Assessment of PCXMC for patients with different body size in chest and abdominal x ray examinations: a Monte Carlo simulation study

被引:25
作者
Borrego, David [1 ]
Lowe, Erin M. [1 ]
Kitahara, Cari M. [1 ]
Lee, Choonsik [1 ]
机构
[1] NCI, Div Canc Epidemiol & Genet, NIH, 9609 Med Ctr Dr, Bethesda, MD 20892 USA
基金
美国国家卫生研究院;
关键词
Monte Carlo; dosimetry; epidemiology; PCXMC; hybrid phantom; DIAGNOSTIC-RADIOLOGY; DOSIMETRY; SPECTRA; ORGAN; PHANTOM; POPULATION; MODELS; FAMILY; TOOL;
D O I
10.1088/1361-6560/aab13e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A PC Program for x ray Monte Carlo (PCXMC) has been used to calculate organ doses in patient dosimetry and for the exposure assessment in epidemiological studies of radiogenic health related risks. This study compared the dosimetry from using the built-in stylized phantoms in the PCXMC to that of a newer hybrid phantom library with improved anatomical realism. We simulated chest and abdominal x ray projections for 146 unique body size computational phantoms, 77 males and 69 females, with different combinations of height (125-180 cm) and weight (20-140 kg) using the built-in stylized phantoms in the PCXMC version 2.0.1.4 and the hybrid phantom library using the Monte Carlo N-particle eXtended transport code 2.7 (MCNPX). Unfortunately, it was not possible to incorporate the hybrid phantom library into the PCXMC. We compared 14 organ doses, including dose to the active bone marrow, to evaluate differences between the built-in stylized phantoms in the PCXMC and the hybrid phantoms (Cristy and Eckerman 1987 Technical Report ORNL/TM-8381/V1, Oak Ridge National Laboratory, Eckerman and Ryman 1993 Technical Report 12 Oak Ridge, TN, Geyer et al 2014 Phys. Med. Biol. 59 5225-42). On average, organ doses calculated using the built-in stylized phantoms in the PCXMC were greater when compared to the hybrid phantoms. This is most prominent in AP abdominal exams by an average factor of 2.4-, 2.8-, and 2.8-fold for the 10-year-old, 15-year-old, and adult phantoms, respectively. For chest exams, organ doses are greater by an average factor of 1.1-, 1.4-, and 1.2-fold for the 10-year-old, 15-year-old, and adult phantoms, respectively. The PCXMX, due to its ease of use, is often selected to support dosimetry in epidemiological studies; however, it uses simplified models of the human anatomy that fail to account for variations in body morphometry for increasing weight. For epidemiological studies that use PCXMC dosimetry, associations between radiation-related disease risks and organ doses may be underestimated, and to a greater degree in pediatric, especially obese pediatric, compared to adult patients.
引用
收藏
页数:15
相关论文
共 33 条
  • [1] Assessment of different computational models for generation of x-ray spectra in diagnostic radiology and mammography
    Ay, MR
    Sarkar, S
    Shahriari, M
    Sardari, D
    Zaidi, H
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1660 - 1675
  • [2] Bays H, 2015, CURR MED RES OPIN, V28, P1
  • [3] Baysson H, 2013, BMC PUBLIC HEALTH, V13, P670
  • [4] COMPUTATION OF BREMSSTRAHLUNG X-RAY-SPECTRA AND COMPARISON WITH SPECTRA MEASURED WITH A GE(LI) DETECTOR
    BIRCH, R
    MARSHALL, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) : 505 - 517
  • [5] Bontrager KL., 2010, Textbook of radiographic positioning and related anatomy, V7th
  • [6] Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV
    Boone, JM
    Seibert, JA
    [J]. MEDICAL PHYSICS, 1997, 24 (11) : 1661 - 1670
  • [7] THYROID RADIATION DOSE TO PATIENTS FROM DIAGNOSTIC RADIOLOGY PROCEDURES OVER EIGHT DECADES: 1930-2010
    Chang, Lienard A.
    Miller, Donald L.
    Lee, Choonsik
    Melo, Dunstana R.
    Villoing, Daphnee
    Drozdovitch, Vladimir
    Thierry-Chef, Isabelle
    Winters, Sarah J.
    Labrake, Michael
    Myers, Charles F.
    Lim, Hyeyeun
    Kitahara, Cari M.
    Linet, Martha S.
    Simon, Steven L.
    [J]. HEALTH PHYSICS, 2017, 113 (06): : 458 - 473
  • [8] Cristy M., 1987, ORNLTM8381V1
  • [9] Eckerman K F, 1993, 12 TN
  • [10] Francis P, 2009, INSPIRING WRITING IN ART AND DESIGN: TAKING A LINE FOR A WRITE, P1