PURPOSE: To evaluate in keratoconus eyes the intrasubject repeatability of anterior and posterior corneal curvature and of other anterior segment anatomic measurements obtained with a new topography system combining Scheimpflug-photography and Placido-disk technology. SETTING: Vissum Corporation, Alicante, Spain. DESIGN: Evaluation of technology. METHODS: All keratoconus eyes had a comprehensive ophthalmologic examination including analysis with the Sirius system. Three consecutive measurements were obtained to assess the intrasubject repeatability of the following parameters: anterior and posterior corneal curvature and shape factor, white-to-white (WTW) corneal diameter, central and minimum corneal thickness, and anterior chamber depth (ACD). The within-subject standard deviation (S-w) and intraclass correlation coefficient (ICC) were calculated. RESULTS: This study comprised 61 eyes of 61 patients ranging in age from 14 to 64 years. For anterior and posterior corneal curvatures and power vector components, the S-w was 0.29 mm or less in all cases. The ICC was above 0.990 in all cases except the flattest curvature of the posterior corneal surface at 3.0 mm, which was 0.840 (moderate agreement), and the posterior power vector J0, which was 0.665 (poor agreement), 0.752, and 0.758 (moderate agreement) for 3.0 mm, 5.0 mm, and 7.0 mm, respectively. In shape factor measurements, the S-w was 0.12 or less in all cases and the ICC ranged between 0.989 and 0.999. Pachymetry, ACD, and WTW had ICC values very close to 1. CONCLUSION: The new topography system provided repeatable measurements of corneal shape and other anatomic parameters in eyes with keratoconus.
引用
收藏
页码:211 / 218
页数:8
相关论文
共 30 条
[1]   Keratoconus-integrated characterization considering anterior corneal aberrations, internal astigmatism, and corneal biomechanics [J].
Alio, Jorge L. ;
Pinero, David P. ;
Aleson, Alicia ;
Teus, Miguel A. ;
Barraquer, Rafael I. ;
Murta, Joaquim ;
Maldonado, Miguel J. ;
Castro de Luna, Gracia ;
Gutierrez, Ramon ;
Villa, Cesar ;
Uceda-Montanes, Antonio .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (03) :552-568
[2]   Measurement error and correlation coefficients [J].
Bland, JM ;
Altman, DG .
BRITISH MEDICAL JOURNAL, 1996, 313 (7048) :41-42
[3]  
Bland JM, 1996, BRIT MED J, V313, P744
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
Bühren J, 2012, OPHTHALMOLOGE, V109, P37, DOI 10.1007/s00347-011-2446-2
[6]   Intrasession and intersession repeatability of the Pentacam system on posterior corneal assessment in the normal human eye [J].
Chen, Davie ;
Lam, Andrew K. C. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (03) :448-454
[7]   Reliability and repeatability of the Pentacam on corneal curvatures [J].
Chen, Davie ;
Lam, Andrew K. C. .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2009, 92 (02) :110-118
[8]   Reproductivity and repeatability of central corneal thickness measurement in keratoconus using the rotating scheimpflug camera and ultrasound pachymetry [J].
De Sanctis, Ugo ;
Missolungi, Alessandro ;
Mutani, Bernardo ;
Richiardi, Lorenzo ;
Grignolo, Federico M. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 144 (05) :712-718
[9]   Asphericity of the anterior human cornea with different corneal diameters [J].
Gonzalez-Meijome, Jose Manuel ;
Villa-Collar, Cesar ;
Montes-Mico, Robert ;
Gomes, Antonio .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (03) :465-473
[10]   Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement [J].
Kawamorita, Takushi ;
Uozato, Hiroshi ;
Kamiya, Kazutaka ;
Bax, Leon ;
Tsutsui, Kenta ;
Aizawa, Daisuke ;
Shimizu, Kimiya .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01) :127-133