressed sensing acceleration enables the time-neutral use of SEMAC [J].
Fritz, Jan ;
Fritz, Benjamin ;
Thawait, Gaurav K. ;
Raithel, Esther ;
Gilson, Wesley D. ;
Nittka, Mathias ;
Mont, Michael A. .
SKELETAL RADIOLOGY, 2016, 45 (10) :1345-1356
[8]   Apparent diffusion coefficients in GEC ESTRO target volumes for image guided adaptive brachytherapy of locally advanced cervical cancer [J].
Haack, Soren ;
Pedersen, Erik Morre ;
Jespersen, Sune N. ;
Kallehauge, Jesper F. ;
Lindegaard, Jacob Christian ;
Tanderup, Kari .
ACTA ONCOLOGICA, 2010, 49 (07) :978-983
[9]   Applicator reconstruction in MRI 3D image-based dose planning of brachytherapy for cervical cancer [J].
Haack, Soren ;
Nielsen, Soren Kynde ;
Lindegaard, Jacob Christian ;
Gelineck, John ;
Tanderup, Kari .
RADIOTHERAPY AND ONCOLOGY, 2009, 91 (02) :187-193
[10]   Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group: Considerations and pitfalls in commissioning and applicator reconstruction in 3D image-based treatment planning of cervix cancer brachytherapy [J].
Hellebust, Taran Paulsen ;
Kirisits, Christian ;
Berger, Daniel ;
Perez-Calatayud, Jose ;
De Brabandere, Marisol ;
De Leeuw, Astrid ;
Dumas, Isabelle ;
Hudej, Robert ;
Lowe, Gerry ;
Wills, Rachel ;
Tanderup, Kari .
RADIOTHERAPY AND ONCOLOGY, 2010, 96 (02) :153-160