Monte Carlo calculations of the ionization chamber wall correction factors for 192Ir and 60Co gamma rays and 250 kV x-rays for use in calibration of 192Ir HDR brachytherapy sources

被引:13
作者
Ferreira, IH
de Almeida, CE
Marre, D
Marechal, MH
Bridier, A
Chavaudra, J
机构
[1] Inst Gustave Roussy, Serv Phys, F-94805 Villejuif, France
[2] Univ Estado Rio de Janeiro, Lab Ciencias Radiol, BR-20550013 Rio De Janeiro, Brazil
关键词
D O I
10.1088/0031-9155/44/8/304
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As in the method for the calibration of Ir-192 high-dose-rate (HDR) brachytherapy sources, the ionization chamber wall correction factor A(w) is needed for Ir-192 and Co-60 gamma rays and 250 kV x-rays. This factor takes into account the variation in chamber response due to the attenuation of the photon beam in the chamber wall and build-up cap and the contribution of scattered photons. Monte Carlo calculations were performed using the ESG4 code system with the PRESTA algorithm, to calculate the A(w) factor for 51 commercial ionization chambers and build-up caps exposed to the typical energy spectrum of Ir-192 and 60Co gamma rays and 250 kV x-rays. The calculated A(w) correction factors for Ir-192 and 60Co sources and 250 kV x-rays agree very well to within 0.1% with published experimental data (the statistical uncertainty is less than 0.1% of the calculated correction factor value). For the Ir-192, sources, A(w) varies from 0.973 to 0.943 and for the 250 kV x-rays the minimum value of A(w) for all chambers studied is 0.983. The calculated A(w) correction factors can be used to calculate the air kerma calibration factor of HDR brachytherapy sources, when interpolative methods are considered, contributing to the reduction in the overall uncertainties in the calibration procedure.
引用
收藏
页码:1897 / 1904
页数:8
相关论文
共 21 条
[1]   CHAMBER-DEPENDENT WALL CORRECTION FACTORS IN DOSIMETRY [J].
ANDREO, P ;
NAHUM, A ;
BRAHME, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (11) :1189-1199
[2]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[3]  
Duane S., 1989, PIRS0173 NRCC PIRS0173 NRCC
[4]  
EZZELL G, 1989, BRACHYTHERAPY, V2, P61
[5]   Perturbation corrections for flat and thimble-type cylindrical standard ionization chambers for 60Co gamma rays:: Monte Carlo calculations [J].
Ferreira, IH ;
Bridier, A ;
de Almeida, CE ;
Marre, D ;
Chavaudra, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :2721-2727
[6]   CALIBRATION OF IR-192 HIGH-DOSE-RATE AFTERLOADING SYSTEMS [J].
GOETSCH, SJ ;
ATTIX, FH ;
PEARSON, DW ;
THOMADSEN, BR .
MEDICAL PHYSICS, 1991, 18 (03) :462-467
[7]  
*IAEA, 1987, 277 IAEA
[8]  
*IAEA, 1997, 381 IAEA
[9]  
ICRU, 1984, 37 ICRU
[10]  
International Atomic Energy Agency, 1999, 1079 IAEA TECDOC