Functional intercomparison of intraoperative radiotherapy equipment - Photon Radiosurgery System

被引:28
作者
Armoogum, Kris S. [1 ]
Parry, John M. [1 ]
Souliman, Salam K. [1 ]
Sutton, David G. [1 ]
Mackay, Colin D. [1 ]
机构
[1] Tayside Univ Hosp NHS Trust, Dept Med Phys, Ninewells Hosp & Med Sch, Dundee, Scotland
关键词
D O I
10.1186/1748-717X-2-11
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Intraoperative Radiotherapy (IORT) is a method by which a critical radiation dose is delivered to the tumour bed immediately after surgical excision. It is being investigated whether a single high dose of radiation will impart the same clinical benefit as a standard course of external beam therapy. Our centre has four Photon Radiosurgery Systems (PRS) currently used to irradiate breast and neurological sites. Materials and methods: The PRS comprises an x-ray generator, control console, quality assurance tools and a mobile gantry. We investigated the dosimetric characteristics of each source and its performance stability over a period of time. We investigated half value layer, output diminution factor, internal radiation monitor (IRM) reproducibility and depth-doses in water. The half value layer was determined in air by the broad beam method, using high purity aluminium attenuators. To quantify beam hardening at clinical depths, solid water attenuators of 5 and 10 mm were placed between the x-ray probe and attenuators. The ion chamber current was monitored over 30 minutes to deduce an output diminution factor. IRM reproducibility was investigated under various exposures. Depth-dose curves in water were obtained at distances up to 35 mm from the probe. Results: The mean energies for the beam attenuated by 5 and 10 mm of solid water were derived from ICRU Report 17 and found to be 18 and 24 keV. The average output level over a period of 30 minutes was found to be 99.12%. The average difference between the preset IRM limit and the total IRM count was less than 0.5%. For three x-ray sources, the average difference between the calculated and actual treatment times was found to be 0.62% (n = 30). The beam attenuation in water varied by approximately I/r(3). Conclusion: The x-ray sources are stable over time. Most measurements were found to lie within the manufacturer's tolerances and an intercomparison of these checks suggests that the four x-ray sources have similar performance characteristics.
引用
收藏
页数:9
相关论文
共 21 条
[1]  
[Anonymous], 1970, 17 ICRU
[2]   Performance reproducibility of intra-operative radiotherapy equipment? Photon radiosurgery system [J].
Armoogum, K ;
Parry, J ;
Mackay, C ;
Souliman, S .
RADIOTHERAPY AND ONCOLOGY, 2005, 76 :S192-S192
[3]   Preclinical studies with the Photon Radiosurgery System (PRS) [J].
Astor, MB ;
Hilaris, BS ;
Gruerio, A ;
Varricchione, T ;
Smith, D .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (03) :809-813
[4]   Multicentricity and recurrence of breast cancer [J].
Baum, M ;
Vaidya, JS ;
Mittra, I .
LANCET, 1997, 349 (9046) :208-208
[5]   A new miniature x-ray device for interstitial radiosurgery: Dosimetry [J].
Beatty, J ;
Biggs, PJ ;
Gall, K ;
Okunieff, P ;
Pardo, FS ;
Harte, KJ ;
Dalterio, MJ ;
Sliski, AP .
MEDICAL PHYSICS, 1996, 23 (01) :53-62
[6]   Radiation properties of a miniature X-ray device for radiosurgery [J].
Biggs, DS ;
Thomison, ES .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (822) :544-547
[7]   Clinical relative biological effectiveness of low-energy x-rays emitted by miniature x-ray devices [J].
Brenner, DJ ;
Leu, CS ;
Beatty, JF ;
Shefer, RE .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02) :323-333
[8]   Interstitial Radiosurgery with the Photon Radiosurgery System in the Minimally-Invasive Treatment of Selected Deep-Seated Brain Tumors [J].
Di Lorenzo, N. ;
Cavedon, C. ;
Paier, F. ;
Compagnucci, A. ;
Buccoliero, A. M. ;
Pansini, G. ;
Francescon, P. ;
Gallina, P. .
JOURNAL OF CHEMOTHERAPY, 2004, 16 :70-74
[9]   Dosimetric characteristics of a low-kV intra-operative x-ray source: Implications for use in a clinical trial for treatment of low-risk breast cancer [J].
Ebert, MA ;
Carruthers, B .
MEDICAL PHYSICS, 2003, 30 (09) :2424-2431
[10]  
*EUR COMM, 1997, RAD PROT EUR COMM, V91