Evaluation of Surface Finish and Dimensional Control of Tribological Metal-Ultra High Molecular Weight Polyethylene Pair of Commercially Available Hip Implants

被引:12
作者
Cubillos, Patricia O. [1 ]
dos Santos, Vinicius O. [1 ]
Pizzolatti, Andre Luiz A. [1 ]
da Rosa, Edison [1 ]
Roesler, Carlos R. M. [1 ]
机构
[1] Univ Fed Santa Catarina, Dept Mech Engn, Univ Hosp, Lab Biomech Engn LEBm, Florianopolis, SC, Brazil
关键词
femoral head; acetabular liner; wear rate; osteolysis; dimensional analysis; surface finish; UHMWPE ACETABULAR CUPS; CONTACT PRESSURE; BEARING SURFACE; FEMORAL HEADS; WEAR; SIMULATOR; PROSTHESES; ARTHROPLASTY; ROUGHNESS; THICKNESS;
D O I
10.1016/j.arth.2017.10.032
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Dimensional control and surface finish of the femoral head and acetabular liner are critical factors in the manufacturing process due to the risk of increased polyethylene wear, which is the primary cause of aseptic failure of a metal/polymer hip prosthesis. The aim of this study is to perform a dimensional and surface finish analysis to evaluate the reproduction and accuracy of the manufacturing processes of metal femoral heads and ultra high molecular weight polyethylene acetabular liners. Methods: Four femoral heads and acetabular liners from 5 manufacturers were evaluated. The methods of evaluation followed the standards ISO 7206-2:2011 and ISO 21535:2010. Results: The diameter, sphericity, and roughness of the femoral heads from all the manufacturers were in accordance with the standard requirements. Only the sphericity showed a lack of repeatability among the manufacturers. The variability in sphericity was high among some manufacturers and low in others. The diameters of the acetabular liners of 2 manufacturers were not in accordance with the standard requirements. The repeatability of sphericity, thickness, and roughness of the acetabular liners were heterogeneous among the manufacturers, which means that some manufacturers need to improve quality control. Conclusion: Our results showed a good dimensional and surface finish control of the manufacturing processes of the femoral heads. However, the same control was not shown during the manufacturing of the acetabular liners although all samples had a thickness and sphericity in accordance with the standard. A better quality control of the manufacturing process of ultra high molecular weight polyethylene acetabular liners should be made to improve the dimensional parameters of the acetabular liners and the tribological pair. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:939 / 944
页数:6
相关论文
共 35 条
[11]   The sphericity in total of the bearing surface hip arthroplasty [J].
Ito, H ;
Minami, A ;
Matsuno, T ;
Tanino, H ;
Yuhta, T ;
Nishimura, I .
JOURNAL OF ARTHROPLASTY, 2001, 16 (08) :1024-1029
[12]   Reduction of polyethylene wear by concave dimples on the frictional surface in artificial hip joints [J].
Ito, H ;
Kaneda, K ;
Yuhta, T ;
Nishimura, I ;
Yasuda, K ;
Matsuno, T .
JOURNAL OF ARTHROPLASTY, 2000, 15 (03) :332-338
[13]   Effect of head surface roughness and sterilization on wear of UHMWPE acetabular cups [J].
Jedenmalm, A. ;
Affatato, S. ;
Taddei, P. ;
Leardini, W. ;
Gedde, U. W. ;
Fagnano, C. ;
Viceconti, M. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2009, 90A (04) :1032-1042
[14]   The role of backside polishing, cup angle, and polyethylene thickness on the contact stresses in metal-backed acetabular components [J].
Kurtz, SM ;
Edidin, AA ;
Bartel, DL .
JOURNAL OF BIOMECHANICS, 1997, 30 (06) :639-642
[15]   Femoral component wear in retrieved hip prostheses [J].
Kusaba, A ;
Kuroki, Y .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02) :331-336
[16]   Early polyethylene wear and osteolysis in cementless total hip arthroplasty - The influence of femoral head size and polyethylene thickness [J].
Lee, PC ;
Shih, CH ;
Chen, WJ ;
Tu, YK ;
Tai, CL .
JOURNAL OF ARTHROPLASTY, 1999, 14 (08) :976-981
[17]   Lubrication and wear modelling of artificial hip joints: A review [J].
Mattei, L. ;
Di Puccio, F. ;
Piccigallo, B. ;
Ciulli, E. .
TRIBOLOGY INTERNATIONAL, 2011, 44 (05) :532-549
[18]   The effect of clearance in total hip endoprostheses [J].
Munteanu, Florin ;
Sirbu, Paul ;
Barbarosie, Constantin ;
Munteanu, Anca ;
Botez, Paul .
AT-EQUAL 2009: 2009 ECSIS SYMPOSIUM ON ADVANCED TECHNOLOGIES FOR ENHANCED QUALITY OF LIFE: LAB-RS AND ARTIPED 2009, 2009, :57-+
[19]   Effect of head size on wear properties of metal-on-metal bearings of hip prostheses, and comparison with wear properties of metal-on-polyethylene bearings using hip simulator [J].
Okazaki, Yoshimitsu .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 31 :152-163
[20]   Influence of the manufacturing process in performance concerning wear of total hip prostheses [J].
Oliveira, A. L. L. .
JOURNAL OF THE BRAZILIAN SOCIETY OF MECHANICAL SCIENCES AND ENGINEERING, 2014, 36 (04) :787-795