Preliminary validation of a mobile force Sensing device for clinical and telerehabilitation

被引:2
作者
Bechard, Laura [1 ]
Bell, Kevin [2 ,3 ]
Lynch, Andrew [4 ]
机构
[1] Univ Pittsburgh, Sch Hlth & Rehabil Sci, Dept Phys Therapy, Pittsburgh, PA 15219 USA
[2] Univ Pittsburgh, Sch Med, Dept Orthopaed Surg, Ferguson Lab Orthopaed Res, Pittsburgh, PA 15213 USA
[3] Univ Pittsburgh, Sch Engn, Dept Bioengn, Pittsburgh, PA 15213 USA
[4] Rutgers State Univ, Sch Hlth Profess, Dept Rehabil & Movement Sci, Blackwood, NJ 08012 USA
基金
美国国家科学基金会;
关键词
Muscle strength; Force production; Manual muscle testing; Electromechanical dynamometry; Hand-held dynamometer; STRENGTH;
D O I
10.1016/j.jbiomech.2020.109973
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Muscle strength and force production are important measures of patient progress during physical rehabilitation. Reliable and objective measurements are important to ascertain throughout rehabilitation. Current methods-manual muscle testing, electromechanical dynamometer, and hand-held dynamometer- are accurate and reliable, but have limitations that prevent wide implementation. As healthcare systems adapt to more patient-centered outcome models, changes to the delivery of rehabilitation, whether at-home or in the clinic, must also change to become more cost effective and accessible and provide quantifiable information regarding patient progress. We developed a novel Force Sensing (FoSe) device to quantify either tensile or compressive isometric muscle strength. The device was tested in a laboratory setting with healthy participants (n = 32) and compared to the commonly used hand-held dynamometer (HHD). Participants used both devices to perform several common isometric muscle tests including: hip abduction, knee extension, knee flexion, shoulder external rotation, and shoulder internal rotation. Compared to the HHD, FoSe was found to be an accurate and reliable measurement of force production. Intraclass Correlation Coefficients ranged from 0.58 to 0.89 without a magnitude dependent variation in force measurement. A second round of clinical testing with a patient population is warranted to determine FoSe's ability to measure clinically relevant asymmetry and progress over time. Further usability testing also needs to be conducted to determine the adequacy of FoSe for at-home use by both patients and clinicians. (c) 2020 Elsevier Ltd. All rights reserved.
引用
收藏
页数:6
相关论文
共 9 条
[1]   Manual muscle testing: does it meet the standards of an adequate screening test? [J].
Bohannon, RW .
CLINICAL REHABILITATION, 2005, 19 (06) :662-667
[2]  
Denegar CR., 1993, J SPORT REHABIL, V2, P35, DOI [DOI 10.1123/JSR.2.1.35, 10.1123/jsr.2.1.35]
[3]  
Ellenbecker TS, 2000, J ATHL TRAINING, V35, P338
[4]   CLINICAL RELIABILITY OF MANUAL MUSCLE TESTING - MIDDLE TRAPEZIUS AND GLUTEUS MEDIUS MUSCLES [J].
FRESE, E ;
BROWN, M ;
NORTON, BJ .
PHYSICAL THERAPY, 1987, 67 (07) :1072-1076
[5]   The Management of Injuries to the Medial Side of the Knee [J].
LaPrade, Robert F. ;
Wijdicks, Coen A. .
JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 2012, 42 (03) :221-233
[6]  
Portney L. G., 2000, FDN CLIN RES APPL PR, V2
[7]   Diagnostic Accuracy of Handheld Dynamometry and 1-Repetition-Maximum Tests for Identifying Meaningful Quadriceps Strength Asymmetries [J].
Sinacore, J. Anthony ;
Evans, Andrew M. ;
Lynch, Brittany N. ;
Joreitz, Richard E. ;
Irrgang, James J. ;
Lynch, Andrew D. .
JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 2017, 47 (02) :97-107
[8]   Muscle strength and hop performance criteria prior to return to sports after ACL reconstruction [J].
Thomee, Roland ;
Kaplan, Yonatan ;
Kvist, Joanna ;
Myklebust, Grethe ;
Risberg, May Arna ;
Theisen, Daniel ;
Tsepis, Elias ;
Werner, Suzanne ;
Wondrasch, Barbara ;
Witvrouw, Erik .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2011, 19 (11) :1798-1805
[9]  
WIKHOLM J B, 1991, Journal of Orthopaedic and Sports Physical Therapy, V13, P191