Evaluation of Three Turbidimetric Assays for Automated Determination of Hemoglobin A1c

被引:5
|
作者
Barrot, Amandine [1 ]
Dupuy, Anne Marie [1 ]
Badiou, Stephanie [1 ]
Bargnoux, Anne Sophie [1 ]
Cristol, Jean Paul [1 ]
机构
[1] Hosp Lapeyronie, Dept Biochem, F-34295 Montpellier 5, France
关键词
Hemoglogin A1c; immunoassay; chromatography; turbidimetry; S-TRAITS;
D O I
10.7754/Clin.Lab.2012.111222
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: To compare the results of HbA1c determination obtained through immunoassays versus the HPLC method currently used routinely in our laboratory. Methods: We evaluated immunoturbidimetric assays for the HbA1c measure on three analyzers, specifically the Roche Cobas Integra 400+(R) (Roche Diagnostics, Indianapolis, IN, USA), Ortho Clinical Diagnostics Vitros 5.1 FS (R) (Ortho Clinical Diagnostics, NY, USA), and Siemens Dimension RxL (R) (Siemens Healthcare Diagnostics, NY, USA), in comparison with the HPLC Menarini HA 8140 (R) (Menarini Diagnostics, Rungis, France) currently used in our laboratory. Results: Analytical performances including precision, analytical range, recovery, carryover, erythrosedimentation and comparison studies were acceptable leading to results with a level of exactitude in accordance with the recommendations of the National Glycohemoglobin Standardization Program (NGSP). Conclusions: The three immunoassays tested can be used interchangeably and will be satisfactory for laboratories who cannot invest in a HPLC analyzer. (Clin. Lab. 2012;58:1171-1177. DOI: 10.7754/Clin.Lab.2012.111222)
引用
收藏
页码:1171 / 1177
页数:7
相关论文
共 50 条
  • [31] Determination of reference intervals in adults for hemoglobin A1c (HbA1c).
    Lehmann, P
    Junge, W
    Wilke, B
    Halabi, A
    Matouk, H
    Klein, G
    CLINICAL CHEMISTRY, 2003, 49 (06) : A113 - A113
  • [32] Multicenter evaluation of hemoglobin A1c assay on capillary electrophoresis
    Marinova, Mariela
    Altinier, Sara
    Caldini, Anna
    Passerini, Gabriella
    Pizzagalli, Giorgio
    Brogi, Marco
    Zaninotto, Martina
    Ceriotti, Ferruccio
    Plebani, Mario
    CLINICA CHIMICA ACTA, 2013, 424 : 207 - 211
  • [33] MORTALITY, GLYCEMIC CONTROL AND HEMOGLOBIN A1C (A1C) IN THE ICU
    Carpenter, David
    Gregg, Sara
    Xu, Kejun
    Coopersmith, Craig
    CRITICAL CARE MEDICINE, 2015, 43 (12)
  • [34] A COMPARISON OF ANALYTICAL GOALS FOR HEMOGLOBIN A1C ASSAYS DERIVED USING DIFFERENT STRATEGIES
    LARSEN, ML
    FRASER, CG
    PETERSEN, PH
    ANNALS OF CLINICAL BIOCHEMISTRY, 1991, 28 : 272 - 278
  • [35] Evaluation of the tosoh HLC-723GHb V A1c 2.2 hemoglobin A1c analyzer
    Chevenne, D
    Marle, N
    Chauffert, M
    Noel, M
    Ducrocq, R
    Trivin, F
    CLINICAL BIOCHEMISTRY, 1999, 32 (06) : 487 - 490
  • [36] Led Astray by Hemoglobin A1c: A Case of Misdiagnosis of Diabetes by Falsely Elevated Hemoglobin A1c
    Chen, Jean
    Diesburg-Stanwood, Amy
    Bodor, Geza
    Rasouli, Neda
    JOURNAL OF INVESTIGATIVE MEDICINE HIGH IMPACT CASE REPORTS, 2016, 4 (01):
  • [37] SEPARATION OF THE DIFFERENT HUMAN HEMOGLOBINS - APPLICATION TO THE DETERMINATION OF HEMOGLOBIN A1C
    PERIER, C
    CHAMSON, A
    FARJANEL, J
    FREY, J
    LYON PHARMACEUTIQUE, 1979, 30 (06): : 457 - 459
  • [38] Evaluation of a dual hemoglobin A2/A1c quantitation kit on the bio-rad variant II automated hemoglobin analyzer
    Lafferty, JD
    McFarlane, AG
    Chui, DHK
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2002, 126 (12) : 1494 - 1500
  • [39] Evaluation of D10® hemoglobin testing system for hemoglobin A1C assay
    Marzullo, C.
    Minery, M.
    ANNALES DE BIOLOGIE CLINIQUE, 2008, 66 (01) : 95 - 99
  • [40] Evaluation of hemoglobin A1c measurement on the analyzer Cobas® c311
    El Boukhrissi, F.
    Balouch, H.
    Bamou, Y.
    IMMUNO-ANALYSE & BIOLOGIE SPECIALISEE, 2012, 27 (01): : 28 - 32