Average glandular dose in paired digital mammography and digital breast tomosynthesis acquisitions in a population based screening program: effects of measuring breast density, air kerma and beam quality

被引:27
作者
Osteras, Bjorn Helge [1 ,2 ]
Skaane, Per [2 ,3 ]
Gullien, Randi [3 ]
Martinsen, Anne Catrine Traegde [1 ,4 ]
机构
[1] Oslo Univ Hosp, Dept Diagnost Phys, Oslo, Norway
[2] Univ Oslo, Inst Clin Med, Oslo, Norway
[3] Oslo Univ Hosp, Dept Radiol & Nucl Med, Oslo, Norway
[4] Univ Oslo, Inst Phys, Oslo, Norway
关键词
mammography; digital breast tomosynthesis; average glandular dose; breast density; screening; DOSIMETRY PROTOCOLS; COMPUTATION; IMAGES; FFDM; AGE; UK;
D O I
10.1088/1361-6560/aaa614
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The main purpose was to compare average glandular dose (AGD) for same-compression digital mammography (DM) and digital breast tomosynthesis (DBT) acquisitions in a population based screening program, with and without breast density stratification, as determined by automatically calculated breast density (Quantra (TM)). Secondary, to compare AGD estimates based on measured breast density, air kerma and half value layer (HVL) to DICOM metadata based estimates. AGD was estimated for 3819 women participating in the screening trial. All received craniocaudal and mediolateral oblique views of each breasts with paired DM and DBT acquisitions. Exposure parameters were extracted from DICOM metadata. Air kerma and HVL were measured for all beam qualities used to acquire the mammograms. Volumetric breast density was estimated using Quantra T. AGD was estimated using the Dance model. AGD reported directly from the DICOM metadata was also assessed. Mean AGD was 1.74 and 2.10 mGy for DM and DBT, respectively. Mean DBT/DM AGD ratio was 1.24. For fatty breasts: mean AGD was 1.74 and 2.27 mGy for DM and DBT, respectively. For dense breasts: mean AGD was 1.73 and 1.79 mGy, for DM and DBT, respectively. For breasts of similar thickness, dense breasts had higher AGD for DM and similar AGD for DBT. The DBT/DM dose ratio was substantially lower for dense compared to fatty breasts (1.08 versus 1.33). The average c-factor was 1.16. Using previously published polynomials to estimate glandularity from thickness underestimated the c-factor by 5.9% on average. Mean AGD error between estimates based on measurements (air kerma and HVL) versus DICOM header data was 3.8%, but for one mammography unit as high as 7.9%. Mean error of using the AGD value reported in the DICOM header was 10.7 and 13.3%, respectively. Thus, measurement of breast density, radiation dose and beam quality can substantially affect AGD estimates.
引用
收藏
页数:14
相关论文
共 37 条
[1]  
[Anonymous], 2007, 457 IAEA TRS
[2]  
[Anonymous], 2011, IAEA Human Health Series.
[3]  
Baorui Ren, 2012, Breast Imaging. Proceedings 11th International Workshop, IWDM 2012, P775, DOI 10.1007/978-3-642-31271-7_100
[4]   Dosimetric implications of age related glandular changes in screening mammography [J].
Beckett, JR ;
Kotre, CJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (03) :801-813
[5]   Breast cancer screening with tomosynthesis (3D mammography) with acquired or synthetic 2D mammography compared with 2D mammography alone (STORM-2): a population-based prospective study [J].
Bernardi, Daniela ;
Macaskill, Petra ;
Pellegrini, Marco ;
Valentini, Marvi ;
Fanto, Carmine ;
Ostillio, Livio ;
Tuttobene, Paolina ;
Luparia, Andrea ;
Houssami, Nehmat .
LANCET ONCOLOGY, 2016, 17 (08) :1105-1113
[6]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[7]   Average glandular dose in digital mammography and digital breast tomosynthesis: comparison of phantom and patient data [J].
Bouwman, R. W. ;
van Engen, R. E. ;
Young, K. C. ;
den Heeten, G. J. ;
Broeders, M. J. M. ;
Schopphoven, S. ;
Jeukens, C. R. L. P. N. ;
Veldkamp, W. J. H. ;
Dance, D. R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (20) :7893-7907
[8]  
Carney PA, 2003, ANN INTERN MED, V138, P168, DOI 10.7326/0003-4819-138-3-200302040-00008
[9]   Automated Breast Density Computation in Digital Mammography and Digital Breast Tomosynthesis: Influence on Mean Glandular Dose and BIRADS Density Categorization [J].
Castillo-Garcia, Maria ;
Chevalier, Margarita ;
Garayoa, Julia ;
Rodriguez-Ruiz, Alejandro ;
Garcia-Pinto, Diego ;
Valverde, Julio .
ACADEMIC RADIOLOGY, 2017, 24 (07) :802-810
[10]   In vivo average glandular dose evaluation: one-to-one comparison between digital breast tomosynthesis and full-field digital mammography [J].
Cavagnetto, Francesca ;
Taccini, Gianni ;
Rosasco, Raffaella ;
Bampi, Rossana ;
Calabrese, Massimo ;
Tagliafico, Alberto .
RADIATION PROTECTION DOSIMETRY, 2013, 157 (01) :53-61