Tracked Ultrasound Snapshots in Percutaneous Pedicle Screw Placement Navigation: A Feasibility Study

被引:17
作者
Ungi, Tamas [1 ]
Moult, Eric [1 ]
Schwab, Joseph H. [2 ]
Fichtinger, Gabor [1 ]
机构
[1] Queens Univ, Lab Percutaneous Surg, Sch Comp, Kingston, ON K7M2N8, Canada
[2] Massachusetts Gen Hosp, Dept Orthoped Surg, Boston, MA 02114 USA
关键词
INTRAOPERATIVE COMPUTED-TOMOGRAPHY; CT REGISTRATION; SYSTEM; INSERTION; ACCURACY; FIXATION; GUIDANCE;
D O I
10.1007/s11999-013-3239-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Computerized navigation improves the accuracy of minimally invasive pedicle screw placement during spine surgery. Such navigation, however, exposes both the patient and the staff to radiation during surgery. To avoid intraoperative exposure to radiation, tracked ultrasound snapshots-ultrasound image frames coupled with corresponding spatial positions-could be used to map preoperatively defined screw plans into the intraoperative coordinate frame. The feasibility of such an approach, however, has not yet been investigated. Are there vertebral landmarks that can be identified using tracked ultrasound snapshots? Can tracked ultrasound snapshots allow preoperative pedicle screw plans to be accurately mapped-compared with CT-derived pedicle screw plans-into the intraoperative coordinate frame in a simulated setting? Ultrasound visibility of registration landmarks was checked on volunteers and phantoms. An ultrasound machine with integrated electromagnetic tracking was used for tracked ultrasound acquisition. Registration was performed using 3D Slicer open-source software (www.slicer.org). Two artificial lumbar spine phantoms were used to evaluate registration accuracy of pedicle screw plans using tracked ultrasound snapshots. Registration accuracy was determined by comparing the ultrasound-derived plans with the CT-derived plans. The four articular processes proved to be identifiable using tracked ultrasound snapshots. Pedicle screw plans were registered to the intraoperative coordinate system using landmarks. The registrations were sufficiently accurate in that none of the registered screw plans intersected the pedicle walls. Registered screw plan positions had an error less than 1.28 +/- A 1.37 mm (average +/- A SD) in each direction and an angle difference less than 1.92A degrees A A +/- A 1.95A degrees around each axis relative to the CT-derived positions. Registration landmarks could be located using tracked ultrasound snapshots and permitted accurate mapping of pedicle screw plans to the intraoperative coordinate frame in a simulated setting. Tracked ultrasound may allow accurate computer-navigated pedicle screw placement while avoiding ionizing radiation in the operating room; however, further studies that compare this approach with other navigation techniques are needed to confirm the practical use of this new approach.
引用
收藏
页码:4047 / 4055
页数:9
相关论文
共 22 条
[1]  
Bandela JR, 2011, WORLD NEUROSURG
[2]   Spinal Navigation: Standard Preoperative Versus Intraoperative Computed Tomography Data Set Acquisition for Computer-Guidance System Radiological and Clinical Study in 100 Consecutive Patients [J].
Costa, Francesco ;
Cardia, Andrea ;
Ortolina, Alessandro ;
Fabio, Galbusera ;
Zerbi, Alberto ;
Fornari, Maurizio .
SPINE, 2011, 36 (24) :2094-2098
[3]   Application of Intraoperative Computed Tomography With or Without Navigation System in Surgical Correction of Spinal Deformity A Preliminary Result of 59 Consecutive Human Cases [J].
Cui, Geng ;
Wang, Yan ;
Kao, Ting-Hsien ;
Zhang, Yonggang ;
Liu, Zhengsheng ;
Liu, Baowei ;
Li, Jie ;
Zhang, Xifeng ;
Zhu, Shourong ;
Lu, Ning ;
Mao, Keya ;
Wang, Zheng ;
Zhang, Xuesong ;
Yuan, Xincheng ;
Dong, Tianxiang ;
Xiao, Songhua .
SPINE, 2012, 37 (10) :891-900
[4]   Clinically relevant complications related to pedicle screw placement in thoracolumbar surgery and their management: a literature review of 35,630 pedicle screws [J].
Gautschi, Oliver P. ;
Schatlo, Bawarjan ;
Schaller, Karl ;
Tessitore, Enrico .
NEUROSURGICAL FOCUS, 2011, 31 (04)
[5]   Biomechanically constrained groupwise ultrasound to CT registration of the lumbar spine [J].
Gill, Sean ;
Abolmaesumi, Purang ;
Fichtinger, Gabor ;
Boisvert, Jonathan ;
Pichora, David ;
Borshneck, Dan ;
Mousavi, Parvin .
MEDICAL IMAGE ANALYSIS, 2012, 16 (03) :662-674
[6]   Ultrasound-guided lumbar facet nerve block - A sonoanatomic study of a new methodologic approach [J].
Greher, M ;
Scharbert, G ;
Kamolz, LP ;
Beck, H ;
Gustorff, B ;
Kirchmair, L ;
Kapral, S .
ANESTHESIOLOGY, 2004, 100 (05) :1242-1248
[7]  
Hu X, 2012, EUR SPINE J
[8]   APPLICATION OF FRAMELESS STEREOTAXY TO PEDICLE SCREW FIXATION OF THE SPINE [J].
KALFAS, IH ;
KORMOS, DW ;
MURPHY, MA ;
MCKENZIE, RL ;
BARNETT, GH ;
BELL, GR ;
STEINER, CP ;
TRIMBLE, MB ;
WEISENBERGER, JP .
JOURNAL OF NEUROSURGERY, 1995, 83 (04) :641-647
[9]   Development of a New Technique for Pedicle Screw and Magerl Screw Insertion Using a 3-Dimensional Image Guide [J].
Kawaguchi, Yoshiharu ;
Nakano, Masato ;
Yasuda, Taketoshi ;
Seki, Shoji ;
Hori, Takeshi ;
Kimura, Tomoatsu .
SPINE, 2012, 37 (23) :1983-1988
[10]   The Accuracy of Navigation and 3D Image-Guided Placement for the Placement of Pedicle Screws in Congenital Spine Deformity [J].
Larson, A. Noelle ;
Polly, David W., Jr. ;
Guidera, Kenneth J. ;
Mielke, Cary H. ;
Santos, Edward R. G. ;
Ledonio, Charles Gerald T. ;
Sembrano, Jonathan N. .
JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2012, 32 (06) :E23-E29