Evaluation of Surface Finish and Dimensional Control of Tribological Metal-Ultra High Molecular Weight Polyethylene Pair of Commercially Available Hip Implants

被引:12
作者
Cubillos, Patricia O. [1 ]
dos Santos, Vinicius O. [1 ]
Pizzolatti, Andre Luiz A. [1 ]
da Rosa, Edison [1 ]
Roesler, Carlos R. M. [1 ]
机构
[1] Univ Fed Santa Catarina, Dept Mech Engn, Univ Hosp, Lab Biomech Engn LEBm, Florianopolis, SC, Brazil
关键词
femoral head; acetabular liner; wear rate; osteolysis; dimensional analysis; surface finish; UHMWPE ACETABULAR CUPS; CONTACT PRESSURE; BEARING SURFACE; FEMORAL HEADS; WEAR; SIMULATOR; PROSTHESES; ARTHROPLASTY; ROUGHNESS; THICKNESS;
D O I
10.1016/j.arth.2017.10.032
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Dimensional control and surface finish of the femoral head and acetabular liner are critical factors in the manufacturing process due to the risk of increased polyethylene wear, which is the primary cause of aseptic failure of a metal/polymer hip prosthesis. The aim of this study is to perform a dimensional and surface finish analysis to evaluate the reproduction and accuracy of the manufacturing processes of metal femoral heads and ultra high molecular weight polyethylene acetabular liners. Methods: Four femoral heads and acetabular liners from 5 manufacturers were evaluated. The methods of evaluation followed the standards ISO 7206-2:2011 and ISO 21535:2010. Results: The diameter, sphericity, and roughness of the femoral heads from all the manufacturers were in accordance with the standard requirements. Only the sphericity showed a lack of repeatability among the manufacturers. The variability in sphericity was high among some manufacturers and low in others. The diameters of the acetabular liners of 2 manufacturers were not in accordance with the standard requirements. The repeatability of sphericity, thickness, and roughness of the acetabular liners were heterogeneous among the manufacturers, which means that some manufacturers need to improve quality control. Conclusion: Our results showed a good dimensional and surface finish control of the manufacturing processes of the femoral heads. However, the same control was not shown during the manufacturing of the acetabular liners although all samples had a thickness and sphericity in accordance with the standard. A better quality control of the manufacturing process of ultra high molecular weight polyethylene acetabular liners should be made to improve the dimensional parameters of the acetabular liners and the tribological pair. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:939 / 944
页数:6
相关论文
共 35 条
[1]   Failure analysis on retrieved ultra high molecular weight polyethylene (UHMWPE) acetabular cups [J].
Burger, N. D. L. ;
de Vaal, P. L. ;
Meyer, J. P. .
ENGINEERING FAILURE ANALYSIS, 2007, 14 (07) :1329-1345
[2]   Size, shape, and composition of wear particles from metal-metal hip simulator testing: Effects of alloy and number of loading cycles [J].
Catelas, I ;
Bobyn, JD ;
Medley, JB ;
Krygier, JJ ;
Zukor, DJ ;
Huk, OL .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2003, 67A (01) :312-327
[3]   What Are the Risks Accompanying the Reduced Wear Benefit of Low-clearance Hip Resurfacing? [J].
Daniel, Joseph ;
Ziaee, Hena ;
Kamali, Amir ;
Pradhan, Chandra ;
McMinn, Derek .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (10) :2800-2809
[4]   A literature review of the association between wear rate and osteolysis in total hip arthroplasty [J].
Dumbleton, JH ;
Manley, MT ;
Edidin, AA .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :649-661
[5]   The quantitative assessment of UHMWPE wear debris produced in hip simulator testing: the influence of head material and roughness, motion and loading [J].
Elfick, APD ;
Smith, SL ;
Green, SM ;
Unsworth, A .
WEAR, 2001, 249 (5-6) :517-527
[6]   Comparison of hip simulator wear of 2 different highly cross-linked ultra high molecular weight polyethylene acetabular components using both 32-and 38-mm femoral heads [J].
Estok, Daniel M., II ;
Burroughs, Brian R. ;
Muratoglu, Orhun K. ;
Harris, William H. .
JOURNAL OF ARTHROPLASTY, 2007, 22 (04) :581-589
[7]  
Gispert MP, 2007, J BIOMED MATER RES B, V83, P340
[8]  
Green G, 2014, ORTHOP TRAUMA, V29, P79
[9]   The effect of surface topograph of retrieved femoral heads on the wear of UHMWPE sockets [J].
Hall, RM ;
Siney, P ;
Unsworth, A ;
Wroblewski, BM .
MEDICAL ENGINEERING & PHYSICS, 1997, 19 (08) :711-719
[10]   Impingement of acetabular cups in a hip simulator - Comparison of highly cross-linked and conventional polyethylene [J].
Holley, KG ;
Furman, BD ;
Babalola, OM ;
Lipman, JD ;
Padgett, DE ;
Wright, TM .
JOURNAL OF ARTHROPLASTY, 2005, 20 (07) :77-86