Dosimetric evaluation of a new OneDose MOSFET for Ir-192 energy

被引:18
作者
Kinhikar, RA [1 ]
Sharma, PK [1 ]
Tambe, CM [1 ]
Deshpande, DD [1 ]
机构
[1] Tata Mem Hosp, Dept Med Phys, Bombay 400012, Maharashtra, India
关键词
D O I
10.1088/0031-9155/51/5/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate dosimetry (reproducibility, energy correction, relative response with distance from source, linearity with threshold dose, rate of fading, temperature and angular dependence) of a newly designed OneDose (TM) MOSFET patient dosimetry system for use in HDR brachytherapy with Ir-192 energy. All measurements were performed with a MicroSelectron HDR unit and OncDose MOSFET detectors. All dosimeters were normalized to 3 min post-irradiation to minimize fading effects. All dosimeters gave reproducible readings with mean deviation of 1.8% (SD 0.4) and 2.4% (SD 0.6) for 0 degrees and 180 degrees incidences, respectively. The mean energy correction factor was found to be 1.1 (range 1.06-1.12). Overall, there was 60% and 40% mean response of the MOSFET at 2 and 3 cm, respectively, from the source. MOSFET results showed good agreement with TLD and parallel plate ion chamber. Linear dose response with threshold voltage shift was observed with applied doses of 0.3 Gy-5 Gy with Ir-192 energy. Linearity (R-2 = 1) was observed in the MOSFET signal with the applied dose range of 0.3 Gy-5 Gy with Ir-192 energy. Fading effects were less than 1% after 10 min and the MOSFET detectors stayed stable (within 5%) over a period of 1 month. The MOSFET response was found to be decreased by approximately 1.5% at 37 degrees C compared to 20 degrees C. The isotropic response of the MOSFET was found to be within +/- 6%. A maximum deviation of 5.5% was obtained between 0 degrees and 180 degrees for both the axes and this should be considered in clinical applications. The small size, cable-less, instant readout, permanent storage of dose and ease of use make the MOSFET a novel dosimeter and beneficial to patients for skin dose measurements with HDRBT using an Ir-192 source compared to the labour demanding and time-consuming TLDs.
引用
收藏
页码:1261 / 1268
页数:8
相关论文
共 8 条
  • [1] A new radiotherapy surface dose detector: The MOSFET
    Butson, MJ
    Rozenfeld, A
    Mathur, JN
    Carolan, M
    Wong, TPY
    Metcalfe, PE
    [J]. MEDICAL PHYSICS, 1996, 23 (05) : 655 - 658
  • [2] Cheung T., 2003, AUSTR PHYS ENG SCI M, V26, P6
  • [3] Application of commercial MOSFET detectors for in vivo dosimetry in the therapeutic x-ray range from 80 kV to 250 W
    Ehringfeld, C
    Schmid, S
    Poljanc, K
    Kirisits, C
    Aiginger, H
    Georg, D
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (02) : 289 - 303
  • [4] A MINIATURE MOSFET RADIATION DOSIMETER PROBE
    GLADSTONE, DJ
    LU, XQ
    HUMM, JL
    BOWMAN, HF
    CHI, LM
    [J]. MEDICAL PHYSICS, 1994, 21 (11) : 1721 - 1728
  • [5] Dosimetric evaluation of a new design MOSFET in vivo dosimeter
    Halvorsen, PH
    [J]. MEDICAL PHYSICS, 2005, 32 (01) : 110 - 117
  • [6] Measurement of radiotherapy x-ray skin dose on a chest wall phantom
    Quach, KY
    Morales, J
    Butson, MJ
    Rosenfeld, AB
    Metcalfe, PE
    [J]. MEDICAL PHYSICS, 2000, 27 (07) : 1676 - 1680
  • [7] Clinical dosimetry using MOSFETS
    Ramani, R
    Russell, S
    OBrien, P
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (04): : 959 - 964
  • [8] Calibration of a MOSFET detection system for 6-MV in vivo dosimetry
    Scalchi, P
    Francescon, P
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (04): : 987 - 993