Intraocular polyimide intraocular lens haptic breakage long-term postoperatively

被引:12
作者
Stallings, Shannon [1 ]
Werner, Liliana [1 ]
Chayet, Arturo [3 ]
Masket, Samuel [2 ]
Camacho, Fidel [3 ]
Peck, Carolee Cutler [1 ]
Mamalis, Nick [1 ]
机构
[1] Univ Utah, John Moran Eye Ctr, Salt Lake City, UT 84132 USA
[2] Adv Vis Care, Los Angeles, CA USA
[3] Codet Vis Inst, Tijuana, Mexico
关键词
OCULAR TISSUE-EXTRACTS; BIODEGRADATION; DEGRADATION; LOOPS;
D O I
10.1016/j.jcrs.2013.11.024
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We present 2 cases of 3-piece silicone intraoculanbsp;r lenses (IOLs) with broken polyimide haptics. In the first case, the IOL was implanted in the anterior chamber after posterior capsule rupture. Twelve years later, it was explanted and exchanged because of bullous keratopathy. However, on careful manipulation for explantation, the haptics shattered into multiple pieces. In the second case, initial in-the-bag implantation of the IOL was uneventful. Fifteen years later, the IOL dislocated into the anterior chamber as a result of spontaneous in situ haptic breakage. The haptics of both IOLs appeared very brittle during careful manipulation for gross and light microscopy, with further breakage. Scanning electron microscopy of both IOLs revealed no evidence of haptic degradation. Their surfaces appeared smooth and regular with jagged edges where the breaks occurred. Few studies have examined the long-term biocompatibility of polyimide, and more research is needed to determine the cause of this phenomenon. (C) 2014 ASCRS and ESCRS
引用
收藏
页码:323 / 326
页数:4
相关论文
共 9 条
[1]  
APPLE DJ, 1984, AM INTRA-OCUL IMPL J, V10, P53
[2]   A SCANNING ELECTRON-MICROSCOPIC STUDY OF NYLON DEGRADATION BY OCULAR TISSUE-EXTRACTS [J].
HAYASAKA, S ;
ISHIGURO, S ;
SHIONO, T ;
OKABE, H ;
MIZUNO, K .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1982, 93 (01) :111-117
[3]  
HAYASAKA S, 1984, JPN J OPHTHALMOL, V28, P131
[4]   Loop memory of haptic materials in posterior chamber intraocular lenses [J].
Izak, AM ;
Werner, L ;
Apple, DJ ;
Macky, TA ;
Trivedi, RH ;
Pandey, SK .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07) :1229-1235
[5]  
KRAUSE U, 1986, ACTA OPHTHALMOL, V64, P338
[6]  
LERMAN S, 1983, AM INTRAOCULAR IMPLA, V9, P25
[7]  
MEHTA HK, 1979, T OPHTHAL SOC UK, V99, P183
[8]  
Mowbray S L, 1983, J Am Intraocul Implant Soc, V9, P143
[9]   POLYIMIDES AS BIOMATERIALS - PRELIMINARY BIOCOMPATIBILITY TESTING [J].
RICHARDSON, RR ;
MILLER, JA ;
REICHERT, WM .
BIOMATERIALS, 1993, 14 (08) :627-635