Intraoperative Pachymetry Using Spectral-Domain Optical Coherence Tomography during Accelerated Corneal Collagen Crosslinking

被引:12
作者
Chow, Vanissa W. S. [1 ,2 ]
Biswas, Sayantan [2 ]
Yu, Marco [2 ]
Wong, Victoria W. Y. [1 ,2 ]
Jhanji, Vishal [1 ,2 ,3 ,4 ]
机构
[1] Hong Kong Eye Hosp, Hong Kong, Hong Kong, Peoples R China
[2] Chinese Univ Hong Kong, Dept Ophthalmol & Visual Sci, Hong Kong, Hong Kong, Peoples R China
[3] Prince Wales Hosp, Dept Ophthalmol, Shatin, Hong Kong, Peoples R China
[4] Univ Melbourne, Ctr Eye Res Australia, Melbourne, Vic 3002, Australia
关键词
HYPOOSMOLAR RIBOFLAVIN SOLUTION; ULTRAVIOLET-A; THICKNESS MEASUREMENTS; EYES; KERATOCONUS;
D O I
10.1155/2013/848363
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Purpose. To evaluate the role of spectral-domain optical coherence tomography (SDOCT) to measure corneal thickness during accelerated corneal crosslinking (CXL). Methods. Intraoperative pachymetry was performed using SDOCT and ultrasound pachymetry (USP) in 6 eyes of 6 patients with keratoconus. Pachymetry readings were obtained at baseline, after epithelium removal and after 30 minutes of riboflavin instillation. SDOCT measurements of eyes with and without lid speculum during riboflavin instillation were compared. Results. There was no statistically significant difference in central corneal thickness (CCT) measurements between SDOCT and USP (P > 0.05 for all). A significant decrease in both CCT (P = 0.031) and the thinnest corneal thickness (TCT) (P = 0.031) was observed during CXL. There was a greater reduction in CCT (38 +/- 6%) with the use of lid speculum as compared to the no-speculum eyes (18 +/- 9%) (P = 0.100). TCT was also reduced by a greater extent with the use of lid speculum (40 +/- 5% versus 26 +/- 7%; P = 0.100). Conclusion. SDOCT can be successfully used to measure intraoperative corneal pachymetry during corneal CXL. SDOCT measurements demonstrated corneal thinning intraoperatively during CXL, which was further accentuated by the use of a lid speculum during the procedure.
引用
收藏
页数:6
相关论文
共 27 条
[1]  
Alario A. F., 2013, VET OPHTHALMOLOGY
[2]   Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry [J].
Chen, Shihao ;
Huang, Jinhai ;
Wen, Daizong ;
Chen, Wuhe ;
Huang, David ;
Wang, Qinmei .
ACTA OPHTHALMOLOGICA, 2012, 90 (05) :449-455
[3]   Transepithelial corneal collagen crosslinking: Bilateral study [J].
Filippello, Massimo ;
Stagni, Edoardo ;
O'Brart, David .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (02) :283-291
[4]   Corneal collagen cross-linking using riboflavin and ultraviolet-A irradiation: A review of clinical and experimental studies [J].
Gkika M. ;
Labiris G. ;
Kozobolis V. .
International Ophthalmology, 2011, 31 (4) :309-319
[5]   Assessment of central corneal thickness in normal, keratoconus, and post-laser in situ keratomileusis eyes using Scheimpflug imaging, spectral domain optical coherence tomography, and ultrasound pachymetry [J].
Grewal, Dilraj S. ;
Brar, Gagandeep S. ;
Grewal, Satinder P. S. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (06) :954-964
[6]   Collagen crosslinking with ultraviolet-A and hypoosmolar riboflavin solution in thin corneas [J].
Hafezi, Farhad ;
Mrochen, Michael ;
Iseli, Hans Peter ;
Seiler, Theo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (04) :621-624
[7]   Transient Corneal Thinning in Eyes Undergoing Corneal Cross-Linking [J].
Holopainen, Juha M. ;
Krootila, Kari .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2011, 152 (04) :533-536
[8]   Central Corneal Thickness Measurements With Fourier-Domain Optical Coherence Tomography Versus Ultrasonic Pachymetry and Rotating Scheimpflug Camera [J].
Ishibazawa, Akihiro ;
Igarashi, Sho ;
Hanada, Kazuomi ;
Nagaoka, Taiji ;
Ishiko, Satoshi ;
Ito, Haruna ;
Yoshida, Akitoshi .
CORNEA, 2011, 30 (06) :615-619
[9]   Management of keratoconus: current scenario [J].
Jhanji, Vishal ;
Sharma, Namrata ;
Vajpayee, Rasik B. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2011, 95 (08) :1044-1050
[10]   Intraoperative Corneal Thickness Measurements During Corneal Collagen Cross-Linking With Hypoosmolar Riboflavin Solution in Thin Corneas [J].
Kaya, Vedat ;
Utine, Canan Asli ;
Yilmaz, Omer Faruk .
CORNEA, 2012, 31 (05) :486-490