Evaluation of Three Turbidimetric Assays for Automated Determination of Hemoglobin A1c

被引:5
|
作者
Barrot, Amandine [1 ]
Dupuy, Anne Marie [1 ]
Badiou, Stephanie [1 ]
Bargnoux, Anne Sophie [1 ]
Cristol, Jean Paul [1 ]
机构
[1] Hosp Lapeyronie, Dept Biochem, F-34295 Montpellier 5, France
关键词
Hemoglogin A1c; immunoassay; chromatography; turbidimetry; S-TRAITS;
D O I
10.7754/Clin.Lab.2012.111222
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: To compare the results of HbA1c determination obtained through immunoassays versus the HPLC method currently used routinely in our laboratory. Methods: We evaluated immunoturbidimetric assays for the HbA1c measure on three analyzers, specifically the Roche Cobas Integra 400+(R) (Roche Diagnostics, Indianapolis, IN, USA), Ortho Clinical Diagnostics Vitros 5.1 FS (R) (Ortho Clinical Diagnostics, NY, USA), and Siemens Dimension RxL (R) (Siemens Healthcare Diagnostics, NY, USA), in comparison with the HPLC Menarini HA 8140 (R) (Menarini Diagnostics, Rungis, France) currently used in our laboratory. Results: Analytical performances including precision, analytical range, recovery, carryover, erythrosedimentation and comparison studies were acceptable leading to results with a level of exactitude in accordance with the recommendations of the National Glycohemoglobin Standardization Program (NGSP). Conclusions: The three immunoassays tested can be used interchangeably and will be satisfactory for laboratories who cannot invest in a HPLC analyzer. (Clin. Lab. 2012;58:1171-1177. DOI: 10.7754/Clin.Lab.2012.111222)
引用
收藏
页码:1171 / 1177
页数:7
相关论文
共 50 条
  • [21] Evaluation of a fully automated high-performance liquid chromatography assay for hemoglobin A1c
    Khuu, HM
    Robinson, CA
    Goolsby, K
    Hardy, RW
    Konrad, RJ
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 1999, 123 (09) : 763 - 767
  • [22] Comparison of costs of hemoglobin A1c determination by point-of-care testing (POCT) and standard hemoglobin A1c analyzer
    Wiwanitkit, V.
    ARCHIVES OF HELLENIC MEDICINE, 2010, 27 (06): : 953 - 955
  • [23] Availability of an immunoassay for the determination of glycated hemoglobin (Hb A1c): Influence of hemoglobin variant on Hb A1c values.
    Harano, T
    Harano, K
    Doi, K
    Yoshikawa, K
    Kamada, S
    Mochizuki, K
    Saitoh, H
    CLINICAL CHEMISTRY, 1998, 44 : A27 - A27
  • [24] Components of total measurement error for hemoglobin A1c determination
    Phillipov, G
    Phillips, PJ
    CLINICAL CHEMISTRY, 2001, 47 (10) : 1851 - 1853
  • [25] Determination of Percent Hemoglobin A1c Using a Potentiometric Method
    Liu, Hong
    Crooks, Richard M.
    ANALYTICAL CHEMISTRY, 2013, 85 (03) : 1834 - 1839
  • [26] Standardization of hemoglobin A1c
    Johnson, R
    CLINICAL CHEMISTRY, 1998, 44 (05) : 1068 - 1068
  • [27] Significance of hemoglobin A1c
    Purucker, E
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2001, 126 (20) : 608 - 609
  • [28] HEMOGLOBIN A1C AND UREMIA
    LUND, L
    MOURITSANDERSEN, T
    SORENSEN, PJ
    CLINICAL NEPHROLOGY, 1988, 29 (03) : 161 - 162
  • [29] Evaluation of Accu-Chek A1c™ test for blood hemoglobin A1c measurements.
    Azhar, A
    Hon, K
    Verdun, R
    CLINICAL CHEMISTRY, 1999, 45 (06) : A65 - A65
  • [30] Glucose and Hemoglobin A1c
    Freeman, Vicki S.
    LABMEDICINE, 2014, 45 (01): : E21 - E24