Enamel Cracks and Ceramic Bracket Failure during Debonding In Vitro

被引:46
作者
Bishara, Samir E. [1 ]
Ostby, Adam Wade
Laffoon, John
Warren, John J. [2 ]
机构
[1] Univ Iowa, Coll Dent, Dept Orthodont, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Prevent & Community Dent, Iowa City, IA 52242 USA
关键词
Ceramic brackets; Debonding; Enamel cracks;
D O I
10.2319/112007-540.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To test the null hypothesis that no difference in bracket failure characteristics is noted when use of a new ceramic bracket debonding instrument is compared with the use of conventional pliers. Materials and Methods: Thirty maxillary premolars were randomly assigned to one of two groups. In group 1, Clarity collapsible ceramic brackets (3M Unitek, Monrovia, Calif) were debonded with the use of conventional Utility/Weingart (3M Unitek, Monrovia, Calif) pliers. In group 2, Clarity brackets were debonded with a new Debonding Instrument (3M Unitek). For all teeth, the same bracket bonding system was used. Following debonding, teeth and brackets were examined under 10x magnification for assessment of bracket failure (fracture) and of residual adhesive on the enamel surface. Enamel surfaces were visualized with transillumination prior to bonding and after removal of the residual adhesive, so the effect of the debonding forces could be determined. Results: The results of Adhesive Remnant Index comparisons indicated that a statistically significant difference (chi(2) = 8.73-1 P = .013) in bond failure patterns was apparent when the two groups were compared. Brackets debonded with the new instrument showed a greater tendency for the adhesive to be removed from the tooth during debonding. Conclusions: The hypothesis is rejected. Although the incidence of enamel damage following debonding was similar in the two groups, the use of the new Debonding Instrument decreased the incidence of bracket fracture.
引用
收藏
页码:1078 / 1083
页数:6
相关论文
共 27 条
[1]   EVALUATION OF FRICTION BETWEEN CERAMIC BRACKETS AND ORTHODONTIC WIRES OF 4 ALLOYS [J].
ANGOLKAR, PV ;
KAPILA, S ;
DUNCANSON, MG ;
NANDA, RS .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1990, 98 (06) :499-506
[2]   The force levels required to mechanically debond ceramic brackets: an in vitro comparative study [J].
Arici, S ;
Minors, C .
EUROPEAN JOURNAL OF ORTHODONTICS, 2000, 22 (03) :327-334
[3]   A post-treatment evaluation of multibonded ceramic brackets in orthodontics [J].
Artun, J .
EUROPEAN JOURNAL OF ORTHODONTICS, 1997, 19 (02) :219-228
[4]  
Birnie D, 1990, Br J Orthod, V17, P71
[5]   THE USE OF DEBONDING PLIERS IN THE REMOVAL OF CERAMIC BRACKETS - FORCE LEVELS AND ENAMEL CRACKS [J].
BISHARA, SE ;
ORTHO, D ;
FONSECA, JM ;
BOYER, DB .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1995, 108 (03) :242-248
[6]   COMPARISONS OF DIFFERENT DEBONDING TECHNIQUES FOR CERAMIC BRACKETS - AN INVITRO STUDY .1. BACKGROUND AND METHODS [J].
BISHARA, SE ;
TRULOVE, TS .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1990, 98 (02) :145-153
[7]   COMPARISONS OF DIFFERENT DEBONDING TECHNIQUES FOR CERAMIC BRACKETS - AN INVITRO STUDY .2. FINDINGS AND CLINICAL IMPLICATIONS [J].
BISHARA, SE ;
TRULOVE, TS .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1990, 98 (03) :263-273
[8]   A COMPARATIVE-STUDY OF THE DEBONDING STRENGTHS OF DIFFERENT CERAMIC BRACKETS, ENAMEL CONDITIONERS, AND ADHESIVES [J].
BISHARA, SE ;
FEHR, DE ;
JAKOBSEN, JR .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1993, 104 (02) :170-179
[9]   SHEAR BOND STRENGTH OF CERAMIC ORTHODONTIC BRACKETS TO ENAMEL [J].
BRITTON, JC ;
MCINNES, P ;
WEINBERG, R ;
LEDOUX, WR ;
RETIEF, DH .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1990, 98 (04) :348-353
[10]  
CHACONAS S, 1990, ANGLE ORTHOD, V60, P35