Optimizing the sensitivity and radiological properties of the PRESAGE® dosimeter using metal compounds

被引:21
作者
Alqathami, Mamdooh [1 ]
Blencowe, Anton [2 ]
Qiao, Greg [2 ]
Adamovics, John [3 ]
Geso, Moshi [1 ]
机构
[1] Univ Melbourne, Sch Med Sci, RMIT, Bundoora, Vic 3083, Australia
[2] Univ Melbourne, Dept Chem & Biomol Engn, Melbourne, Vic 3052, Australia
[3] Rider Univ, Dept Biol & Chem, Lawrenceville, NJ 08648 USA
关键词
Metal compounds; Dosimetry; Radiation; SOLID POLYMER DOSIMETER; LEUCO DYES;
D O I
10.1016/j.radphyschem.2012.06.004
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The aim of this study is to investigate the radiation-modifying effects of incorporating commercially available bismuth-, tin- and zinc-based compounds in the composition of the PRESAGE (R) dosimeter, and the feasibility of employing such compounds for radiation dose enhancement. Furthermore, we demonstrate that metal compounds can be included in the formulation to yield water-equivalent PRESAGE (R) dosimeters with enhanced dose response. Various concentrations of the metal compounds were added to a newly developed PRESAGE (R) formulation and the resulting dosimeters were irradiated with 100 kV and 6 MV photon beams. A comparison between sensitivity and radiological properties of the PRESAGE (R) dosimeters with and without the addition of metal compounds was carried out. Optical density changes of the dosimeters before and after irradiation were measured using a spectro-photometer. In general, when metal compounds were incorporated in the composition of the PRESAGE (R) dosimeter, the sensitivity of the dosimeters to radiation dose increased depending on the type and concentration of the metal compound, with the bismuth compound showing the highest dose enhancement factor. In addition, these metal compounds were also shown to improve the retention of the post-response absorption value of the PRESAGE (R) dosimeter over a period of 2 weeks. Thus, incorporating 1-3 mM (ca. 0.2 wt%) of any of the three investigated metal compounds in the composition of the PRESAGE (R) dosimeter is found to be an efficient way to enhance the sensitivity of the dosimeter to radiation dose and stabilize its post-response for longer times. Furthermore, the addition of small amounts of the metal compounds also accelerates the polymerization of the PRESAGE (R) dosimeter precursors, significantly reducing the fabrication time. Finally, a novel water-equivalent PRESAGE (R) dosimeter formula optimized with metal compounds is proposed for clinical use in both kilovoltage and megavoltage radiotherapy dosimetry. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1688 / 1695
页数:8
相关论文
共 33 条
[1]   Characterisation of PRESAGE™:: A new 3-D radiochromic solid polymer dosemeter for ionising radiation [J].
Adamovics, J. ;
Maryanski, M. J. .
RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) :107-112
[2]  
Adamovics J, 2003, MED PHYS, V30, P1349
[3]  
Adamovics J., 2006, Journal of Physics: Conference Series, V56, P172, DOI 10.1088/1742-6596/56/1/020
[4]  
ADAMOVICS J, 2006, Patent No. 7098463
[5]   Optimization of the sensitivity and stability of the PRESAGE™ dosimeter using trihalomethane radical initiators [J].
Alqathami, Mamdooh ;
Blencowe, Anton ;
Qiao, Greg ;
Butler, Duncan ;
Geso, Moshi .
RADIATION PHYSICS AND CHEMISTRY, 2012, 81 (07) :867-873
[6]   Polymer gel dosimetry [J].
Baldock, C. ;
De Deene, Y. ;
Doran, S. ;
Ibbott, G. ;
Jirasek, A. ;
Lepage, M. ;
McAuley, K. B. ;
Oldham, M. ;
Schreiner, L. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :R1-R63
[7]   Cancer and Radiation Therapy: Current Advances and Future Directions [J].
Baskar, Rajamanickam ;
Lee, Kuo Ann ;
Yeo, Richard ;
Yeoh, Kheng-Wei .
INTERNATIONAL JOURNAL OF MEDICAL SCIENCES, 2012, 9 (03) :193-199
[8]   Catalysis of the isocyanate-hydroxyl reaction by non-tin catalysts [J].
Blank, WJ ;
He, ZA ;
Hessell, ET .
PROGRESS IN ORGANIC COATINGS, 1999, 35 (1-4) :19-29
[9]   Investigation into the feasibility of using PRESAGE™/optical-CT dosimetry for the verification of gating treatments [J].
Brady, Samuel L. ;
Brown, William E. ;
Clift, Corey G. ;
Yoo, Sua ;
Oldham, Mark .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (08) :2187-2201
[10]   Radiological properties of the PRESAGE and PAGAT polymer dosimeters [J].
Brown, S. ;
Venning, A. ;
De Deene, Y. ;
Vial, P. ;
Oliver, L. ;
Adamovics, J. ;
Baldock, C. .
APPLIED RADIATION AND ISOTOPES, 2008, 66 (12) :1970-1974